Medical device recalls
39229 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2018-04-25FDA-DeviceStryker Medical Division of Stryker CorporationClass IIPower-LOAD Cot Fastening System, Model #6390, Catalog #6390-000-000, UDI: 07613327261523 Power-LOAD is intended to assist with loading and unloading of a compatible wheeled stretcher (ambulance cot) to and from a transport vehicle and to secure the ambulance cot during transport. The device has a maximum safe working load of 870 lb (395 kg), which includes the weight of the ambulance cot, patient, and equipment attached to the cot (such as oxygen bottles, monitors, and pumps). The intended us
The power load floor plates which hold the ambulance cot anchors in place during patient transport, may be broken.
- 2018-04-25FDA-DeviceLife Technologies CorporationClass IIHuman Fetal Hemoglobin (HBF-1) FITC Conjugate
Three lots of Human Fetal Hemoglobin FITC Conjugate are displaying high background and unusual bright staining in the controls and patient samples which could potentially lead to erroneous results (i.e. false positives)
- 2018-04-25FDA-DevicePreservation Solutions, Inc.Class IICold Storage Solution, 2L, labeled as Waters IGL Cold Storage Solution, Catalog No. CSS/2000/US; and, Belzer UW Cold Storage Solution, Catalog No. BUW_002. Intended for the hypothermic flushing and storage of specific organs in preparation for storage, transportation, and transplantation into a recipient.
Bags of solution may have a small leak at the base of the injection port where the port is welded into the body of the bag..
- 2018-04-25FDA-DeviceSouthern Implants, (Pty.) Ltd.Class IISOUTHERN IMPLANTS REF IBR12d-13, Implant, Ext. Hex, tapered 12o Co-Axis(TM) intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment of crowns, bridges or overdentures utilizing delayed or immediate loading.
The UDI label correctly reads IBR12d-13, however, the inner label incorrectly read IBNT12d-13.
- 2018-04-25FDA-DeviceDatascope CorporationClass IIIntra-Aortic Balloon Catheter Mega 8FR 50cc Product Usage: The intra-aortic balloon catheter and accessories are used to provide counter pulsation therapy in the aorta, whereby balloon inflation during diastole and deflation during systole increases blood supply to the heart muscle and decreases the work of the left ventricle.
The U.S. Instructions For Use for the Mega Intra-Aortic Balloon (IAB) Catheter was missing the Contraindications section.
- 2018-04-25FDA-DeviceDatascope CorporationClass IIIntra-Aortic Balloon Catheter Mega 7.5FR 30cc Product Usage: The intra-aortic balloon catheter and accessories are used to provide counter pulsation therapy in the aorta, whereby balloon inflation during diastole and deflation during systole increases blood supply to the heart muscle and decreases the work of the left ventricle.
The U.S. Instructions For Use for the Mega Intra-Aortic Balloon (IAB) Catheter was missing the Contraindications section.
- 2018-04-25FDA-DeviceDatascope CorporationClass IIIntra-Aortic Balloon Catheter Mega 7.5FR 40cc Product Usage: The intra-aortic balloon catheter and accessories are used to provide counter pulsation therapy in the aorta, whereby balloon inflation during diastole and deflation during systole increases blood supply to the heart muscle and decreases the work of the left ventricle.
The U.S. Instructions For Use for the Mega Intra-Aortic Balloon (IAB) Catheter was missing the Contraindications section.
- 2018-04-25FDA-DeviceDraegar Medical Systems, Inc.Class IIInfinity Acute Care System (IACS) Monitoring Solution, Catalog Numbers: MS18841, MS20401, MS20407, MS20724, MS22956, MS25510, MS25520, MS25643, MS26196, MS26372, MS31818; UDI Information: 4049098054454, 4049098054447, 4049098054409, 4049098009799, 4049098009751 Multi-parameter, physiologic patient monitoring of adult, pediatric and neonatal patients in environments where patient care is provided by trained healthcare professsionals. The lACS obtains the physiologic, multi-parameter data from
The invasive pressure visual and audible alarms may not activate.
- 2018-04-25FDA-DeviceZimmer Biomet, Inc.Class IIZimmer Segmental System Proximal Femoral Component 38mm Offset, Model Number 00-5850-030-38 The Segmental System is used for solving complex salvage cases to address patients with severe bone loss associated with disease, trauma, or revision.
There is a potential for debris in the hole on the superior lateral aspect of the device as a result of machining which may not have been adequately removed during the subsequent cleaning process.
