- 2017-11-08FDA-DeviceAlcon Research, LTD.Class IIAlcon Custom Pak
Alcon is initiating a medical device recall, after receiving notification from a supplier of a recall for x-ray detectable sponges. The one product referenced in the supplier recall notification, is used within Alcon Custom Paks.
- 2017-11-08FDA-DeviceBios S.r.l.Class IITHERMI250 Radiofrequency System. Intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions.
In some systems, the unit continues to emit RF energy after the PAUSE button is pressed.
- 2017-11-01FDA-DeviceNucletron BVClass IIOncentra Brachy 4.5 radiation therapy software The firm name on the label is Nucletron B.V. Oncentra is a radiation therapy planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer. The treatment plans provide estimates of dose distributions expected during the proposed treatment and may be used to administer treatments after review and approval by qualified medical personnel.
Incorrect source step size may occur in the software plans
- 2017-11-01FDA-DeviceEdap Technomed Inc.Class IIAblatherm(R) Integrated Imaging, High Intensity Ultrasound System For Prostate Tissue Ablation Transrectal high intensity focused ultrasound (HIFU) ablation of prostate tissue
The US FDA has requested the optional energy treatment settings, "medium" and "low" (Software Protocols 02-Medium and 03-Low) be removed from all Ablatherm Integrated Imagining devices in the U.S. until supporting clinical data can be submitted and evaluated by FDA.
- 2017-11-01FDA-DeviceW L Gore & Associates, Inc.Class IIGORE TAG, THORACIC ENDOPROSTHESIS, CONFORMABLE. The GORE TAG Thoracic Endoprosthesis is intended for endovascular repair of all lesions of the descending thoracic aorta.
Incomplete and/or partial deployment of the CTAG Device during the endovascular procedure.
- 2017-11-01FDA-DeviceSiemens Medical Solutions USA, Inc.Class IITorque wrench used on the bolts for the Symbia T2 model 10275008, Symbia Intevo Excel model 10764801, Symbia Intevo 16 model 10764804, Symbia Evo model 10910500, and Symbia Evo Excel model 10910501 devices.
The torque wrench used to tighten system bolts during installation was found to be out of tolerance
- 2017-11-01FDA-DeviceBeckman Coulter Inc.Class IIPower Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.
Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard can initiate a retrieval of a sample tube during the rack loading process, which should not occur. This issue and associated complaint were discovered and filed internally.
- 2017-11-01FDA-DeviceSysmex America, Inc.Class IISysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers
There is potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%, #) results obtained using the lots listed above may exhibit lower than actual PLT-F and IPF values. The issue may be recognized by a large discrepancy between the impedance platelet (PLT-I) result and the PLT-F result. In most cases, the falsely decreased PLT-F results display a PLT Abn Scattergram message with an asterisk beside the result or an Action Message: Difference between PLT and PLT- F. Check the results. These are indicators of unreliable data. In rare instances, platelet flags may be absent. The PLT-I results are unaffected.
- 2017-11-01FDA-DeviceBard Peripheral Vascular IncClass IIBARD MAX-CORE Disposable Core Biopsy Instrument, Product Codes - MC1410, 14 g x 10 cm; MC1616, 16 g x 16 cm; MC1816, 18 g x 16 cm; MC1820, 18 g x 20 cm; MC1825, 18 g x 25 cm. The core needle biopsy device is intended for use in obtaining biopsies from soft tissues.
During the review of May 2017 complaint data, a significant increase in MAX-CORE Disposable Core Biopsy Instrument complaints was observed for Failure to Prime, Failure to Fire, Failure to Obtain Sample, and Self-Activation.
- 2017-11-01FDA-DeviceMicroPort Orthopedics Inc.Class IIMicroPort Orthopedics SCHANZ SCREW SELF DRILL, SIZE 4mm, REF 20070057, STERILE R Product Usage: MPO Total Hip Systems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2. inflammatory degenerative joint disease such as rheumat
The item is being recalled due to an inadequate pouch seal. This defect has the potential to lead to a loss of sterile pouch integrity, which could impact the products sterility assurance and may lead to serious adverse events such as infections, both localized and systemic.
