Medical device recalls
39229 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2017-09-27FDA-DeviceMedtronic Navigation, Inc.Class IISpine Referencing Instrumentation, Spinous Process Clamp, Short When used with a Medtronic StealthStation Navigation System, the Spine Referencing fixation devices are intended to provide rigid attachment between patient and patient reference frame for the duration of the surgery. The devices are intended to be reusable. The spine referencing clamps provide bony fixation through two styles of clamps; a single spinous process clamp and a double spinous process clamp.
Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.
- 2017-09-27FDA-DeviceMedtronic Navigation, Inc.Class IISpine Referencing Instrumentation, Spinous Process Clamp, Tall When used with a Medtronic StealthStation Navigation System, the Spine Referencing fixation devices are intended to provide rigid attachment between patient and patient reference frame for the duration of the surgery. The devices are intended to be reusable. The spine referencing clamps provide bony fixation through two styles of clamps; a single spinous process clamp and a double spinous process clamp.
Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.
- 2017-09-27FDA-DeviceKamiya Biomedical Company, LLCClass IIK-Assay(R) Multi-Analyte Calibrator, REF KAI--016C, KAMIYA BIOMEDICAL COMPANY, Seattle, WA, IVD The K-ASSAY(R) Multi-Analyte Calibrator is intended to be used for the calibration of the K-ASSAY(R) Alpha-1 Acid Glycoprotein, Alpha-1 Anti-Trypsin, Complement C3, Complement C4, Haptoglobin, lgA, lgG, lgM, and Transferrin immunoturbidimetric assays. For in vitro diagnostic use
The AlAT calibrator values in the package insert for the Multi-Analyte Calibrator, lot D755, lots N653 and lot N654 were not correct.
- 2017-09-27FDA-DeviceMedtronic Vascular Galway DBA Medtronic IrelandClass IISolarice Rapid Exchange Balloon Dilatation Catheter, a Percutaneous Transluminal Coronary Angioplasty (PTCA) Rapid Exchange System.
Stylette removal difficulties on the Euphora and Solarice products.
- 2017-09-27FDA-DeviceMedtronic Vascular Galway DBA Medtronic IrelandClass IIEuphora Rapid Exchange Balloon Dilatation Catheter, a Percutaneous Transluminal Coronary Angioplasty (PTCA) Rapid Exchange System.
Stylette removal difficulties on the Euphora and Solarice products.
- 2017-09-27FDA-DeviceBeckman Coulter Inc.Class IIAQUIOS CL Flow Cytometer System The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping.
Beckman Coulter has identified that an incorrect setting using a proprietary software configuration tool can result in the Sample Prep Probe of the AQUIOS CL drawing aspirant from the wrong tube in a cassette which manifests in incorrect values being reported out of the laboratory for some patient samples.
- 2017-09-27FDA-DeviceBiomerieux IncClass IIVITEK(R) 2 GP, REF 21342, IVD, UDI 03573026131920, bioMerieux, Inc. VITEK(R) 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK¿ 2 utilizes growth based biochemical patterns to determine identification. The VITEK¿ 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consist
Customer reports have indicated occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC(R) 700327(TM) in association with VITEK(R) 2 GP ID Lot 2420240403 and 2420303403, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. The expected O129R reaction for ATCC¿ 700327(TM) strain is positive.
- 2017-09-20FDA-DeviceBaxter Healthcare CorpClass IIBaxter Amia Automated Peritoneal Dialysis Set with Cassette, REF 5C5479
The firm received increased customer complaints for Missing Red Line, Patient Slow Flow, Solution Slow Flow, and Inadequate Drain alerts on certain lots of the AMIA Automated Peritoneal Dialysis Set with Cassette.
- 2017-09-20FDA-DeviceBoston Scientific CorporationClass IIFFR Link-FFR Signal Processing Module, Material Number H7495551000 It is intended to condition physiological signals from measuring devices (BSC Pressure Guidewire or an external pressure transducer), transmit and receive via radiofrequency, and recondition the signals so they can be displayed on and/or recorded in a receiving device (iLab POLARIS Multi-Modality Guidance System or other monitoring device). The physiological signals can also be distributed by cable
The device history record (DHR) was missing its test documentation for final HIPOT (high potential) electrical testing.
- 2017-09-20FDA-DeviceMako Surgical CorporationClass IITotal Hip Application (THA) Product Usage: The Robotic Arm Interactive Orthopedic System (RIO) is intended to assist the surgeon in providing software define spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
Software discrepancy of not showing all the EE constants, when the screen is filled.
- 2017-09-20FDA-DeviceMako Surgical CorporationClass IIMakoplasty Partial Knee Application (PKA) Product Usage: The Robotic Arm Interactive Orthopedic System (RIO) is intended to assist the surgeon in providing software define spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
Software discrepancy of not showing all the EE constants, when the screen is filled.