- 2018-04-25FDA-DeviceGetinge Disinfection AbClass IIGetinge AGS, FSLC, FSUC, RC accessory to Washer- Disinfector 88 Series UDI Info: 07340153700109, 07340153700116 Getinge automated loading/unloading systems are intended to be used as an automatic conveyor/transport system for unloading carriers (wash carts) from Getinge 86/88 series Washer-disinfectors. The automated loading/unloading systems in no way affect the cleaning/disinfection performance of the washer-disinfector. The automated loading/unloading equipment is considered an accessor
There is potential for a cart with disinfected goods to unload without the trolley being docked. Disinfected goods could end up falling to the floor causing risk to nearby users and damage to goods and equipment.
- 2018-04-25FDA-DeviceGetinge Disinfection AbClass IIGetinge AGS, FSLC, FSUC, RC accessory to Washer- Disinfector 86 Series UDI Info: 86 Series Washer Disinfector 07340153700246, 07340153700253, 07340153700260, 07340153700277 Getinge automated loading/unloading systems are intended to be used as an automatic conveyor/transport system for unloading carriers (wash carts) from Getinge 86/88 series Washer-disinfectors. The automated loading/unloading systems in no way affect the cleaning/disinfection performance of the washer-disinfector. The au
There is potential for a cart with disinfected goods to unload without the trolley being docked. Disinfected goods could end up falling to the floor causing risk to nearby users and damage to goods and equipment.
- 2018-04-25FDA-DeviceCook Inc.Class IIMINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
- 2018-04-25FDA-DeviceCook Inc.Class IIMINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000 US) G46022 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
- 2018-04-25FDA-DeviceWright Medical Technology, Inc.Class IIWRIGHT, Wright Medical Technology, VALOR(R) Nail, Right, Small, OD: 11.5mm, L: 150mm, REF415101115R
Tow lot of VALOR(TM) Nail could potentially be missing the assembled compression screw.
- 2018-04-25FDA-DeviceWright Medical Technology, Inc.Class IIWRIGHT, Wright Medical Technology, VALOR(R) Nail, Right, Medium, OD: 10mm, L: 200mm, REF415101020
Tow lot of VALOR(TM) Nail could potentially be missing the assembled compression screw.
- 2018-04-25FDA-DeviceTornier, IncClass IIFornier Phantom Fiber(TM) Sutures: Item Number Description SMB000425 4.5 MM PHANTOM FT BIOCOMPOSITE SUTURE ANCHOR WITH NEEDLES SMB000525 5.5 MM PHANTOM FT BIOCOMPOSITE SUTURE ANCHOR WITH NEEDLES SMP000425 4.5 MM PHANTOM FT PEEK SUTURE ANCHOR WITH NEEDLES SMP000525 5.5 MM PHANTOM FT PEEK SUTURE ANCHOR WITH NEEDLES SMSB0110 1PK PHANTOM FIBER SZ 2 WHITE SMSB0110N lPK BIOFIBER SUTURE SZ 2 WHT NEEDLED SMSB0112 10 PK PHANTOM FIBER SZ 2 WHITE SMSB0112N 10 PK PHANTOM FIBER SUTURE SZ 2WHT N
Data has shown that degradation of a raw material used in Phantom Fiber may occur more quickly than expected if improperly stored. This may impact the shelf-life of the product which could cause the suture to lose strength more quickly than expected.
- 2018-04-18FDA-DeviceTriMed Inc.Class IIHex Driver Part Number HXDRIVER-2.0 AO; Hex Driver Part Number HXDRIVR-2.5 AO; Hex Driver Part Number HXDRIVR-2.5CL AO; Hex Driver Part Number DVHX-2.5/130C AO; Hex Driver Part Number HXDRIVR-3.0 AO; Hex Driver Part Number HXDRIVER-3.5AO
Following the completion of sterilization cycle(s) of the impacted Hex Driver, the epoxy color band on the impacted Hex Driver can migrate down the shaft and/or potentially onto the sterilization tray or caddy.
- 2018-04-18FDA-DeviceHamilton Medical AGClass IIHamilton-C3 Ventilator Unit Intended to provide ventilator support to adults and pediatrics and optionally infants and neonates in the hospital.
After two and a half years, the battery fuel gauge may indicate a higher battery charge than what is available.
- 2018-04-18FDA-DeviceHamilton Medical AGClass IIHamilton-C2 Ventilator Unit Intended to provide ventilator support to adults and pediatrics and optionally infants and neonates in the hospital.
After two and a half years, the battery fuel gauge may indicate a higher battery charge than what is available.
