Medical device recalls
39229 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2018-03-07FDA-DeviceMobius Imaging, LLCClass IIAIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO¿ is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding pediatric patients and patients weighing over 400 lbs. (182 kg).
If the tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended or unpredictable device motion while the CT system is homing the axis at startup or preparing for a scan.
- 2018-03-07FDA-DeviceSiemens Medical Solutions USA, IncClass IISOMATOM Go.Now computed tomography (CT) x-ray system table, Model Number: 11061618 Product Usage: Intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by these systems can be used by a trained physician as an aid in diagnosis. These CT systems can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies. Please refer to
The table may become unstable if the nuts on the bolts were not adequately tightened and thus have the potential to loosen and several bolts may gradually loosen at the same time.
- 2018-03-07FDA-DeviceSiemens Medical Solutions USA, IncClass IISOMATOM Go.Up computed tomography (CT) x-ray system table, Model Number: 11061628 Product Usage: Intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by these systems can be used by a trained physician as an aid in diagnosis. These CT systems can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies. Please refer to c
The table may become unstable if the nuts on the bolts were not adequately tightened and thus have the potential to loosen and several bolts may gradually loosen at the same time.
- 2018-03-07FDA-DeviceAtrium Medical CorporationClass IIProLite Mesh - mesh, surgical, polymeric ProLite Mesh is intended for use in soft tissue deficiencies including hernia repair, traumatic or surgical wounds and chest wall reconstruction procedures requiring reinforcement with a non-absorbable supportive material.
An incorrect ProLite Mesh configuration was packaged as product code 1010306-06, with lot number 412298. ProLite Mesh manufactured as product code 1010306-04 was inadvertently placed inside the pouch of packages labeled with product code 1010306-06 and lot number 412298.
- 2018-03-07FDA-DeviceSiemens Medical Solutions USA, IncClass IIAxiom Artis model Interventional fluoroscopic x-ray foot switch accessory; Model Numbers: 4787797, 4787805, and 4787813.
The possibility exists for the wireless foot switch to fail due to impacts by external factors, such as electrostatic discharge that exceed a certain intensity. If the foot switch fails, it will no longer be able to be used to release radiation. The hand switch provided can be used to release exposures; however, fluoroscopic examinations will not be possible.
- 2018-03-07FDA-DeviceSiemens Medical Solutions USA, IncClass IIArtis zee and Artis zeego model Interventional fluoroscopic x-ray foot switch accessory; Model Numbers: 4787797, 4787805, and 4787813.
The possibility exists for the wireless foot switch to fail due to impacts by external factors, such as electrostatic discharge that exceed a certain intensity. If the foot switch fails, it will no longer be able to be used to release radiation. The hand switch provided can be used to release exposures; however, fluoroscopic examinations will not be possible.
- 2018-03-07FDA-DeviceSiemens Medical Solutions USA, IncClass IIArtis Q and Q.zen model Interventional fluoroscopic x-ray system wireless foot switch accessory; Model Numbers: 4787797, 4787805, and 4787813.
The possibility exists for the wireless foot switch to fail due to impacts by external factors, such as electrostatic discharge that exceed a certain intensity. If the foot switch fails, it will no longer be able to be used to release radiation. The hand switch provided can be used to release exposures; however, fluoroscopic examinations will not be possible.
- 2018-03-07FDA-DeviceSiemens Medical Solutions USA, IncClass IIAxiom Artis, Model Numbers: 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555373, 7727717, 7728350 & 7728392
Potential for liquid entry
- 2018-03-07FDA-DeviceSiemens Medical Solutions USA, IncClass IIArtis Zee and Zeego, Model Numbers: 10094135, 10094137, 10094141, 10094142, 10094143, 10280959, 10848283 & 10848354
Potential for liquid entry
- 2018-03-07FDA-DeviceSiemens Medical Solutions USA, IncClass IIArtis Q and Q.Zen, Model Numbers:10848280, 10848281, 10848282, 10848353 & 10848355
Potential for liquid entry
- 2018-03-07FDA-DeviceTeleflex MedicalClass IIRusch Greenlite MAC 1, Catalog Number 004551001. Used during tracheal intubation.