- 2017-09-20FDA-DeviceFujifilm Medical Systems U.S.A., Inc.Class IIFujifilm Endoscope Models ED-250XL5 ED-250XT5 ED-450XT5 ED-450XL5 Product Usage: This product is a medical endoscope for the duodenum and upper GI tract. It is intended for observation, diagnosis, and endoscopic treatment of the esophagus, stomach and duodenum. This product is not used for any other purposes.
Fujifilm is conducting a corrective action due to an FDA inspection.
- 2017-09-20FDA-DeviceFujifilm Medical Systems U.S.A., Inc.Class IIFujifilm Endoscope Model ED-530XT Product Usage: This product is a medical endoscope for the duodenum and upper GI tract. It is intended for observation, diagnosis, and endoscopic treatment of the esophagus, stomach and duodenum. This product is not used for any other purposes.
Fujifilm is reporting a corrective action due to an FDA inspection.
- 2017-09-20FDA-DeviceOrthofix, IncClass IIFirebird NXG Spinal Fixation System Decorticating Planer, Catalog #36-1334, non-sterile, Rx. .A Decorticating Planer is used to decorticate bone surrounding the spherical head of an implanted bone screw in order to provide clearance to allow for proper seating of the modular body, as well as to provide full range of modular body angulation. It is placed axially over the spherical head of the bone screw and rotated clockwise and counterclockwise to achieve the needed decortication
The plastic core of the Decorticating Planer was found to be missing, which could result in scoring of the bone screw neck if used at or beyond a 23 degree angle.
- 2017-09-20FDA-DeviceLeica Biosystems Richmond Inc.Class IICell Marque, SOX-10 (Polyclonal), 7 mL, REF PA0813, IVD, This antibody is intended for in vitro diagnostic (IVD) use. SOX-10 Rabbit Polyclonal Primary Antibody is intended for laboratory use in the detection of the SOX-10 protein in formalin-fixed, paraffin-embedded tissue stained in qualitative immunohistochemistry (IHC) testing..
A delay in transit may have contributed to nonconformity as the product was stored outside of defined shipping and handling temperature requirements.
- 2017-09-20FDA-DeviceLeica Biosystems Richmond Inc.Class IICell Marque, PAX8 (MRQ-50 Mab.), 7 mL, REF PA0808, IVD, This antibody is intended for in vitro diagnostic (IVD) use. PAX-8 (MRQ-50) Mouse Monoclonal Primary Antibody is intended for laboratory use in the detection of the PAX-8 protein in formalin-fixed, paraffin-embedded tissue stained in qualitative immunohistochemistry (IHC) testing
A delay in transit may have contributed to nonconformity as the product was stored outside of defined shipping and handling temperature requirements.
- 2017-09-20FDA-DeviceLeica Biosystems Richmond Inc.Class IICell Marque, Napsin A (MRQ-60 Mab.), 7 mL, REF PA0805, IVD, The antibody is intended for in vitro diagnostic (IVD) use. Napsin A (MRQ-60) Mouse Monoclonal Primary Antibody is intended for laboratory use in the detection of the Napsin A protein in formalin-fixed, paraffin-embedded tissue stained in qualitative immunohistochemistry (IHC) testing.
A delay in transit may have contributed to nonconformity as the product was stored outside of defined shipping and handling temperature requirements.
- 2017-09-20FDA-DeviceLeica Biosystems Richmond Inc.Class IICell Marque, MSH6 (44 Mab.), 7 mL, REF PA0804, IVD, The antibody is intended for in vitro diagnostic (IVD) use. MSH6 is a mismatch repair gene which is deficient in a high proportion of patients with microsatellite instability (MSI-H). This finding is associated with the autosomal dominant condition known as Hereditary Non-Polyposis Colon Cancer (HNPCC).
A delay in transit may have contributed to nonconformity as the product was stored outside of defined shipping and handling temperature requirements.
- 2017-09-20FDA-DeviceLeica Biosystems Richmond Inc.Class IICell Marque, Microphthalmia Transcription Factor (MiTF) (C5/D5 Mab.), 7 mL, REF PA0803, IVD, The antibody is intended for in vitro diagnostic (IVD) use. The Microphthalmia Transcription Factor (MiTF) (C5/D5) antibody is intended for qualified laboratories to qualitatively identify by light microscopy the presence of associated antigens in sections of formalin-fixed, paraffin-embedded tissue sections using IHC test methods.
A delay in transit may have contributed to nonconformity as the product was stored outside of defined shipping and handling temperature requirements.
