Medical device recalls
39229 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2017-10-25FDA-DeviceSt. Jude Medical, Inc.Class IIInfinity 5 Implantable Pulse Generator and Infinity 7 Implantable Pulse Generator Models 6660, 6661, 6662, 6663
Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations.
- 2017-10-25FDA-DeviceSt. Jude Medical, Inc.Class IIProclaim 5 Elite Implantable Pulse Generator and Proclaim 7 Elite Implantable Pulse Generator Models 3660, 3661, 3662, 3663, 3665, 3667
Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations.
- 2017-10-25FDA-DevicePhilips Electronics North America CorporationClass IIPhilips Healthcare Brilliance iCT SP X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
- 2017-10-25FDA-DevicePhilips Electronics North America CorporationClass IIPhilips Healthcare Brilliance iCT X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
- 2017-10-25FDA-DevicePhilips Electronics North America CorporationClass IIPhilips Healthcare Ingenuity Core128 X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
- 2017-10-25FDA-DevicePhilips Electronics North America CorporationClass IIPhilips Healthcare Ingenuity CT X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
- 2017-10-25FDA-DevicePhilips Electronics North America CorporationClass IIPhilips Healthcare Ingenuity Core X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
- 2017-10-25FDA-DevicePhilips Electronics North America CorporationClass IIPhilips Healthcare Brilliance 64 System, X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
- 2017-10-25FDA-DeviceToshiba American Medical Systems IncClass IIToshiba Medical Kalare Fluoroscopic X-Ray System (DREX-KL80) Kalare is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations. It is intended for use by qualified/trained doctors or technologists on both adult and pediatric subjects taking radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities, and other body parts. Applications ca
During an examination images were displayed on the live monitor,but the images were not displayed on the system monitor nor were they saved to the hard disk.
- 2017-10-18FDA-DeviceHandicare ABClass IICarina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009
A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.
- 2017-10-18FDA-DeviceHandicare Usa IncClass IIPRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered
During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.
- 2017-10-18FDA-DeviceAbbott Laboratories, IncClass IIARCHITECT c4000 Cuvette Segment, List number 02P75-01. The firm name on the label is Abbott Laboratories, Abbott Park, IL. Cuvette segments are a component to the ARCHITECT c4000 system. Cuvette segments are racks that sit in the reaction carousel and hold cuvettes. Each cuvette segment holds 11 cuvettes. With 9 segments, the reaction carousel holds 99 cuvettes. The Abbott ARCHITECT System is intended for In Vitro diagnostic use only. The system is designed to perform automated chemistry te
There is a potential to generate falsely-depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000 instrument.
- 2017-10-18FDA-DeviceSt Jude Medical Inc.Class IUnify, Model Numbers: CD3231-40 and CD3231-40Q; Unify Quadra, Model Numbers: CD3249-40 and CD3249-40Q; Unify Assura, Model Numbers: CD3257-40, CD3257-40Q, CD3357-40C, CD3357-40Q; Quadra Assura, Model Numbers: CD3265-40, CD3265-40Q, CD3365-40C, CD3365-40Q; Quadra Assura MP, Model Numbers: CD3269-40, CD3269-40Q, CD3369-40C, CD3369-40Q Unify and Unify Assura cardiac resynchronization therapy defibrillators (CRTDs) are intended to provide ventricular anti-tachycardia pacing and ventricular defibril
The device may exhibit premature battery depletion.
- 2017-10-18FDA-DeviceSt Jude Medical Inc.Class IFortify VR - Model Numbers: CD1231-40 and CD1231-40Q; Fortify ST VR, Model Numbers: CD1241-40 and CD1241-40Q; Fortify Assura VR, Model Numbers: CD1257-40, CD1257-40Q, CD1357-40C, CD1357-40Q; Fortify Assura ST VR, Model Numbers: CD1263-40, CD1263-40Q, CD1363-40C, CD1363-40Q; Fortify DR, Model Numbers: CD2231-40 and CD2231-40Q; Fortify ST DR, Model Numbers: CD2241-40 and CD2241-40Q; Fortify Assura DR, Model Numbers: CD2257-40, CD2257-40Q, CD2357-40C, CD2357-40Q; Fortify Assura ST DR, Model Numbers
The device may exhibit premature battery depletion.
- 2017-10-11FDA-DeviceThe Metrix CompanyClass IISECURE(R) 100mL EMPTY VINYL CONTAINER With 2 Ports - (5 Boxes of 40), REF 58719, STERILE R,
The hot stamp label on the bag incorrectly states the material of the vinyl (bag) container. The bags are correctly made with PVC material, as ordered and as labeled on the inner and outer boxes.
- 2017-10-11FDA-DeviceFirefly Medical, Inc.Class IIIVEA Product Part Number 16100 Model 500A
The dowel pins (P/N 16082) that fit in the left and right leg pivot castings (P/Ns 16049/16056) can rotate and gradually maneuver out of the leg casting after several hundred deployment and folding cycles (designed for easy storage) of the IVEA unit.