- 2018-04-18FDA-DeviceTeleflex Medical Europe LtdClass IIBaker Jejunostomy Tubes: Product Code Equivalent Code (a) 655300160 655316 (b) 655500200 655520
These products contain latex, but the following statement is not included on any level of product packaging: Caution: This Product Contains Natural Rubber Latex Which May Cause Allergic Reactions.
- 2018-04-18FDA-DeviceTeleflex Medical Europe LtdClass IISimplastic Suprapubic Puncture Instruments: Product Code Equivalent Code (a) 650704100 551310 (b) 650704120 551312 (c) 650704160 551316 Product Usage: RUSCH Simplastic Suprapubic Puncture Instruments are indicated for emergency suprapubic drainage of the bladder.
These products contain latex, but the following statement is not included on any level of product packaging: Caution: This Product Contains Natural Rubber Latex Which May Cause Allergic Reactions.
- 2018-04-18FDA-DeviceTeleflex Medical Europe LtdClass IIRectal Catheters; Product Code Equivalent Code 580114 580114 Product Usage: Rectal Balloon Catheter is inserted into the rectum to measure the abdominal pressure as bladder filled.
These products contain latex, but the following statement is not included on any level of product packaging: Caution: This Product Contains Natural Rubber Latex Which May Cause Allergic Reactions.
- 2018-04-18FDA-DeviceTeleflex Medical Europe LtdClass IILatex Rebreathing Bags: Product Code Equivalent Code (a) 151174050 151174 (b) 153000005 153000005 (c) 153000010 153000010 (d) 153000020 153000020 (e) 153000030 153000030 (f) 153000050 153000050 (g) 210700050 2107050 (h) 210700102 2107102 (I) 210700202 2107202 (j) 210700302 2107302 (k) 210700402 2107402 (l) 210700502 2107502 (m) 210700602 2107602 Product Usage: Rebreathing Bag is indicated for manual patient ventilation or in connection with anaesthesia apparatus and/or patient
These products contain latex, but the following statement is not included on any level of product packaging: Caution: This Product Contains Natural Rubber Latex Which May Cause Allergic Reactions.
- 2018-04-18FDA-DeviceBeckman Coulter Inc.Class IIiQ200 Series Urine Microscopy Analyzer with Barcode Reader Model NFT-2100, Catalog #s: 800-3046, 800-3047, 800-3052, 800-3053, 800-3802, 800-3803, 800-3900, 800-3920, 800-3925, 800-3931, 800-3933, 800-3934, 800-3935, 800-3937, 800-3950, 800-3951, 800-3042, 800-3043, 800-3044, 800-3049, 800-3050, 800-7190, 800-7713, 800-7714, 800-7715, 800-7101, 800-7102, 800-3938, 800-7157
iQ200 Series Urine Microscopy Analyzer may intermittently fail to read urine sample dilution barcode labels causing erroneous results that could delay treatment for health conditions such as hematuria.
- 2018-04-18FDA-DeviceFresenius Medical Care Renal Therapies Group, LLCClass IILiberty Select Cycler (SW v.2.8.7), Material Number RTLR108343 Product Usage: The device is indicated for acute and chronic peritoneal dialysis.
The recalling firm identified a software issue related to the Patient Line Check (PLC) which may result in an increased risk of Overfill (also known as Increased Intraperitoneal Volume, IIPV). Overfill/IIPV may result in serious injury or death.
- 2018-04-18FDA-DeviceOrthofix, IncClass IIORTHOFIX, Connector System, LARGE SET SCREW, REF 79-2002
Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.
- 2018-04-18FDA-DeviceOrthofix, IncClass IIORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003
Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.
- 2018-04-18FDA-DeviceBeckman Coulter Inc.Class IIBECKMAN COULTER iQ200 Series Urine Microscopy Analyzer, Catalog Numbers: 700-3345, C10684, 700-3375, C10683, 700-3325, 700-3347, 700-3320, 700-3322, 700-3370
Beckman Coulter has determined that there is a potential for under-reporting casts. This can occur if per high-power field (/HPF) units of measurement for casts are selected in the iQ200 software, but the abnormal threshold and/or grading format is set up based on reporting per low-power field (/LPF)* or *per microliter. This may occur during initial method validation or if settings are altered after the initial validation.
- 2018-04-18FDA-DeviceZimmer GmbHClass IINormed Charcot Osteotome Chisel (Various sizes)
Product may contain a gap between the handle and metal chisel which can lead to failure to properly disinfect or sterilize between use.
- 2018-04-18FDA-DeviceAesculap Implant Systems LLCClass IIIStrauss Penis Clamp 130MM/General Instruments
The plastic sleeves of the Penis Clamp and Mouth Gag may stick together during or after reprocessing rendering the instruments unusable.