- 2018-03-07FDA-DeviceSteris CorporationClass IIVis-U-All High Temperature Tubing 9"x100'. One roll per box; 10 boxes per case 6 rolls/cs Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed to maintain sterility of properly processed medical devices during normal handling and storage until the pouch is opened and the medical device is removed for use.
On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
- 2018-03-07FDA-DeviceSteris CorporationClass IIVis-U-All High Temperature Tubing 6"x100'. One roll per box; 10 boxes per case 10 rolls/cs Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed to maintain sterility of properly processed medical devices during normal handling and storage until the pouch is opened and the medical device is removed for use.
On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
- 2018-03-07FDA-DeviceSteris CorporationClass IIVis-U-All High Temperature Tubing 4"x100'. One roll per box; 10 boxes per case 10 rolls/cs Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed to maintain sterility of properly processed medical devices during normal handling and storage until the pouch is opened and the medical device is removed for use.
On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
- 2018-03-07FDA-DeviceSteris CorporationClass IIVis-U-All High Temp 7.5"x13" Self Seal Pouch 200 pouches per box; 4 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed to maintain sterility of properly processed medical devices during normal handling and storage until the pouch is opened and the medical device is removed for use.
On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
- 2018-03-07FDA-DeviceAngiodynamics, Inc.Class IISoft-Vu Omni Flush Angiographic Catheter (UPN H787107322015); SV OF 5F x 65 cm .035 NB PG; Catalog Number: 10732201 AngioDynamics Angiographic Catheters are for use where angiographic diagnosis is indicated.
Product was placed into distribution prior to completion of all required post sterilization release activities.
- 2018-03-07FDA-DeviceCosmedent, Inc.Class IICosmedent's Etching Gel, 3g, 37% Phosphoric Acid, Rx only. Packaged in 3g Luer Lock syringes. Used for etching enamel and dentin.
Product packaged in a syringe that is not a lure lock syringe and the tip cold come off when expressing the material.
- 2018-03-07FDA-DeviceVentana Medical Systems IncClass INEXES VEN IVIEW DAB DET KT JPN-US EXPORT, Ventana Part Number 760-041, Roche GMMI 05266084001 in vitro diagnostic
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- 2018-03-07FDA-DeviceVentana Medical Systems IncClass IultraView SISH DNP Detection Kit, Ventana Part Number 800-098, Roche GMMI 05907136001 in vitro diagnostic
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- 2018-03-07FDA-DeviceVentana Medical Systems IncClass IHematoxylin II, Ventana Part Number 790-22087, Roche GMMI 05277965001 in vitro diagnostic
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- 2018-03-07FDA-DeviceVentana Medical Systems IncClass IOptiView Amplification Kit (250 Test), Ventana Part Number 860-099, Roche GMMI 06718663001 in vitro diagnostic
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- 2018-03-07FDA-DeviceVentana Medical Systems IncClass IOptiView Amplification Kit, Ventana Part Number 760-099, Roche GMMI 06396518001 in vitro diagnostic
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- 2018-03-07FDA-DeviceVentana Medical Systems IncClass ICINtec PLUS Cytology Kit (Canada/Japan), Ventana Part Number 805-100, Roche GMMI 06889549001 in vitro diagnostic
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- 2018-03-07FDA-DeviceVentana Medical Systems IncClass ICINtec PLUS Cytology Kit (CE-IVD), Ventana Part Number 605-100, Roche GMMI 06889565001 in vitro diagnostic
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- 2018-03-07FDA-DeviceVentana Medical Systems IncClass IultraView SISH DNP Detection Kit US, Ventana Part Number 760-098, Roche GMMI 05572037001 in vitro diagnostic
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- 2018-03-07FDA-DeviceVentana Medical Systems IncClass IultraView SISH Detection Kit, Ventana Part Number 780-001, Roche GMMI 05271967001 in vitro diagnostic
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- 2018-03-07FDA-DeviceVentana Medical Systems IncClass IiView DAB IHC Detection Kit, Ventana part Number 760-091, Roche GMMI 05266157001 in vitro diagnostic
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- 2018-03-07FDA-DeviceVentana Medical Systems IncClass IultraView Universal DAB Detection Kit, Ventana part Number 760-500, Roche GMMI 05269806001 in vitro diagnostic
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- 2018-03-07FDA-DeviceVentana Medical Systems IncClass IOptiView DAB IHC Detection Kit, Ventana Part Number 760-700, Roche GMMI 06396500001 in vitro diagnostic
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- 2018-03-07FDA-DevicePerkinelmer Life Sciences, Inc.Class IIIJANUS & JANUS G3 Automated Workstation is an automated programmable liquid handling instrument intended for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation work flows.
We have become aware through internal testing that JANUS and Janus G3 Automated Workstations were failing production final acceptance testing due to a Horizontal Motion Error message displayed on the system. Further investigation concluded that two printed circuit boards (PCB) used in MDT models of Janus and Janus G3 were manufactured with the incorrect resistor values. This anomaly affects component part number 7101640 (PCB-Quad Motor Controller MDT) and part number 7101647 (PCB-Quad Motor Controller Gripper). This anomaly could result in the Horizontal Motion Error being displayed and result in instrument down time. Until PerkinElmer s corrective measures are completed, we request you; ¿ Be aware that there is potential for the Horizontal Motion Error message to display which could result in instrument down time. ¿ Follow guidance provided in the Janus Automated Workstation User Manual when the Horizontal Motion Error message displays. ¿ If you have additional questions, contact PerkinElmer Technical Support for further assistance (800) 762-4000.