- 2017-09-20FDA-DeviceLeica Biosystems Richmond Inc.Class IICell Marque, Glypican-3 (1G12 Mab), 7 mL, REF PA0800, IVD, The antibody is intended for in vitro diagnostic (IVD) use. Glypican-3 (1G12) Mouse Monoclonal Primary Antibody is intended for laboratory use in the detection of the Glypican-3 protein in formalin-fixed, paraffin-embedded tissue stained in qualitative immunohistochemistry (IHC) testing.
A delay in transit may have contributed to nonconformity as the product was stored outside of defined shipping and handling temperature requirements.
- 2017-09-20FDA-DeviceLeica Biosystems Richmond Inc.Class IICell Marque, Podoplanin (D2-40 Mab), 7 mL, REF PA0796, IVD, The antibody is intended for in vitro diagnostic (IVD) use. The Podoplanin (D2-40) antibody is intended for qualified laboratories to qualitatively identify by light microscopy the presence of associated antigens in sections of formalin-fixed, paraffin-embedded tissue sections using IHC test methods
A delay in transit may have contributed to nonconformity as the product was stored outside of defined shipping and handling temperature requirements.
- 2017-09-20FDA-DevicebioMerieux, Inc.Class IIeMAG System, Ref 418591 It is an In Vitro Diagnostic Medical device intended for the automated isolation (purification and concentration) of total nucleic acids (RNA/DNA) from biological specimens, with liquid and homogeneous properties (as part of their natural characteristics or after pre-treatment).
Some anomalies have been identified during manufacturing controls.
- 2017-09-20FDA-DeviceBeckman Coulter Inc.Class IISekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnostics (UK) Limited, United Kingdom, and Sekisui Diagnostics P.E.I. Inc., Canada. Used in laboratory testing.
Reagent carryover from Sekisui ACTH when utilized on the BEC AU 680 analyzer may lead to falsely elevated BEC Lipase recovery exceeding the total precision specification.
- 2017-09-20FDA-DeviceZimmer Biomet, Inc.Class IIHerga foot switch, model 6289-WS, a component of the ROSA Robotized Stereotactic Assistant Surgical Device, Model 2.5.8. The firm name on the foot switch label is Herga Electric Limited, Bury, St. Edmunds, Suffolk IP32 6NN. ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a pre-operative plan developed with three-dimensional
Complaints were received reporting the system would freeze/shut down while in Fulgurate mode.
- 2017-09-20FDA-DeviceBecton Dickinson & CompanyClass IIBD Needle 22GA 1-1/2in SafetyGlide(TM) Non Sterile, Bulk, REF 301701 The BD SafetyGlide(TM) needle is intended to be used for general purpose injection, aspiration of fluids from parts of the body below the surface of the skin, injection into intravenous ports and tubings external to the patient. The needle stick prevention feature of the device contains a mechanism that covers the needlepoint after use. Bulk Non Sterile configuration is intended for further manufacturing, processing or repac
Presence of loose polypropylene foreign matter above release specification.
- 2017-09-20FDA-DeviceImplant Direct Sybron Manufacturing, LLCClass IISpectra-System/Legacy Guided Surgery Handle Kit, Device Listing Number D245883. The Guided Handle Kit (G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consist of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the g
The inside diameter of the 2.3mm small drill guide inserts have a smaller diameter than the specifications require.
- 2017-09-20FDA-DeviceToshiba American Medical SystemsClass IIToshiba Medical Radrex - i Digital Radiography X-ray system Radrex-i is intended for use in conjunction with the ceiling suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck, and limbs. This system is used for image acquisition, image display and trans mission/ output of images to external devices.
It was discovered during a procedure that when the operator made an exposure on the wireless x-ray detector and the image data was sent to the digital radiography system, and error message was displayed " System Error (2063)" which required a reboot and loss of the image.
- 2017-09-20FDA-DeviceLeonhard Lang Medizintechnik GmbHClass ISkintact Electrode for Defibrillation, Pacing, Monitoring, Cardioversion. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only. Used with automated external defibrillators
Certain lots of the defibrillation electrodes model DF59N and DF59NC may have a delay in connecting or not connect at all with the defibrillator ZOLL AED PLUS. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.
- 2017-09-20FDA-DeviceX-Laser LlcClass IIX-Laser Laser Light Show projector models Aurora, Caliente Aurora, PSX-400, Mobile Beat Max, Mobile Beat Max MKII, X-Beam, Hawk 500 High-power laser light show projector.
Failure to comply with performance standard requirements (21 CFR 1040.10(f))when operated in the user accessible auto and music modes of operation
- 2017-09-13FDA-DeviceBeckman Coulter Inc.Class IICoulter HmX CP and Coulter HmX AL Analyzers The purpose of the HmX Hematology Analyzer is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies. These studies include further measurements of cell size and cell distribution, biochemical investigation or any other test that helps diagnose the abnormality
Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the LH 750, LH 780, LH 500 and HmX Analyzers may not flag or detect blasts in some blood samples.