- 2018-02-21FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIIngenuity TF PET/CT (model 882442) running software version 4.0.2 This device is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data. The PET subsystem produces images of the distribution of PET radiopharmaceuticals in the patient body (specific radiopharmaceuticals are used for whole body, brain, heart and ot
A software issue causes PET reconstructions to fail intermittently. It was determined that reconstructions fail due to a negative table position (-1 value is inserted) in the raw data list file, rather than the actual table position. This error has been found to occur in two scenarios: 1) When the system operator cancels an acquisition a. The error will occur every time a scan is cancelled by the operator. 2) Couch position requests within the software sequence were delayed a. The error occurs intermittently, but has been found to occur more frequently when the gantry s network is heavily loaded with multiple retrospective reconstructions running in parallel. b. The error manifests to the technologist by an error message and Failed status on the Reconstruction Monitor and the error message Result {0} failed to reconstruct" on the Acquisition Workflow window during reconstruction after the patient scan has been completed. In both scenarios, the acquisition data will not be able to be reconstructed and will therefore be unusable.
- 2018-02-21FDA-DeviceArrow International IncClass IIRadial Artery Catheterization Set, 1) NA-04550-1A & 2) NA-04550-X1A, device permits access to the peripheral arterial circulation or to other small vessels.
Needle supplier indicated the presence of loose polypropylene foreign matter above release specification.
- 2018-02-14FDA-DeviceCovidien LLCClass IICovidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm (reinforced), Material CA30L2
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
- 2018-02-14FDA-DeviceCovidien LLCClass IICovidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm, Material CA30L1
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
- 2018-02-14FDA-DeviceCovidien LLCClass IICovidien Emprint Percutaneous Antenna with Thermosphere Technology, 20cm (reinforced), Material CA20L2
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
- 2018-02-14FDA-DeviceCovidien LLCClass IICovidien Emprint Percutaneous Antenna with Thermosphere Technology, 20cm, Material CA20L1
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
- 2018-02-14FDA-DeviceCovidien LLCClass IICovidien Emprint Percutaneous Antenna with Thermosphere Technology, 15cm (reinforced), Material CA15L2
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
- 2018-02-14FDA-DeviceCovidien LLCClass IICovidien Emprint Percutaneous Antenna with Thermosphere Technology, 15cm, Material CA15L1
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
- 2018-02-14FDA-DeviceCyberonics, IncClass IIVNS(R) Therapy Programmer, GMDN: 44077, Model 3000 v1.0.2.2 Intended Use/Indications: United States Indications for Use: Epilepsy : The VNS Therapy System is indicated for use as an adjunctive therapy in reducing the frequency of seizures in patients 4 years of age and older with partial onset seizures that are refractory to antiepileptic medications. Depression : VNS Therapy is indicated for the adjunctive long-term treatment of chronic or recurrent depression for patients 18 years of age
Two Model 3000 Programmers were distributed in error by prior to FDA approval of version 1.0.2.2 software.
- 2018-02-14FDA-DeviceMindray DS USA, Inc. dba Mindray North AmericaClass IITD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00
A damaged circuit board in the battery charger can cause the battery to overcharge and the battery case to swell and separate.
- 2018-02-14FDA-DeviceThe Binding Site Group, Ltd.Class IIRheumatoid Factor (RF) Kit for use on SPAPLUS, Product #LK151.S.A Intended for the quantitative in-vitro measurement of rheumatoid factor in serum using the Binding Site SPAPLUS analyser. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis. This test should be used in conjunction with other laboratory and clinical findings
The low and high controls for the product are returning results above the assigned values.
- 2018-02-14FDA-Device3M Company - Health Care BusinessClass IIACE(TM) BRAND, ULTRA LITE ANKLE brace: (a) Size Selection Medium, Cat #901014, UPC 051131019984 2; (b) Size Selection Large, Cat #901015, UPC 0 51131019985 9
Mold was found on the non-patient contact surfaces of the product and was identified as Aspergillus candidus.