- 2017-10-11FDA-DeviceBomimedClass IIBOMImed Disposable Fiber Optic Medium Laryngoscope Handle. Each handle is packaged in an individual peel pouch, 20 handles per box. The rigid laryngoscope handle is a device used along with a blade to examine and visualize a patient's airway and aid placement of a tracheal tube.
Potential for intermittent lighting (flickering) during use.
- 2017-10-11FDA-DeviceRadiometer America IncClass IIAutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality control (QC) ampoules are specifically designed for Radiometer America s blood gas analyzers
Radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a result close to or just outside of the low end of the control range. (4.92 5.85 kPa or 36.9 46.9 mmHg) " pO2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kPa (104-107 mmHg)
- 2017-10-11FDA-DeviceImplant Direct Sybron Manufacturing, LLCClass IIImplant Direct Swish Duo-Grip Healing Collar, Length: 3mm, Platform: 4.8 mm, Rx only, Sterile Dental Implant component.
The Swish Duo-Grip Healing Collar part number 9048-13H (4.8mm) package contained a 9037-13H (3.7 mm) component.
- 2017-10-11FDA-DevicePermobil, AbClass IIElectric Wheelchair bases: a) C300 b) C350 c) C400 d) C500
There is a potential failure of the top plate assembly, which is the component that connects the seating system of the wheelchair to the base.
- 2017-10-11FDA-DevicePermobil, AbClass IIElectric Wheelchairs: a) C300 CORPUS II b) C350 CORPUS II c) C400 CORPUS II d) C400 CORPUS II LR e) C500 CORPUS II f) C500 CORPUS II LR
There is a potential failure of the top plate assembly, which is the component that connects the seating system of the wheelchair to the base.
- 2017-10-11FDA-DeviceThe Binding Site Group, Ltd.Class IICheck The Binding Site antitrypsin test kit lot #400900
The high control for the kit is recovering low and may be out of range for assigned values.
- 2017-10-11FDA-DeviceCarl Zeiss Meditec AGClass IICheck Carl Zeiss INTRABEAM calibration certificate
The device had an incorrect printed calibration value on the calibration certificate.
- 2017-10-04FDA-DeviceNOVADAQ TECHNOLOGIES INC.Class IINOVADAQ SPY-PHI Drape, 7" x 104" (18 cm x 265 cm), Ref. HH 2020 (box of 20 HH2000 drapes), Sterile, Rx.. Manufactured for: Novadaq Technologies Inc., Burnaby, BC, Canada. The SPY-PHI Drape is intended to cover the SPY-PHI System to maintain a sterile field, and as an aid in the clean-up of equipment after surgery or other medical procedures. The product is not intended to be used as patient drapes and they have no patient contact.
The sterile drape material may contain a pinhole compromising its integrity resulting in a breach of the sterile barrier.
- 2017-10-04FDA-DeviceNOVADAQ TECHNOLOGIES INC.Class IINOVADAQ SPY-PHI PAQ Intended for use with the SPY-PHI System, Product Code HH9006, containing 6 Single-Procedure Kits (Product Code HH9001), each containing the sterile drape (HH2000), as well as single-use drug components, Manufactured for and Distributed by Novadaq Technologies Inc., Burnaby, BC, Canada, Assembled by Carestream Health, Inc., Rochester, NY. The SPY-PHI Drape is intended to cover the SPY-PHI System to maintain a sterile field, and as an aid in the clean-up of equipment after
The sterile drape material may contain a pinhole compromising its integrity resulting in a breach of the sterile barrier.
- 2017-10-04FDA-DeviceNOVADAQ TECHNOLOGIES INC.Class IINOVADAQ SPY-PHI Drape, 7" x 104" (18 cm x 265 cm), Ref. HH2000 (single drape), Sterile, Rx. Manufactured for: Novadaq Technologies Inc., Burnaby, BC, Canada. The SPY-PHI Drape is intended to cover the SPY-PHI System to maintain a sterile field, and as an aid in the clean-up of equipment after surgery or other medical procedures. The product is not intended to be used as patient drapes and they have no patient contact.
The sterile drape material may contain a pinhole compromising its integrity resulting in a breach of the sterile barrier.
- 2017-10-04FDA-DeviceSorin Group Italia SRL - CRFClass IIPlatinium CRT-D 1741, Model Number TDF037U, UDI/GTIN 08031527015484, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of consc
There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.
- 2017-10-04FDA-DeviceSorin Group Italia SRL - CRFClass IIPlatinium DR DF4 1540, Model Number TDF036U, UDI/GTIN 08031527015477, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of cons
There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac ResynchronizationTherapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.
- 2017-10-04FDA-DeviceSorin Group Italia SRL - CRFClass IIPlatinium VR DF4 1240, Model Number TDF035U, UDI/GTIN 08031527015460, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of cons
There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.
- 2017-10-04FDA-DeviceSorin Group Italia SRL - CRFClass IIPlatinium CRT DF1 1711, Model Number TDF033U, UDI/GTIN 08031527015538, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of con
There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.