- 2018-08-29FDA-DeviceEuro Diagnostica ABClass IIIis a quantitative/qualitative enzyme-linked immunosorbent assay (ELISA) for the detection of autoantibodies specific for Anti-Ro (SS-A) antigen in human serum or plasma. EDTA, lithium heparin, citrated plasma. It is intended to aid in the diagnosis of systemic rheumatic diseases, particularly Sj¿gren's syndrome and is not definitive in isolation. Autoantibody levels represent one parameter in a multi-criterion diagnostic process.
Incorrectly labeled Manufactured by DiaSorin, The correct term should be Distributed by DiaSorin.
- 2018-08-29FDA-DeviceEuro Diagnostica ABClass IIImmunoscan CCPlus test kit is an enzyme-linked immunosorbent assay (ELISA) for qualitative and semiquantitative determination of IgG antibodies to Cyclic Citrullinated Peptides (CCP) in human sera.
Decreasing performance overtime.
- 2018-08-29FDA-DeviceElekta, Inc.Class IIMOSAIQ Oncology Information System
Order Status Remains Approved When Should Indicate Complete.
- 2018-08-29FDA-DeviceJohnson & Johnson Vision Care, Inc.Class IIJohnson and Johnson Vision Care 1-Day ACUVUE MOIST Daily Wear, 8.5 Base Curve, -3.25 Power. Master Lot, containing 12 split lots, repackaged in 30 and 90-packs . Intended for daily disposable wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may have 1.00D or less of astigmatism.
Johnson and Johnson Vision Care received a number of reports of a partially formed contact lens, which appears to the customer or consumer as visibly smaller in diameter, thicker/stiffer to the touch and more darkly tinted blue than the normal contact lens.
- 2018-08-29FDA-DeviceRichard Wolf GmbHClass IIArtholine Punch B 4, 6MM H 1, 6MM, used for the removal of soft tissue, cartilage tissue or other non-osseous tissue via natural or surgically created passages. Product Usage: Device is a standard punch; standard punches are used for the removal of soft tissue, cartilage tissue or other non-osseous tissue via natural or surgically created passages.
Punch may have a burr in one section of the joint and when the punch is open, there is an increase potential for injury.
- 2018-08-29FDA-DevicePhilips Electronics North America CorporationClass IIPhilips SureSigns VS2 monitor, 863079 863080 863081 863082 863283 863286 The Sure Signs VS2 vital signs monitor is for use by healthcare professionals whenever there is a need for monitoring the physiological parameter of patients. Standard and optional parameters include: NBP, Sp02, Temperature. The Sure Signs VS2 vital signs monitor is for the monitoring, recording and alarming of multiple parameters of adult, pediatric and neonates in healthcare environments. Additionally, this monit
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.
- 2018-08-29FDA-DevicePhilips Electronics North America CorporationClass IIPhilips SureSigns VM 3/4/6/8 Patient Monitors, 863063 863064 863065 863066 863068 863077 863085 863086 863317 863287 863288 863289 The SureSigns VM4, VM6, and VM8 patient monitors are for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Standard and optional parameters include: " ECG " Respiration " NBP " S2pO " IBP " CO2 " Temperature
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.
- 2018-08-29FDA-DevicePhilips Electronics North America CorporationClass IIPhilips SureSigns VS4 Vital Signs Monitor, 863283 863286 The SureSigns VS4 vital signs monitor is for use by healthcare professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSigns VS4 monitor is for monitoring, recording, and alarming of multiple physiological parameters in healthcare environments for the patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.
- 2018-08-29FDA-DevicePhilips Electronics North America CorporationClass IIPhilips SureSigns VS3 Vital Signs Monitor, 863069 863070 863071 863072 863073 863074 The SureSigns VS3 monitor is for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility.
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.
- 2018-08-29FDA-DevicePhilips Electronics North America CorporationClass IIPhilips SureSigns VSV (Vital Signs Viewer), 863067 Indicated for central monitoring of multiple adult, pediatric and neonatal patients, and where the clinician decides to monitor cardiac arrhythmia of adult, pediatric and neonatal patients to gain information for the treatment, to monitor adequacy of treatment, or to exclude causes of symptoms.
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.
- 2018-08-29FDA-DeviceOrtho Kinematics, IncClass IIVertebral Motion Analyzer (VMA), 2.3.239 - 2.3.254. Product Usage: VMA software is a quantitative imaging software application intended to be used to process digital image files.
Ortho Kinematics notified customers that errors were contained in Radiological Read Report and VMA Report for the VMA test. The error resulted from an administrative error of VMA user account settings and has been fixed.
- 2018-08-29FDA-DeviceOrtho Kinematics, IncClass IIVertebral Motion Analyzer (VMA), VMA DataLink Software.2.3.231, 2.3.232. Product Usage: VMA software is a quantitative imaging software application intended to be used to process digital image files.