- 2018-02-14FDA-DeviceZimmer Biomet, Inc.Class IIZimmer Natural Nail System Tear Drop Guide Wire 2.4 mm Diameter, 70 cm length, Item Number 47-2490-098-01 Product Usage: The Ball Tip and Tear Drop Guide Wires are used during the initial reaming of the intramedullary canal. They also assists in guiding the nail during implantation. The guide wire incorporates either a ball or tapered tip to aid in removal of the intramedullary reamer if it becomes lodged or fails.
The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed.
- 2018-02-14FDA-DeviceZimmer Biomet, Inc.Class IIZimmer Natural Nail System Tear Drop Guide Wire 3.0 mm Diameter, 70 cm length, Item Number 47-2490-098-00 Product Usage: The Ball Tip and Tear Drop Guide Wires are used during the initial reaming of the intramedullary canal. They also assists in guiding the nail during implantation. The guide wire incorporates either a ball or tapered tip to aid in removal of the intramedullary reamer if it becomes lodged or fails.
The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed.
- 2018-02-14FDA-DeviceZimmer Biomet, Inc.Class IIHumeral Guide Wire Ball Tip 2.4 mm Diameter 70 cm Length, Item Number 47-2255-008-00 Product Usage: The Ball Tip and Tear Drop Guide Wires are used during the initial reaming of the intramedullary canal. They also assists in guiding the nail during implantation. The guide wire incorporates either a ball or tapered tip to aid in removal of the intramedullary reamer if it becomes lodged or fails.
The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed.
- 2018-02-14FDA-DeviceZimmer Biomet, Inc.Class IIM/DN Intramedullary Fixation Humeral Guide Wire - Bullet Tip 2.4 mm Diameter 70 cm Length Single Use Only, Item Number 00-2255-026-00 Product Usage: The Ball Tip and Tear Drop Guide Wires are used during the initial reaming of the intramedullary canal. They also assists in guiding the nail during implantation. The guide wire incorporates either a ball or tapered tip to aid in removal of the intramedullary reamer if it becomes lodged or fails.
The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed.
- 2018-02-14FDA-DeviceZimmer Biomet, Inc.Class IIM/DN Intramedullary Fixation Humeral Guide Wire - Smooth 2.4 mm Diameter 70 cm Length Single Use Only, Item Number 00-2255-025-00 Product Usage: The Ball Tip and Tear Drop Guide Wires are used during the initial reaming of the intramedullary canal. They also assists in guiding the nail during implantation. The guide wire incorporates either a ball or tapered tip to aid in removal of the intramedullary reamer if it becomes lodged or fails.
The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed.
- 2018-02-14FDA-DeviceRobert Busse & Co. Inc. d.b.a. Busse Hospital DisposablesClass IISafety Deluxe Bone Marrow Biopsy/Aspiration Tray, Catalog # 986 Product Usage: Surgical procedure
BD SafetyGlide Needle 22GA 1-1/2 inch may contain the presence of loose polyprophylene foreign matter above release specification.
- 2018-02-14FDA-DeviceRobert Busse & Co. Inc. d.b.a. Busse Hospital DisposablesClass IILumbar Puncture Tray, Catalog # 1862 Product Usage: Surgical procedure
BD SafetyGlide Needle 22GA 1-1/2 inch may contain the presence of loose polyprophylene foreign matter above release specification.
- 2018-02-14FDA-DeviceRobert Busse & Co. Inc. d.b.a. Busse Hospital DisposablesClass IILaceration Tray, Catalog # 1892R1 Product Usage: Surgical procedure
BD SafetyGlide Needle 22GA 1-1/2 inch may contain the presence of loose polyprophylene foreign matter above release specification.
- 2018-02-14FDA-DeviceRobert Busse & Co. Inc. d.b.a. Busse Hospital DisposablesClass IISafety Deluxe Bone Marrow Biopsy/ Aspiration Tray with Biopsy Needle and Aspiration Needle Sterile, Catalog # 985 Product Usage: Surgical procedure
BD SafetyGlide Needle 22GA 1-1/2 inch may contain the presence of loose polyprophylene foreign matter above release specification.