The pouch may contain the incorrect Rusch Green Lite product size. Pouches that are labeled as containing RUSCH GREEN LITE MAC 1 may instead contain RUSCH GREEN LITE MAC 2.
- 2018-03-07FDA-DevicePhilips Electronics North America CorporationClass IIdigital x-ray detector ProGrade R1 - solid state X ray imager (flat panel/digital imager) As a part of a radiographic system, the Philips ProGrade is intended to acquire, process, store, display, and export digital radiographic images. The Philips ProGrade is suitable for all routine radiographic examinations, including specialist area like intensive care, trauma, or pediatric work, excluding mammography. United States only: The Eleva Workspot is not intended for fluoroscopy and angiograph
It the WiFi connection between the SkyPlate detector and HP transfer point is weak, an image may fail to transfer from the SkyPlate detector to the system. The image remains in the memory of the detector, but cannot be transferred wirelessly or by use of the backup cable. To continue, the operator can reset the SkyPlate detector by removing its batteries, but the acquired image is lost and a re-take is necessary.
- 2018-03-07FDA-DeviceOakworks IncClass IIFoot Controls impacted for the following devices: OAKWORKS¿ Medical CFPM 300 Imaging Table OAKWORKS¿ Medical CFPM 301 Imaging Table OAKWORKS¿ Medical CFPM 400 Imaging Table OAKWORKS¿ Medical CFPM 401 Imaging Table OAKWORKS¿ Medical CFPMB 301 Bariatric Imaging Table OAKWORKS¿ Medical CFUR 301 Urology Table OAKWORKS¿ Medical CFUR 401 Urology Table OAKWORKS¿ Medical CFLU 401 Lithotripsy/Urology Table
The foot control used on the OAKWORKS¿ Inc., CFPM300, CFPM301, CFPM400, CFPM401, CFPMB301, CFUR301, CFUR401 and CFLU401 Imaging Tables can become lodged under the base or under the column of the table.
- 2018-03-07FDA-DeviceZevex Incorporated (dba Moog Medical Medical Devices Group)Class IIMOOG Curlin Infusion Administration Set, Ref 340-4169, Non-DEHP Microbore Tubing with Non-Vented Bag Spike and distal Ultrasite Needleless Injection Port (Positive Pressure), Anti-Siphon Valve included for Free Flow Protection, packaged 20/box, for use with the Moog Curlin Ambulatory Volumetric Infusion Pump, RX, Sterile. Product Usage: Administration set to be used with the Moog Curlin infusion Pump. The pump can be used for intravenous, intra-arterial, epidural, subcutaneous, or enteral
Particulate found on the Curlin spike.
- 2018-03-07FDA-DeviceZevex Incorporated (dba Moog Medical Medical Devices Group)Class IIMOOG Curlin Infusion Administration Set, Ref 340-4168, Non-DEHP Microbore Tubing with Non-Vented Bag Spike and distal Female Luer Y-Site with Check Valve, Anti-Siphon Valve included for Free Flow Protection, packaged 20/box, for use with the Moog Curlin Ambulatory Volumetric Infusion Pump, RX, Sterile. Product Usage: Administration set to be used with the Moog Curlin infusion Pump. The pump can be used for intravenous, intra-arterial, epidural, subcutaneous, or enteral infusions.
Particulate found on the Curlin spike.
- 2018-03-07FDA-DeviceZevex Incorporated (dba Moog Medical Medical Devices Group)Class IIMOOG Curlin Infusion Administration Set, Ref 340-4166, Non-DEHP Microbore Tubing with Non-Vented Spike, Anti-Siphon Valve included for Free Flow Protection, packaged 20/box, for use with the Moog Curlin Ambulatory Volumetric Infusion Pump, RX, Sterile. Product Usage: Administration set to be used with the Moog Curlin infusion Pump. The pump can be used for intravenous, intra-arterial, epidural, subcutaneous, or enteral infusions.
Particulate found on the Curlin spike.
- 2018-03-07FDA-DeviceZevex Incorporated (dba Moog Medical Medical Devices Group)Class IIMOOG Curlin Infusion Administration Set, Ref 340-4134, Non-DEHP Microbore Tubing with Non-Vented Bag Spike and Y-Site with Check Valve, packaged 20/box, for use with the Moog Curlin Ambulatory Volumetric Infusion Pump, RX, Sterile. Product Usage: Administration set to be used with the Moog Curlin infusion Pump. The pump can be used for intravenous, intra-arterial, epidural, subcutaneous, or enteral infusions.