- 2017-11-01FDA-DeviceEncore Medical, LpClass IIEMPOWR PS KNEE(TM) e+, REF 343-19-707, SIZE 7, 19mm, QTY 01, GREY, STERILE Product Usage The EMPOWR PS Inserts are manufactured from ultra high molecular weight polyethylene (UHMWPE) infused with pure liquid pharmaceutical grade alpha-tocopheral and then cross-linked to conform to ASTM F2695. This insert is posterior stabilized and is intended to more closely match the kinematics of the knee, allowing some rotation along the medial condyle and increased congruency along the lateral condyle.
It was reported that an implant labeled as 343-13-708 was packaged with implant 343-19-707.
- 2017-11-01FDA-DeviceEncore Medical, LpClass IIEMPOWR PS KNEE(TM) e+, REF 343-13-708, SIZE 8, 13mm, QTY 01, SKY BLUE, STERILE Product Usage The EMPOWR PS Inserts are manufactured from ultra high molecular weight polyethylene (UHMWPE) infused with pure liquid pharmaceutical grade alpha-tocopheral and then cross-linked to conform to ASTM F2695. This insert is posterior stabilized and is intended to more closely match the kinematics of the knee, allowing some rotation along the medial condyle and increased congruency along the lateral cond
It was reported that an implant labeled as 343-13-708 was packaged with implant 343-19-707.
- 2017-11-01FDA-DeviceOriGen Biomedical, Inc.Class IThe OriGen Reinforced Dual Lumen catheter (RDLC) is a duel lumen catheter with a wire reinforcement layer on the tip of the device. A thin sheath is used to cover and secure the wire layer and is heat shrunk into a permanent position.
Two users have reported to OriGen that a VV28F OriGen Reinforced Dual Lumen ECMO Catheter has had a separation of the clear extension tube from the hub connection. These adverse events resulted in patient injury (blood loss).
- 2017-10-25FDA-DeviceGenesys Orthopedics Systems, LLCClass IIGenesys spine, Binary Screwdriver, GCP221, component of the Binary Anterior Cervical Plating System The Genesys Spine Binary Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine. It is to be used in skeletally mature patients as an adjunct to fusion of the cervical spine (C2 to C7). The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusion in patients with degenerative disc di
Unanticipated technique for a cervical screw insertion caused the tip of the threaded inner shaft to break off.
- 2017-10-25FDA-DeviceTeleflex MedicalClass IIPilling Knife Handles: a) REF 352950 b) REF 352953 c) REF 352951 d) REF 352957 e) REF 352952 : Reusable surgical knife handles, which are intended for use with blades which are inserted into a slot in the handle.
The knife handles slot depth is out of specification, so blades may not fit on the handles properly.
- 2017-10-25FDA-DeviceSt Jude Medical Inc.Class IIPM3544 QUADRA RELIEVE and PM3562 QUADRA ALLURE MP These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
- 2017-10-25FDA-DeviceSt Jude Medical Inc.Class IIPM 1260 ASSURITY + These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders.
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
- 2017-10-25FDA-DeviceSt Jude Medical Inc.Class IIPM 1250 and PM 2250 ZENEX; PM1282 and PM2282 ZENEX MRI These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders.
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
- 2017-10-25FDA-DeviceSt Jude Medical Inc.Class IIPM1226 ACCENT ST MRI SR RF and PM2222 ACCENT ST These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders.
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
- 2017-10-25FDA-DeviceSt Jude Medical Inc.Class IIPM1214 NUANCE SR RF; PM1230 and PM2230 NUANCE MRI; PM2214 NUANCE DR RF These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders.
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
- 2017-10-25FDA-DeviceSt Jude Medical Inc.Class IIMerlin@Home Software Model EX2000 v8.2.2 for Merlin@Home Transmitter Models EX1150, EX1150W, EX1100, and EX1100W. The Merlin@home Transmitters are a component of St. Jude Medicals (SJM) Remote Care System. These Merlin@home transmitters are used to retrieve information from the SJM implantable medical device (IMD) using RF telemetry. The data is then sent back to the Merlin.net Patient Care Network through a built-in modem or, alternatively, through a USB cellular adapter or a USB based Broad
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
- 2017-10-25FDA-DeviceSt Jude Medical Inc.Class IIMerlin PCS 3650 Programmer with Model 3330 Software Version 23.1.1 The Merlin Patient Care System (PCS) Model 3650 Programmer is a portable, dedicated programming system designed to interrogate program display data and test St. Jude Medical implantable devices.
