Medical device recalls
39226 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2018-09-19FDA-DeviceDiagnostica Stago, Inc.Class IIDiagnostica Stago STA N¿oplastine¿ CI Plus, 10 ml vial. In-Vitro Diagnostic for determination of the prothrombin time (PT) in plasma. (REF 00667) The STA¿ - N¿oplastine¿ CI Plus kits provide reagents for the determination of the prothrombin time (PT) in plasma with STA-R¿, STA Compact¿ and STA Satellite¿. (In the USA this procedure has been assigned to the moderate complexity category per CLIA 1988 - CDC Analyte Code 4922; CDC Test System Codes 4677 and 4875).
QC values outside of the assigned ranges (prolonged Prothrombin time / decreased PT %)
- 2018-09-12FDA-DeviceHitachi Healthcare Americas Corp Informatics DivisionClass IIVidiStar(TM) PACS & DICOM Viewer Software system
The secure filesystem client software used in the interface between the Vidistar PACS and an EHR system may cause intermixed images from multiple patients showing in a single study.
- 2018-09-12FDA-DeviceGE Medical Systems Ultrasound & Primary Care Diagnostics, LLClass IIGE Vivid E7 ultrasound system Product Usage: Vivid E7 is a general-purpose ultrasound system, specialized for use in cardiac imaging. It is intended for use by, or under the direction of a qualified physician for ultrasound imaging and analysis of Fetal/Obstetrics; Abdominal (including renal and GYN); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional; Musculo-skeletal Supe
The system side of the power cord may break and expose the electrical conductors leading to the risk of shock.
- 2018-09-12FDA-DeviceGE Medical Systems Ultrasound & Primary Care Diagnostics, LLClass IIGE Vivid 7 ultrasound system Product Usage: Vivid E7 is a general-purpose ultrasound system, specialized for use in cardiac imaging. It is intended for use by, or under the direction of a qualified physician for ultrasound imaging and analysis of Fetal/Obstetrics; Abdominal (including renal and GYN); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional; Musculo-skeletal Super
The system side of the power cord may break and expose the electrical conductors leading to the risk of shock.
- 2018-09-12FDA-DeviceGE Medical Systems Ultrasound & Primary Care Diagnostics, LLClass IIGE Vivid E95 ultrasound system Product The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal (including renal and GYN); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional; Urology (including prostate), Transesophageal; Transrectal (TR); Transvaginal (TV); and Intraoperative (abdominal, thoracic, & vascular).
The system side of the power cord may break and expose the electrical conductors leading to the risk of shock.
- 2018-09-12FDA-DeviceGE Medical Systems Ultrasound & Primary Care Diagnostics, LLClass IIGE Vivid E90 ultrasound system Product Usage: Vivid E80 / E90 / E95 ultrasound system is general-purpose ultrasound systems, specialized for use in cardiac imaging. It is intended for use by, or under the direction of a qualified physician for ultrasound imaging and analysis of fetal/Obstetrics; Abdominal (including renal and GYN); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conv
The system side of the power cord may break and expose the electrical conductors leading to the risk of shock.
- 2018-09-12FDA-DeviceGE Medical Systems Ultrasound & Primary Care Diagnostics, LLClass IIGE Vivid E80 ultrasound system Vivid E80 / E90 / E95 ultrasound system is general-purpose ultrasound systems, specialized for use in cardiac imaging. It is intended for use by, or under the direction of a qualified physician for ultrasound imaging and analysis of fetal/Obstetrics; Abdominal (including renal and GYN); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional; Urolo
The system side of the power cord may break and expose the electrical conductors leading to the risk of shock.
- 2018-09-12FDA-DeviceGE Medical Systems Ultrasound & Primary Care Diagnostics, LLClass IIGE Vivid E9 ultrasound system Product Usage: GE Vivid E9 ultrasound system is a general purpose ultrasound system for use in cardiac imaging. It is intended for use by, or under direction of a qualified physician for ultrasound imaging and analysis of Fetal; Abdominal (including renal and GYN); Pediatric; Small Organ (breast, Testes, thyroid; Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional; Urology (including prost
The system side of the power cord may break and expose the electrical conductors leading to the risk of shock.
- 2018-09-12FDA-DeviceGE Medical Systems Ultrasound & Primary Care Diagnostics, LLClass IIGE LOGIQ 9 ultrasound system Product Usage: The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic and vascular).
The system side of the power cord may break and expose the electrical conductors leading to the risk of shock.
- 2018-09-12FDA-DeviceGE Medical Systems Ultrasound & Primary Care Diagnostics, LLClass IIGE LOGIQ E9 ultrasound system Product Usage: The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic and vascular).
The system side of the power cord may break and expose the electrical conductors leading to the risk of shock.