Particulate found on the Curlin spike.
- 2018-03-07FDA-DeviceZevex Incorporated (dba Moog Medical Medical Devices Group)Class IIMOOG Curlin Infusion Administration Set, Ref 340-4133, Non-DEHP Microbore Tubing with Non-Vented Bag Spike and Needleless Injection Port (Positive Pressure), packaged 20/box, for use with the Moog Curlin Ambulatory Volumetric Infusion Pump, RX, Sterile. Product Usage: Administration set to be used with the Moog Curlin infusion Pump. The pump can be used for intravenous, intra-arterial, epidural, subcutaneous, or enteral infusions.
Particulate found on the Curlin spike.
- 2018-03-07FDA-DeviceZevex Incorporated (dba Moog Medical Medical Devices Group)Class IIMOOG Curlin Infusion Administration Set, Ref 340-4114, Non-DEHP Microbore Tubing with Non-Vented Bag Spike, packaged 20/box, for use with the MOOG Curlin Ambulatory Volumetric Infusion Pump, RX, Sterile. Product Usage: Administration set to be used with the Moog Curlin infusion Pump. The pump can be used for intravenous, intra-arterial, epidural, subcutaneous, or enteral infusions.
Particulate found on the Curlin spike.
- 2018-03-07FDA-DeviceOlympus Corporation of the AmericasClass IIA20975A "Working insert, with ramp, one way" (Lots 13ZW to 172W; A20976A "Bridge, one way" (Lots 146W to 172W) and A20977A "Bridge, two way" (Lots 146W to 174W) Product Usage: A20975A: Working insert for endoscopic diagnosis and treatment in urologic applications. A20976A: Bridge for endoscopic diagnosis and treatment in urologic applications. A20977A: Bridge for endoscopic diagnosis and treatment in urologic applications.
Olympus has received complaints about fragments of adhesive which detached from inside the working channel of the referenced products. Cracking, chipping, missing pieces, and delamination of the adhesive have also been reported. In one case, a fragment of the adhesive was detected inside a patient's urethra during a procedure and was retrieved without injury to the patient. The manufacturer, Olympus Winter & lbe GmbH ("OWi") has determined that adhesive can be dislodged during the intended use of the cystoscopy bridge or working insert, e.g., when inserting an instrument through the working channel. As a result, a fragment of the adhesive may fall inside the patient's bladder, urethra or uterus and would need to be retrieved, extending the length of the procedure or requiring additional surgical treatment. The adhesive is used to seal gaps inside the working channel of the cystoscopy bridges and working inserts. In 2013, the adhesive used until then was discontinued and had to be replaced. Current investigations revealed that the durability of the new adhesive is insufficient
- 2018-03-07FDA-DeviceDePuy Mitek, Inc., a Johnson & Johnson Co.Class II1113 Gravity Tube Set Pk/24 -SH Product Usage: The FMS Gravity Tube Set is intended to deliver irrigation fluid from irrigation containers to the operative site during arthroscopic procedures.
Leakage of FMS gravity tub sets (281113) at the bulb and tube connection
- 2018-03-07FDA-DeviceIntegra LifeSciences Corp.Class IIFootswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation System; Catalogue No: C7000 Product Usage: The footswitch is a component of the CUSA Clarity Ultrasonic Surgical Aspirator System, which is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable. The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in: Neurosurgery, Plastic and Reconstructive surgery, Orthopedic Surgery, Gynecological Surgery and
A faulty footswitch may result in unintended ultrasonic fragmentation activation. There are two scenarios when the failure mode may occur: 1) Ultrasonic fragmentation may be inadvertently activated at start-up or at the initial use once the footswitch is connected to the console and after the user primes the system, performs a handpiece test, and enters the Main Screen. 2) Ultrasonic fragmentation may remain activated during surgical use once the foot pedal is disengaged slowly.