Ortho Kinematics, Inc. notified customers of erroneous results of VMA tests due to a software bug that has been corrected.
- 2018-08-29FDA-DeviceOrtho Kinematics, IncClass IIVertebral Motion Analyzer (VMA), 2.3.250. Product System, image processing, radiological
This email is to provide notification that, due to a software bug that has been corrected, the Radiological Read Report for the VMA test(s) contained an error.
- 2018-08-29FDA-DeviceCivco Medical Instruments Co. Inc.Class IICIVCO lnfiniti Needle Guide Bracket. used for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers. Product Usage: The bracket and disposable guide provide physicians with a tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers.
A cover failed leak testing during validation testing of new needle guides from a replacement tool.
- 2018-08-29FDA-DeviceCAREstream Medical LLCClass IICAREstream ProNox Nitrous Oxide Delivery System labeled as the following: a. CMI-0100 PNX, Pro-Nox Mixer - Canadian; b. CMI-0100-PNX-HS, Pro-Nox Complete Hospital Roll Stand System - Nitrous Oxide Delivery System; c. CMI-0100-PNX-HS-NO2H, Pro-Nox Hospital System without Oxygen Regulator; d. CMI-0100-PNX-HSWM, Pro-Nox Hospital System Wall Mount with Universal Mount (GCX) scavenger; e. CMI-0100-PNX-NHS, Pro-Nox Complete Roll Stand Nitrous Oxide Delivery System (Non-hospital
Two analgesic gas delivery system oxygen/nitrous oxide mixers were reported to have delivered higher levels of nitrous oxide than specification during routine testing.
- 2018-08-29FDA-DeviceArrow International IncClass IIPediatric Two-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter, Cat. No. AK-22502 Product Usage: The multiple-lumen catheter permits venous access to the central circulation of pediatric patients. The ARROWg+ard antimicrobial surface is intended to help provide protection against catheterrelated infections. The catheter is not intended to be used as a treatment for existing infections, nor is it indicated for long-term use. The use of the 4, 5, and 5.5 F
One lot of product is from product code AK-22502 so it contains a 2-lumen, 5 Fr x 5 cm catheter. However, this product is incorrectly labeled as AK-25502, so the lidstock says it contains a 2-lumen, 5 Fr x 8 cm catheter. No patient injuries have been reported related to this issue.
- 2018-08-29FDA-DeviceO-Two Medical Technologies, Inc.Class IIO-Two Equinox¿ Relieve N2O/O2 Analgesic Gas Mixing and Delivery System Model#: 01EQ1000
Higher levels of nitrous oxide delivered than specification
- 2018-08-29FDA-DeviceStryker GmbHClass IIStryker T2 ANKLE ARTHRODESIS NAIL RIGHT 10MMX200MM Item Number: 1819-1020S T2 ANKLE ARTHRODESIS NAIL RIGHT 12MMX200MM 1819-1220S T2 LONG PROXIMAL HUMERAL NAIL LEFT 240mm1832-2824S T2 LONG PROXIMAL HUMERAL NAIL LEFT 280mm 1832-2828S T2 LONG PROXIMAL HUMERAL NAIL RIGHT 260mm 1832-3826S T2 Ankle Arthrd Nail lft 1818-1115S
Incomplete seal on the Tyvek lid may compromise the sterility of the device
- 2018-08-29FDA-DeviceSmith & Nephew, Inc.Class IIGENESIS II NON-POROUS CRUCIATE RETAINING FEMORAL COMPONENT, SIZE 6 RIGHT, REF 71420026 Knee prosthesis
Two lots of Genesis II Non Porous C/R Femoral Size 6 right packaged were distributed containing Genesis II C/R Femoral Size 4 left.
- 2018-08-29FDA-DeviceOrtho Kinematics, IncClass IIVertebral Motion Analyzer (VMA) Version 2.3.252. VMA software is a quantitative imaging software application intended to be used to process digital image files.
Ortho Kinematics Inc. sent a Notice of Correction to Released Testing Results, Radiological Read Report for the Vertebral Motion Analyzer (VMA) test because it contained an error. The error occurred due to a software bug that has been corrected.
- 2018-08-29FDA-DeviceBack 2 Basics Direct LLCClass IIDymaxeon 5.5 Curved Rod Size 5.5 x 110mm (Yellow), Model Number 06.6441B Product Usage: The Dymaxeon Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral/iliac spine (T1 S1/Ileum): degenerative disc disease (defined as discogenic back pain with degeneration of disc confirmed by history and rad
Packages labeled as containing 110mm rods may contain 100mm rods.
- 2018-08-29FDA-DeviceIn2bones USA, LLCClass IICoLink(TM) Lapidus Plate XP, +2 mm, Left, REF P40 ST265, QTY 1, In2Bones Product Usage: The In2Bones USA LLC, CoLink¿ Plating System is indicated for stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction of the small bones in the hand, wrist, foot and ankle in both pediatric and adult patients.