- 2018-09-12FDA-DeviceAlden OpticalClass IIIZen RC Toric-Gas permeable contact lenses for Daily Wear in a 10-cc polycarbonate vial (Rx)as follows: Zen RC TPC (Toric Peripheral Curve) Zen RC FT (Front Toric Optics) Catalog Number: AOZNT Zenlens Toric - Zenlens contact lenses are indicated for daily wear for the correction of refractive arnetropia (myopia, hyperopia, and astigmatism) in aphakic and nonaphakic persons with non-diseased eyes.
Vials labeled for the prescriptions contained incorrect lenses
- 2018-09-12FDA-DeviceAlden OpticalClass IIIZenlens Toric-Gas Permeable Contact lenses for Daily Wear in a 10-cc polycarbonate vial (Rx) as follows: Zenlens FT(Front Toric); Zenlens Toric-Zenlens FT + MV (Front Toric+Microvault); Zenlens TPC-(Toric Peripheral Curve); Zenlens TPC + FT- (Toric Peripheral Curve + FrontToric) Zenlens TPC + MV- (Toric Peripheral Curve+Microvault). Catalog Number: AOZNT Zenlens Toric - Zenlens contact lenses are indicated for daily wear for the correction of refractive arnetropia (myopia, hyperopi
Vials labeled for the prescriptions contained incorrect lenses
- 2018-09-12FDA-DeviceMedline Industries, Inc.Class IILASER CDS
Multiple surgical packs were assembled with a non-sterile CHG applicators labeled as "ChloraPrep" (kit component numbers: 29183, 52115, 82279, and 82281). The non-sterile components 29183, 52115, 82279, and 82281 should have been placed inside the sterile portion of the pack, where it would subsequently be exposed to the ethylene oxide sterilization process. However, the components were not placed in the sterile portion of the pack during assembly and were never exposed to the sterilization process. Components 29183, 52115, 82279, and 82281 were inadvertently released non-sterile. The remainder of the pack components were properly sterilized and only components 29183, 52115, 82279, and 82281 were affected. Components 29183, 52115, 82279, and 82281 are manufactured by Becton, Dickinson (BD) and are labeled "ChloraPrep". ChloraPrep" applicators contain 2% w/v chlorhexidine gluconate (CHG) in 70% v/v isopropyl alcohol (IPA). The applicators are used to facilitate skin preparation for various preoperative and pre-injection procedures.
- 2018-09-12FDA-DeviceMedline Industries, Inc.Class IICardiac Insulation Pad, OPEN HEART CDS
Complete delivery System (CDS) # CDS984746C was assembled using a non-sterile cardiac insulation pad, Medline component number 32369. The non-sterile component (32369) was placed within the non-sterile portion of the CDS assembly. This component should have been placed inside the sterile portion of the pack where it would have subsequently been exposed to an ethylene oxide sterilization process. This component was not placed in the sterile portion of the pack and was not sterilized, resulting in the inadvertent release of this item in a non-sterile state. The contents of the sterile kit were properly sterilized and the remaining contents of the non-sterile subassembly had the appropriate sterility status based on their intended use.
- 2018-09-12FDA-DeviceMedline Industries, Inc.Class IILaparoscopy CDS
One manufacturing lot of a non-sterile pack was assembled with a non-sterile component labeled sterile (kit component #80532NS). The non-sterile component 80532NS was supposed to be placed inside the sterile portion of the pack where it would subsequently be exposed to the ethylene oxide sterilization process. However, the component was not placed in the sterile portion of the pack during assembly and was never exposed to the sterilization process. Component 80532NS was inadvertently released non-sterile. Component 80532NS is packaged by Medline and is an Anti-fog kit that contains 1 bottle of 6g Anti-Fog solution and an adhesive backed sponge.
- 2018-09-12FDA-DeviceMedline Industries, Inc.Class IIPICC Tray, STERILE, 12 packs each per case.
Multiple surgical packs were assembled with a non-sterile CHG disk labeled as BIOPATCH" (kit component #52655) attached to the outside of the sterile pack. This may pose potential risk of infection. However, the kits have been in distribution since March 2016. Many of these kits have been used with no reported infections or adverse patient events. In addition, the BIOPATCH is impregnated with CHG, a well-known antiseptic agent with broad-spectrum anti-microbial and antifungal activity.
- 2018-09-12FDA-DeviceMedline Industries, Inc.Class IIArt-Line Surgery Pack-LF, STERILE, 40 packs each per case.
Multiple surgical packs were assembled with a non-sterile CHG disk labeled as BIOPATCH" (kit component #52655) attached to the outside of the sterile pack. This may pose potential risk of infection. However, the kits have been in distribution since March 2016. Many of these kits have been used with no reported infections or adverse patient events. In addition, the BIOPATCH is impregnated with CHG, a well-known antiseptic agent with broad-spectrum anti-microbial and antifungal activity.
- 2018-09-12FDA-DeviceQuasar Bio-Tech, Inc.Class IIPure Rayz, Light Therapy System, packaging, box, sleeve and user manual. Handheld Wrinkles Reduction Device.
Quasar did not receive FDA clearance for claims of collagen/elastin production as stated on the packaging and in the user manuals.