- 2018-03-07FDA-DeviceEthicon, Inc.Class IICoated VICRYL RAPIDETM (Polyglactin 910) Suture
On August 17, 2017, Ethicon had decided to initiate a medical device recall (removal) of the twenty-five (25) lots of VICRYL RAPIDETM (Polyglactin 910) Suture that were distributed to the field. During finished goods testing, a suture damage was identified on a lot of VICRYL RAPIDETM Suture. During the course of the subsequent investigation, the suture also failed knot tensile strength testing.
- 2018-03-07FDA-DeviceStryker CommunicationsClass IIBerchtold Chromophare Dual Monitor Carrier Arm or Dual Flat Panel, a component of Berchtold Chromophare Ceiling Mounted Surgical Light System.
The mounts on the monitor may have an insufficient (missing or incomplete) weld.
- 2018-03-07FDA-DeviceCooperVision Inc.Class IIIBiofinity XR Toric contact lenses 8.7/14.5 - 6PK Biofinity XR Toric (comfilcon A) Soft (Hydrophilic) Contact Lenses are indicated for the correction of ametropia (myopia or hyperopia) with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -5.75 diopters. The Biofinity (comfilcon A) Soft (Hydrophilic) Contact Lenses have been approved for extended wear for up to 6 nights /7 days of continuo
The firm identified a labeling error affecting Made-To-Order (MTO) Biofinity XR Toric blister labels. The primary blister reads Biofinity Multifocal Toric in error, and should read Biofinity XR Toric. The outer carton correctly reads Biofinity XR Toric.
- 2018-02-28FDA-DeviceArrow International IncClass IIARROW¿ PICC powered by ARROW¿ VPS¿ Stylet
The kit may contain an incorrect catheter component. Kits that should contain a 40cm catheter may contain a 55cm catheter.
- 2018-02-28FDA-DevicePhilips Consumer LifestyleClass IIPhilips Avent Smart Ear Thermometer (SCH740), Philips Ear Thermometer (DL8740)
Device may provide inaccurate measurements.
- 2018-02-28FDA-DeviceMerit Medical Systems, Inc.Class IIPreludeSNAP Splittable Hydrophilic and non-Hydrophilic Sheath Introducers, Catalog Numbers - PLS-1007, PLS-1008, PLS-1009, PLS-1009.5, PLS-1010, PLS-1010.5, PLS-1011, PLS-2506, PLS-2507, PLS-2508, PLS-2509, PLS-2510, PLS-2510.5, PLSH-1006, PLSH-1009, PLSX-1006, PLSX-1007, PLSX-1009, PLSX-1009.5
Merit Medical Systems, Inc. is voluntarily conducting a recall of specific lots of Prelude SNAP Splittable Sheath Introducers due to an intermittent failure of the splittable hub.
- 2018-02-28FDA-DeviceCosman Medical, LLCClass IINitinol TC Reusable Electrodes (TCN), Model Numbers: 1)TCN-10, 2) TCNK-10-R*, 3)TCN-10-3M, 4) TCNK-15, 5) TCN-15, 6) TCNK-15-C, 7) TCN-15-3M, 8) TCNK-15-R*, 9) TCN-20, 10) TCNK-20, 11) TCN-20-3M*, 12) TCNK-20-C*, 13) TCN-5, 14) TCNK-20-R*, 15) TCN-5-3M, 16) TCNK-5, 17) TCNK-10, 18) TCNK-5-C, 19) TCNK-10-C & 20) TCNK-5-R*. The TCN devices are indicated for use in radiofrequency (RF) heat lesioning of peripheral nerve tissue only.
After multiple reprocessing cycles, the epoxy resin which holds the TCN Electrode in the hub can exhibit signs of damage. In some cases, this damage may result in the inability to fully remove blood and/or tissue residuals prior to cleaning and re-sterilizing the device.
- 2018-02-28FDA-DeviceBoston Scientific CorporationClass IIInnova Self-Expanding Stent System, 180mm. Labeled with Material Number and description as follows: Product Usage: The Innova Vascular Self-Expanding Stent System is indicated to improve luminal diameter in the treatment of symptomatic de-novo or restenotic lesions in the native superficial femoral artery (SFA) and/or proximal popliteal artery (PPA) with reference vessel diameters from 4.0 mm to 7.0 mm and lesion lengths up to 190 mm. a. H74939180051870 (Material Description Innova-E
Stent possibly unable to be fully released from the delivery system.