- 2018-03-07FDA-DeviceKeystone Dental IncClass IIStage-1¿ RDS Indexed Abutment for RDS Stage-1¿ Single Stage Implants Keystone s implants are intended for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cement retained, screw retained, or overdenture restorations and terminal or intermediate abutment support for fixed bridgework.
The recalled units are mislabeled with an incorrect expiration date. The expiration date printed on the label is 2202-06. The correct expiration date should be 2022-06.
- 2018-03-07FDA-DeviceTerumo Medical CorporationClass IIDestination Guiding Sheath catheter introducer, Catalog Number: RSR06 Product Usage: The Destination¿ Guiding Sheath is designed to be used for the introduction of interventional and diagnostic devices into the human vasculature, including but not limited to the renal, carotid, and peripheral arteries.
The tucking depth feature of this catheter introducer is marginally outside the upper specification.
- 2018-03-07FDA-DeviceElekta LimitedClass IIIntegrity R1.2
Potential for positional errors following automatic table movement (ATM).
- 2018-03-07FDA-DeviceElekta LimitedClass IIAgility with Integrity R3.2
Potential for positional errors following automatic table movement (ATM).
- 2018-03-07FDA-DeviceZimmer Biomet, Inc.Class IIVersys 6 inch Beaded Full Coat Plus Hip Prosthesis. Labeled as follows: a. VERSYS 6 INCH BEADED FC 19X160MM STD BODY STD NECK (Item No. 00894301906); b. VERSYS 6 INCH BEADED FC 21X160MM STD BODY STD NECK (Item No. 00784302106); c. VERSYS 6 INCH BEADED FC 22X160MM STD BODY STD NECK (Item No. 00784302206); d. VERSYS 6 INCH BEADED FC STEM 19X160MM LM (Item No. 00784301936); e. VERSYS 6 INCH BEADED FC STEM 20X160MM LM (Item No. 00784302036); f. VERSYS 6 INCH BEADED FC STEM 21X160M
The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.
- 2018-03-07FDA-DeviceZimmer Biomet, Inc.Class IIZimmer Segmental System, ZSS Distal Femur. Sizes as follows: a. Size B-LT (Item No. 00585001201); b. Size B-RT (Item No. 00585001202); c. Size C-LT (Item No. 00585001301); d. Size C-RT (Item No. 00585001302) Components of the system are intended to replace the proximal femur, mid-shaft femur, distal femur, proximal tibia, and/or total knee in cases that require extensive resection and restoration.
The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.
- 2018-03-07FDA-DeviceSiemens Healthcare Diagnostics, IncClass IIADVIA Centaur¿ T3 ADVIA Centaur T3: For in vitro diagnostic use in the quantitative determination of triiodothyronine (T3) in serum using the ADVIA Centaur, ADVIA Centaur XP, and ADVIA Centaur XPT systems. Measurements of triiodothyronine are used in the diagnosis and treatment of thyroid disease.
Siemens has identified an issue when scanning the 20 barcode for the Master Curve Card (MCC) with ADVIA Centaur systems T3 kit lots ending in 198 on the ADVIA Centaur XPT system.
- 2018-03-07FDA-DeviceMedistim AsaClass IIMEDISTIM VERIQ SYSTEM, MODEL VQ4122 An intraoperative diagnostic system that utilizes ultrasonography to guide surgeons to successfully plan and accomplish surgical interventions.
A system component overheated causing melting of the connector and emission of smoke through the systems loudspeaker opening.
- 2018-03-07FDA-DeviceKeystone Dental IncClass IIMax Implant System - Bone Tap for use with 7.0mm diameter Max Implant, Catalog Number D-TAP-MAX7-7 Dental implants are intended for use as anchors of fixed or semi-fixed dental crowns, bridges, and overdentures in patients with partial or full edentulism of the upper or lower jaw.
An 8mm diameter bone tap was laser marked as 7mm diameter and was packed with a 7mm label.
- 2018-03-07FDA-DevicePhilips Electronics North America CorporationClass IIAllura Xper FD and and Xper FD ORT - Philips X-ray Systems Product Usage: The Allura Xper FD10 and Alllura Xper FD10/10 is intendd for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and ElectroPhysiology (EP). The Allura Xper FD20, Allura Xper FD20/10, Allura Xper FD20/15 and Allura Xper FD10/10 is intended for: Dedicated vascular and neurovascular imaging
Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of the support arm can come loose because of vibrations and forces exerted on them. If the set screws have come loose gravity can keep the rods in position. However, if a force is exerted on the monitor pushing it forward, the monitor can topple forward. Once the monitor has toppled over it is only attached with its wires. If the monitor breaks loose of its wires it will drop on the floor.
- 2018-03-07FDA-DevicePhilips Electronics North America CorporationClass IIAllura Xper systems with Certeray X-ray Generator located in regions with high mains (480V) voltage. Product Usage: The Allura Xper series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. Cardiac imaging appl
There is an increase in the failure rate of certain Anode Drive Units (ADU5) used in these products. Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may also generate a peculiar burning smell that may be noticed. In some instances this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality is reduced.