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
- 2017-10-25FDA-DeviceSt Jude Medical Inc.Class IIAssurity MRI Model Numbers: PM1272, PM2272 Implantation of a single-chamber pulse generator, dual-chamber pulse generator, or CRT-P is indicated in one or more of the following permanent conditions: Syncope, Presyncope, Fatigue, Disorientation, Or any combination of those symptoms. Implantation of a CRT-P is indicated for: Maintaining synchrony of the left and right ventricles in patients who have undergone an AV nodal ablation for chronic atrial fibrillation and have NYHA Class II or III heart
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
- 2017-10-25FDA-DeviceSt Jude Medical Inc.Class IIAssurity Model Numbers: PM1240, PM2240; Assurity + Model Number: PM2260; Accent Model Numbers: PM1210, PM2210, PM1222, PM2212; Accent MRI Model Numbers: PM1224, PM2218, PM2224 Implantation of a single-chamber pulse generator, dual-chamber pulse generator, or CRT-P is indicated in one or more of the following permanent conditions: Syncope, Presyncope, Fatigue, Disorientation, Or any combination of those symptoms. Implantation of a CRT-P is indicated for: Maintaining synchrony of the left and rig
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
- 2017-10-25FDA-DeviceSt Jude Medical Inc.Class IIAnthem Model Numbers: PM3210, PM3212; Allure/Relieve Model Numbers: PM3222, PM3242, PM3262, PM3542, PM3224, PM3244 Implantation of a single-chamber pulse generator, dual-chamber pulse generator, or CRT-P is indicated in one or more of the following permanent conditions: Syncope, Presyncope, Fatigue, Disorientation, Or any combination of those symptoms. Implantation of a CRT-P is indicated for: Maintaining synchrony of the left and right ventricles in patients who have undergone an AV nodal abla
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
- 2017-10-25FDA-DeviceCepheidClass IIXpert EV Assay, Catalog GXEV-100N-10. The Cepheid Xpert EV assay is a reverse transcription polymerase chain reaction (RT-PCR) using the GeneXpert Dx System for the presumptive qualitative detection of enterovirus (EV) RNA in cerebrospinal fluid (CSF) specimens from individuals with signs and symptoms of meningitis.
Customers have reported a higher than expected number of invalid test results when using the recalled product per the package insert.
- 2017-10-25FDA-DeviceHologic, IncClass IIPanther(R) System, Catalog Number 902615, For Use with Hologic Assays, For In Vitro Diagnostic Use
An incorrect incubator was installed on one Panther System which potentially led to incorrect Aptima Combo 2(R) assay (AC2) results.
- 2017-10-25FDA-DeviceCareFusion 303, Inc.Class IIAlaris Pump Module model 8100 manufactured between November 2011 and March 2012; Alaris Pump Module serviced with LVP Mechanism Sub Assembly (P/N) 10942012 between November 2011 and March 2012; Alaris Pump module Bezel Kit Assembly (P/N) 10964559) shipped between November 2011 and March 2012. The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using the IV administration sets for continuous or
The recalling firm has received reports of increased or decreased flows that have occurred in certain pumps.
- 2017-10-25FDA-DeviceMedtronic Navigation, Inc.Class IIVCLAS (Visualase Cooled Laser Ablation System) Part Number: 9735559 Visualase Cooled Laser Applicator System 3mm Tip; Part Number: 9735560 Visualase Cooled Laser Applicator System 10mm Tip; Part Number: 9735561 Visualase Cooled Laser Applicator System 15mm Tip
The VCLAS 3mm, 10mm, and 15mm tip devices were not appropriately tested for saline leakage during the manufacturing process.
- 2017-10-25FDA-DeviceSt. Jude Medical, Inc.Class IIProclaim DRG Implantable Pulse Generator Model 3664
Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations.