Medical device recalls
39229 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2018-02-07FDA-DeviceMicroAire Surgical Instruments, LLCClass IISST TWIST DRILL, STANDARD, 1.6mm Dia. x 127mm, REF 8054-010, STERILE This product line is intended for use in orthopaedic surgery (large and small bone surgeries) to cut bone
One lot was found to contain both the correct twist drill (1.6mm) as well as an incorrect twist drill (2.0mm)
- 2018-02-07FDA-DeviceCarl Zeiss Metrology IncClass IIMetrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal.
Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.
- 2018-01-31FDA-DeviceBaxter Healthcare CorporationClass IIMARS Treatment Kit, Use only with MARS(R) Monitor 1TC, GAMBRO, Product Code 800541
Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.
- 2018-01-31FDA-DeviceSomatex Medical Technologies GmbHClass IITUMARK Q, Length 12 cm, Diameter 18G, GTIN 04250195603081, REF 271500, STERILE
There is a potential risk of puncture to the packaging material which can compromise the devices sterility, making it non-sterile.
- 2018-01-31FDA-DeviceSmith & Nephew, Inc.Class IISmith & Nephew Taylor Spatial Frame Website software, Catalog Number 71070401. A web-based and mobile application software component of a multilateral external fixation system that is intended for fracture fixation, fixation of long bones and for joint fusions and limb lengthening or deformity corrections which involve cutting of the bone.
In rare circumstances, when surgeons use the software to plan the strut adjustment, an inverted image of the frame is generated by the software. The frame will appear inverted in the preview images on the Strut Settings tab.
- 2018-01-31FDA-DeviceSmith & Nephew, Inc.Class IIsmith&nephew, 2.4 MM, 6 HOLE, EVOS, FLEX PLATE, SST Plate, REF CATALOG NUMBER 72442360N, NON-STERILE, Rx only, (01)00885556456446(10), (11)171116 Bone plates and bone screws from the EVOS Mini-Fragment Plating System are intended for use in internal fixation of small bones and small bone fragments and non-load bearing stabilization and reduction of bone fragments in long bones.
One lot of EVOS 2.4MM Flex Plate 6 Hole packaging contained templates.
- 2018-01-31FDA-DeviceInovo, IncClass IIPURE OXYGEN CONCENTRATOR, CHAD Drive, Models CH5000 and CH5000S
Post-market surveillance for Pure Stationary Oxygen Concentrators model CHSOOO and CHSOOOS demonstrate an aggregate complaint/service rate greater than 35% for low oxygen purity (output oxygen below the published specification).
- 2018-01-31FDA-DeviceSmith & Nephew, Inc.Class IIRadial Osteotome Blade, REF 71369314. Used for cutting or preparing bone.
The expiration date provided on the label is incorrect.
- 2018-01-31FDA-DeviceSiemens Medical Solutions USA, Inc.Class IISyngo Imaging version V31 (model # 10014063) is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images, including digital mammography images. It supports the physician in diagnosis and treatment planning.
Siemens is releasing a letter to inform about potential data loss relevant to diagnosis.
- 2018-01-31FDA-DeviceIntuitive Surgical, Inc.Class IIdaVinci X EndoWrist(R) Stapler 45, REF 470298 Product Usage: The Intuitive Surgical EndoWrist Stapler 45, Stapler 45 Reloads, and other Stapler Accessories (including the bladeless obturators) are intended to be used with the da Vinci Surgical System (Model IS4000) for resection, transection, and/or creation of anastomoses in General, Thoracic, Gynecologic and Urologic surgery. The device can be used with the same staple line or tissue buttressing material (natural synthetic).
Specific da Vinci Xi EndoWrist Stapler 45 instruments are potentially impacted by a variation in the manufacturing process. In affected instruments, there is a possibility that the Stapler Release Kit (SRK) Wrench interface feature accessed through release hole 2 is manufactured to an incorrect dimension, which may prevent the SRK Wrench from being able to manually unclamp the instrument.
- 2018-01-31FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIBrilliance iCT - Model 728306 Computed Tomography X-ray systems
Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image volume and other functions.
- 2018-01-31FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIBrilliance iCT SP - Model 728311 Brilliance iCT - Model 728306 Computed Tomography X-ray systems
Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image volume and other functions.
- 2018-01-31FDA-DeviceSorin Group Deutschland GmbHClass IICP5 Centrifugal Pump, made up of the drive unit, the emergency system and the pump control panel. It used with an Electrical Remote-Controlled (ERC) tubing clamp and a bubble sensor.
Possibility that the automatic line closure by the Electronic Remote-Controlled (ERC) clamp after detection of an air bubble could take longer than is intended by design in the CP5 system.
- 2018-01-31FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart FailureClass IIMyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic device that communicates with an implanted heart device
Patients monitored on two (2) or more implanted Medtronic heart devices in the Medtronic CareLink Network may have potential impact on the ability to remotely monitor the patient's heart devices. Potential impacts could lead to missed CareAlert notifications or device reports.
- 2018-01-31FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart FailureClass IIMyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The monitor is an external electronic device that interfaces with Medtronic implanted heart devices and a telecommunications connection to transmit stored implanted heart device data to the physician or clinician.
Patients monitored on two (2) or more implanted Medtronic heart devices in the Medtronic CareLink Network may have potential impact on the ability to remotely monitor the patient's heart devices. Potential impacts could lead to missed CareAlert notifications or device reports.
- 2018-01-31FDA-DeviceThe Anspach Effort, Inc.Class IIOSCILLATING SAW ATTACHMENT Surgical instrument motors and accessories/attachments are intended for use during surgical procedures to cut hard tissue or bone and soft tissue.
This voluntary recall was initiated because a veterinary unit was physically shipped in error to a non-veterinary customer account.
- 2018-01-31FDA-DeviceSmith & Nephew, Inc.Class IIsmith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating system, 2.0MM STOUT STRAIGHT PLATE, 8 HOLE VLP(R) MIN-MOD, REF 74442030, QTY: (1), STERILE R Bone plates and bone screws from the Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System (VLP MINI-MOD) are intended for use in internal fixation of small bones and small bone fragments and non-load bearing stabilization and reduction of bone fragments in long bones.
Devices, incorrectly labeled as a VLP MINI-MOD 2.0MM STRAIGHT PLATE 8 HOLE, contained a VLP MINI-MOD 2.0MM STOUT PLATE 8 HOLE and vice versa
- 2018-01-31FDA-DeviceSmith & Nephew, Inc.Class IIsmith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating system, 2.0MM STRAIGHT PLATE, 8 HOLE VLP(R) MIN-MOD, REF 74442024, QTY: (1), STERILE R Bone plates and bone screws from the Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System (VLP MINI-MOD) are intended for use in internal fixation of small bones and small bone fragments and non-load bearing stabilization and reduction of bone fragments in long bones.
Devices, incorrectly labeled as a VLP MINI-MOD 2.0MM STRAIGHT PLATE 8 HOLE, contained a VLP MINI-MOD 2.0MM STOUT PLATE 8 HOLE and vice versa
- 2018-01-31FDA-DeviceMeasurement Specialties IncClass IIMeasurement Specialties Reusable Temperature Probe Autoclavable. Sold as Esophageal/Rectal (Adult), Esophageal/Rectal (Pediatric), and Skin (Adult) Intended for use by a licensed physician to continuously monitor human esophageal, rectal, or skin temperature. They are designed to be used with any patient monitoring system designed to accept 400 Series temperature probes.
The reusable temperature probes are provided with instructions for sterilization that have not been proven through a validation study.
- 2018-01-31FDA-DeviceBecton Dickinson & Co.Class IIBD MAX" CT/GC/TV. Catalog Number(s): 443824,442970,443590 (RUO).for collection and transport of vaginal and endocervical swab specimens, in addition to the preservation of urine from males and females.
BD recently notified customers in July 2017 regarding an increased rate of unresolved (UNR) and indeterminate (IND) results for the BD MAXTM Vaginal Panel, experienced by some customers. A non-reportable result may lead to repeat of the test and/or re-collection of a specimen.
- 2018-01-31FDA-DeviceBecton Dickinson & Co.Class IIBD MAX" UVE Specimen Collection Kit. Catalog Number(s): 443376 for collection and transport of vaginal and endocervical swab specimens, in addition to the preservation of urine from males and females.
BD recently notified customers in July 2017 regarding an increased rate of unresolved (UNR) and indeterminate (IND) results for the BD MAXTM Vaginal Panel, experienced by some customers. A non-reportable result may lead to repeat of the test and/or re-collection of a specimen.
- 2018-01-31FDA-DeviceBecton Dickinson & Co.Class IIBD MAXTM Vaginal Panel. Catalog Number(s): 443712 443710 443711 (RUO)for collection and transport of vaginal and endocervical swab specimens, in addition to the preservation of urine from males and females.
BD recently notified customers in July 2017 regarding an increased rate of unresolved (UNR) and indeterminate (IND) results for the BD MAXTM Vaginal Panel, experienced by some customers. A non-reportable result may lead to repeat of the test and/or re-collection of a specimen.
- 2018-01-31FDA-DeviceVyaire MedicalClass IAdult AirLife dual-limb, dual-heat circuit (Model AH280) with AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.
Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.
- 2018-01-31FDA-DeviceVyaire MedicalClass IAdult Air-Life single-limb circuit (Model AH202) with AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.
Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.
- 2018-01-31FDA-DeviceVyaire MedicalClass IInfant AirLife Dual-,limb, dual-heat, high-flow circuit (Model AH265) with AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.
Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.
- 2018-01-31FDA-DeviceVyaire MedicalClass IInfant AirLife single-limb, continuous-flow circuit (Model AH132) containing AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.
Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.
- 2018-01-24FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIIngenuity Core 128, (Code 728323) Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.
- 2018-01-24FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIIngenuity Core, (Code 728321) Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.
- 2018-01-24FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIBrilliance 64, (Code 728231) Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.
- 2018-01-24FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIBrilliance iCT Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.