- 2018-02-14FDA-DeviceRobert Busse & Co. Inc. d.b.a. Busse Hospital DisposablesClass IISafety-Deluxe Lumbar Puncture Tray, Catalog # 975 Product Usage: Surgical procedure
BD SafetyGlide Needle 22GA 1-1/2 inch may contain the presence of loose polyprophylene foreign matter above release specification.
- 2018-02-14FDA-DeviceRobert Busse & Co. Inc. d.b.a. Busse Hospital DisposablesClass IISafety-Deluxe Basic Soft Tissue Biopsy Tray without Biopsy Needle, Catalog # 977 Product Usage: Surgical procedure
BD SafetyGlide Needle 22GA 1-1/2 inch may contain the presence of loose polyprophylene foreign matter above release specification.
- 2018-02-14FDA-DeviceRobert Busse & Co. Inc. d.b.a. Busse Hospital DisposablesClass IISafety-Deluxe Lumbar Puncture Tray, Catalog # 974 Product Usage: Surgical procedure
BD SafetyGlide Needle 22GA 1-1/2 inch may contain the presence of loose polyprophylene foreign matter above release specification.
- 2018-02-14FDA-DeviceImplant Direct Sybron Manufacturing, LLCClass IIImplant Direct, INTERACTIVE HEALING COLLAR, PART NUMBER 6530-15, 5.0mmL: 3.0mmD Platform
InterActive Healing Collar, Lot Number 104203, labeled as sterile with distributed prior to being sterilized.
- 2018-02-14FDA-DeviceExactech, Inc.Class IIAlteon HA Femoral Stem, Press-Fit, Extended Offset, Cementless, HA Coated, Collared, Size 5 All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip System
Wrong device description of the label identifying the implant
- 2018-02-14FDA-DeviceMedtronic NeuromodulationClass IIIMedtronic Ascenda(R) Intrathecal Catheter, Models: (a) 8780, UDI: 00643169202122, 00643169783027 (b) 8781, UDI: 00643169202139 (c) 8784, UDI: 00643169202153 The Ascenda Intrathecal Catheter kits (models 8780, 8781 and 8784) contain parts of an infusion system that delivers drugs to the intrathecal space. The implanted system consists of an implantable infusion pump and an Ascenda catheter. The catheter connects to the pump at the catheter port and enters the intrathecal space.
There is a possibility that specific lots of catheters that have recently been distributed may be at risk for an increased potential for kinking at the proximal end where the catheter connects to the drug infusion pump.
- 2018-02-14FDA-DeviceBard Access Systems Inc.Class IIPowerPICC Provena Catheters, as 3 Fr. Single lumen and 4 Fr. Dual Lumen. Packaged in a poly tray with Tyvek lid within a header bag configuration or a breather bag configuration.
Possible leaks on the 3 Fr. Single lumen and 4 Fr. Dual lumen PowerPICC catheters at the luer hub extension leg junction.
- 2018-02-14FDA-DeviceMedtronic Perfusion SystemsClass IIMedtronic Custom Perfusion Systems, indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.
Medtronic custom perfusion tubing packs contain Terumo OPS valves that were recalled due to a lack of flow through the valve.
- 2018-02-14FDA-DeviceCovidien LPClass IICovidien Vistec" X-Ray Detectable Sponges, 7317 Vistec" X-Ray Detectable Sponges Sponge. product Usage: Vistec X-Ray Detectable Sponges are medical absorbent devices intended to be used internally within the body cavity to control bleeding or for applying medication. These sponges contain unique dual-colored elements that combines blue radiopaque strand with white strand, which is visible when sponge is saturated with blood.
Medtronic is voluntarily recalling two production lots of its Covidien Vistec" x-ray detectable sponges de to possible contamination with human blood caused by handling of the product during the manufacturing process and prior to final packaging and Ethylene Oxide sterilization.