- 2018-09-12FDA-DeviceQuasar Bio-Tech, Inc.Class IIQuasar MD Plus, Light Therapy System, packaging, box, sleeve and user manual. Handheld Wrinkles Reduction Device.
Quasar did not receive FDA clearance for claims of collagen/elastin production as stated on the packaging and in the user manuals.
- 2018-09-12FDA-DeviceQuasar Bio-Tech, Inc.Class IIBaby Quasar Plus, Light Therapy System, packaging, box, sleeve and user manual. Handheld Wrinkles Reduction Device.
Quasar did not receive FDA clearance for claims of collagen/elastin production as stated on the packaging and in the user manuals.
- 2018-09-12FDA-DeviceInspire Medical Systems Inc.Class IIInspire II Model 3024 Implantable Pulse Generator (IPG) for Obstructive Sleep Apnea, Inspire Part #900-003-020, Product Usage: The device is an implanted component of the Inspire Upper Airway Stimulation (UAS) system. The UAS therapy is used to treat a subset of patients with moderate to severe obstructive sleep apnea (OSA).
Incorrect use-by date on the device registration/patient file labels.
- 2018-09-12FDA-DeviceMcKesson Israel Ltd.Class IIMcKesson Cardiology Hemo, Release SW version 13.0. Used for physiological monitoring, image and data processing. McKesson Cardiology Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment. McKesson Cardiology Hemo is also intended for patient/procedural data management, such as documentation, logging, reporting, trending, storing, reviewing, carrying out clinical calculations and exporting v
Change Healthcare has identified an issue where, under certain circumstances, the Real Time Monitor (RTM) may not display physiological signals.
- 2018-09-12FDA-DeviceEncore Medical, LpClass IIFoundation Modular Femoral Stem, Part Number, 220-10-110. Knee prosthesis component. The Foundation Revision Knee is a total knee system indicated as knee joint replacement for patients suffering from pain and dysfunction.
After receipt of a product complaint for the labeling, it was determined that this device does not have FDA clearance.
- 2018-09-12FDA-DeviceExactech, Inc.Class IITibial Alignment Guide, REF 351-10-00 Vantage Fixed Ankle system components for ankle replacements.
Exactech is recalling the Tibial Alignment Guide because the Global Trade Item Number (GTIN) on the device is incorrect. The GTIN on the device reads 10885862525629, when it should read 10885862525628.
- 2018-09-12FDA-DeviceZimmer Biomet, Inc.Class IIPeg Fully Threaded, Locking, 2.5mm x 10mm, Item Number 131212610 Orthopedic device used for stabilization and fixation of small bone fragments Intended for stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion and reconstructions of small bones of the hand, foot, wrist, ankle, humerus, scapula, finger, toe, pelvis and craniomaxillofacial skeleton, particularly in osteopenic bone.
A size 22mm threaded peg was found in the package of a 10mm unit.
- 2018-09-12FDA-DeviceCanon Medical System, USA, INC.Class IIDRAD-3000E (Radrex-i) TFP-4336W (Wireless FPD)
It was found when an operator performs a radiography using the wireless flat panel detector (FPD), a message window was displayed on the monitor stating image transmission was not completed and there was no image. It also showed the OK button to reacquire image data form the FPD, and the Cancel button to cancel the reacquisition. When the operator selects the OK button, the same message window appears. The operator then repeated the same operation several times and finally selected the Cancel button to quit the reacquisition mode.
- 2018-09-12FDA-DeviceZimmer Dental IncClass IIOuter Package Label: Tapered Screw-Vent Implant System, 4.1mm X 10mm, 3.5mm, REF:TSVM4B10, Sterile:R, Rx Only; Inner Package Label: Implant, MTX, 3.7mm x 10mm,3.5mm, REF: TSVB10. Zimmer Dental Implants are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
Dental implants have incorrect item/lot numbers and size on the inner package labeling. Correct item/lot numbers and size are on the outer package labeling.
- 2018-09-12FDA-DeviceBayer Medical Care, Inc.Class IIOverhead Counterpoise Systems (OCS), OCS, 580mm CEILING MOUNT, PORTEGRA2, Catalog # OCS CEIL 58, Material Number 84433765
The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.
- 2018-09-12FDA-DeviceBayer Medical Care, Inc.Class IIOverhead Counterpoise Systems (OCS) OCS CONF an accessory device for use with Bayer injector systems as a support for an injector head. , SHORT CEIL MNT, STD HORZ ARM, Catalog # OCS 115AS, Material Number 59880288
The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.
- 2018-09-12FDA-DeviceBayer Medical Care, Inc.Class IIOverhead Counterpoise Systems (OCS), OCS CONF,STD CEIL MNT, STD HORZ ARM, Catalog # OCS 115A, Material Number 59880253
The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.
Get notified about new device recalls
Free weekly digest. We email when a new federal recall in this category — or anything else affecting your household — is published.
Get the Sunday Brief