CoLink(TM) Lapidus Plate +2mm Right and Left, were laser marked incorrectly. In these lots, right plates were laser marked as left plates and the left plates were laser marked as right plates.
- 2018-08-29FDA-DeviceIn2bones USA, LLCClass IICoLink(TM) Lapidus Plate XP, +2 mm, Right, REF P40 ST165, QTY 1, In2Bones Product Usage: The In2Bones USA LLC, CoLink¿ Plating System is indicated for stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction of the small bones in the hand, wrist, foot and ankle in both pediatric and adult patients.
CoLink(TM) Lapidus Plate +2mm Right and Left, were laser marked incorrectly. In these lots, right plates were laser marked as left plates and the left plates were laser marked as right plates.
- 2018-08-29FDA-DeviceTriMed Inc.Class IIUlnar Osteotomy Compression Plate, Short Model: UOCP-S Product Usage: The UOCP-S is provided non-sterile with the required fixation screws and surgical instrumentation. The TriMed Ulnar Osteotomy Plate is intended for use in osteotomy procedures of the ulna.
Two instrument holes located on the side of the plate, which act to align the saw guide are located on the wrong side of the plate
- 2018-08-29FDA-DeviceZimmer Biomet, Inc.Class IIPersona Partial Knee System Spacer Block Alignment Tower, Item No. 42-5399-006-35 Product Usage: Intended to aid in preparatory surgical procedures
The PPK System Spacer Block Alignment Tower mating feature that mates with the spacer block was undersized, which could potentially lead to mating issues between the devices. This could cause a delay in surgery of less than 30 minutes. There is an option in the surgical technique for performing limb alignment without the alignment tower.
- 2018-08-29FDA-DeviceZimmer Biomet, Inc.Class IIConnecting Bolt, Model 14-442093
Product was manufactured from the incorrect material, which shows a potential decrease in the fatigue strength of the connecting bolt that can lead to fracture.
- 2018-08-29FDA-DeviceBiomerieux IncClass IIbioM¿rieux VITEK 2 NH ID Test Kit Product Usage: Is an automated system consisting of instruments, software and reagent cards designed for the ID and antimicrobial susceptibility testing of bacteria and yeast.
Missing desiccant.
- 2018-08-29FDA-DeviceGE Healthcare Japan CorporationClass IIGE Healthcare Optima CT660. The systems are intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications. Sold only outside the U.S. under the following product names: CT OPTIMA 660, CT OPTIMA 660 LONG BJG SKD RUS, CT OPTIMA 660 LONG BJG SKD RUSSIA, CT OPTIMA 660 MID BJG SKD RUSSIA, CT OPTIMA 660 SE, CT OPTIMA 660 SE ASIR, CT660 M40 LONG, CT660 M40 MID, CT660 M40 MID WAUK SKD, CT660 M40 MID WAUK SKD BRAZ, CT660 M40 MID WAUK SKD BRAZ, OPT 660 M40 MID HINO SKD RUS
GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.
- 2018-08-29FDA-DeviceGE Healthcare Japan CorporationClass IIGE Healthcare Optima CT540. The systems are intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications.Sold under the following product names: BEIJING CT OPTIMA 540, BEIJING OPTIMA, BEIJING OPTIMA 540, BEIJING OPTIMA 540 ASIR, CT BEIJING OPTIMA 540, CT BEIJING OPTIMA 540 ASIR, CT GOLDSEAL OPTIMA 540, CT OPTIMA 540, CT OPTIMA 540 ASIR, CT OPTIMA CT540 TANG BEIJING, DISCOVERY 670 3/8 MEA PRO, HINO OPTIMA 540, HINO OPTIMA 540 ASIR, HINO OPTIMA 540 ASIR, OPTIMA 54
GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.
- 2018-08-29FDA-DeviceGE Healthcare Japan CorporationClass IIGE Healthcare Optima CT660. Sold under the following product names: BEIJING OPTIMA CT660 PRO GT170, BEIJING OPTIMA CT660 PRO GT200, BEIJING OPTIMA CT660S PRO 32 S, CT GOLDSEAL, CT GOLDSEAL OPTIMA 660, CT GOLDSEAL OPTIMA 660 UT-BUC, CT OPTIMA 660 LNG HN P20, CT OPTIMA 660 MID HN P20, CT OPTIMA 660 PRO, CT OPTIMA 660 PRO LONG BJG P2., CT OPTIMA 660 PRO MID, CT OPTIMA 660 PRO MID HN P20, CT OPTIMA 660S LNG HN P20, CT OPTIMA CT660 GT 1700, CT OPTIMA CT660 GT1700, CT OPTIMA CT660 GT1700V,
GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.