Medical device recalls
39229 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2017-06-14FDA-DeviceConvaTec, IncClass IINon-Rebreathing Mask, Reservoir Bag, Check Valve, One Side Valve, with Tubing, Product Code 106-E, Product Code 3226MM, and Product Code 9108MM
- 2017-06-14FDA-DeviceConvaTec, IncClass IIRebreathing Mask, Reservoir bag, with Tubing, Product Code 104MM
- 2017-06-14FDA-DeviceConvaTec, IncClass IIOxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E;
Reports of tubing detaching from oxygen masks either prior to use (in packaging) or during use were made by different end users and for different batches. The failures were caused by an improper solvent bond between the oxygen tubing and the oxygen mask connector. While the defect affects less than 2% of recalled product, there is a potential for harm when oxygen supply to the user is disrupted.
- 2017-06-14FDA-DeviceOrthosoft, Inc. dba Zimmer CASClass IISesamoid Plasty, Stereotaxic Instrument The Sesamoid Plasty workstation is used to run the tracking software as part of the software based navigation systems. The affected Sesamoid Plasty components are the central structural column and its base that hold up the computer and the tracking camera on the Sesamoid Plasty workstation. They are also designed to be disassembled such that the column can be folded onto itself for ease of transport after use. This is a retrospective recall report for a r
Zimmer CAS voluntarily conducted a retrospective recall of the Sesamoid Plasty due to the risk of the top of the workstation to potentially fall over during disassembly.
- 2017-06-14FDA-DeviceMagellan Diagnostics, Inc.Class IMagellan Diagnostics LeadCare Ultra and LeadCare Plus test Kits. Product Usage: The LeadCare Blood Lead Test Kit is for in vitro diagnostic use only. The Test Kit is for the quantitative measurement of lead in fresh whole blood. This product is for professional use only. Sensors and Treatment Reagent of the Test Kit are specific for lead only. Contents of this LeadCare Blood Lead Test Kit must only be used with a LeadCare Analyzer and Blood Lead Testing System.
Underestimates the lead concentration of venous blood samples when the sample is analyzed with the LeadCare systems
- 2017-06-14FDA-DeviceMagellan Diagnostics, Inc.Class IMagellan Diagnostics LeadCare and LeadCare II test Kits. Product Usage: The LeadCare Blood Lead Test Kit is for in vitro diagnostic use only. The Test Kit is for the quantitative measurement of lead in fresh whole blood. This product is for professional use only. Sensors and Treatment Reagent of the Test Kit are specific for lead only. Contents of this LeadCare Blood Lead Test Kit must only be used with a LeadCare Analyzer and Blood Lead Testing System.
Underestimates the lead concentration of venous blood samples when the sample is analyzed with the LeadCare systems
- 2017-06-14FDA-DeviceRoche Diagnostics CorporationClass IICobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional connection of the pre-analytical systems cobas p 612/ p 512 to analysis instruments from Roche Diagnostics, Hitachi or Sysmex by means of the Aloka transportation system
a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to involved operators and can also lead to potential cross contamination of samples in the affected racks.
- 2017-06-14FDA-DeviceAcumedia Manufacturers, Inc.Class IIIUrea Base Agar, Acumedia PN 7226 Urea Agar Base is used with agar for the differentiation of microorganisms on the basis of urease production
Storage temperature listed on the label was incorrect. Label stated 2-30¿C. Actual storage temperature is 2-8¿C.
- 2017-06-14FDA-DeviceIba DosimetryClass IILarge Stealth Chamber, Catalog Numbers: DS50-001-XL-T, DS50-002-XL-T, DS50-003-XL-T, DS50-102-XL-T, DS50-103-XL-T
When the Stealth Chamber is used as a reference detector in relative dosimetry, deviation in the measured PDDs at shallow depths and geometric shadowing of measured profiles may appear.
- 2017-06-14FDA-DeviceLaserex Systems Inc.Class IIEllex Tango Reflex with slit lamp Laser Ophthalmic In the SLT mode, the device is intended to be used for selective laser trabeculoplasty (ST) operations (laser trabeculoplasty for primary open angle glaucoma). In the VAG mode it is intended to be used to perform procedures requiring the rupture of tissue in the eye for Iridotomy and Iridectomy, Posterior capsulotomy and Posterior membranectomy.
It was discovered the unit produced a laser emission without pressing the fire button when the slit-lamp was driven to its lowest position prior to use on any patient.
- 2017-06-07FDA-DeviceExact Sciences CorporationClass IIExact Sciences Sample Mixer 2. For In Vitro Diagnostic Use. The Exact Sciences Sample Mixer 2 is intended for mixing stool samples in a Sample Container as part of pre-analytic sample processing for in vitro diagnostic tests
Exact Sciences is correcting all Sample Mixer v2 Upgrade units because of incidents of Sample Mixer door damage, which has occurred when the Sample Holder opens during a run and the Sample Holder lid makes contact with the Sample Mixer door.
- 2017-06-07FDA-DeviceInvacare CorporationClass IIAssist Bed Rail - 6632
A typographical error was noted within the user manual for the 6632 Assist Bed Rail which could result in the Assist Bed Rails being installed incorrectly.
- 2017-06-07FDA-DeviceCardinal Scale Mfg CoClass IICardinal, Digital Bariatric Scale 1) model 6876, Cap 600 and 2) model 6868 Cap 1000
One complaint of a seat on a scale failing unexpectedly which can cause an injury.
- 2017-06-07FDA-DeviceZimmer Biomet, Inc.Class IIModular Hybrid Glenoid Base Large for the Comprehensive Shoulder System, The product is placed into an AlOx coated PET pouch and vacuum sealed with an Argon flush. The vacuum sealed product is then placed into a Tyvek/film pouch and that pouch is heat sealed. The fully pouched product is then labeled and placed into a carton with an IFU. The carton is then labeled and shrink wrapped. The labels also include a sterilization indication dot. This system is intended for primary and revision total s
Zimmer Biomet is conducting a voluntary recall for a single lot of the Modular Hybrid Glenoid Base Large for the Comprehensive Shoulder System after it was determined that two pieces from the lot were moved to finished goods and shipped despite having an open Non-Conformance Report (NCR). The NCR indicated that one piece from the lot had scratches and the other non-confirming threads.
- 2017-06-07FDA-DeviceZimmer Biomet, Inc.Class IIAFFIXUS Hip Fracture Nail Product Usage: The Hip Fracture Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, interochanteric fractures, high subtrochanteric fractures and combinations of these fractures, including non-union, malunion and tumor resections. The Long Nail system is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone (including p
nail not properly engaging with the targeting jig due to certain undersized specification
- 2017-06-07FDA-DeviceAcumedia Manufacturers, Inc.Class IIIBeef Extract Powder, Acumedia PN 7228 Product Usage: Beef Extract Powder is prepared and standardized for use in microbiological culture media, where it is generally used to replace infusion of meat. Culture media containing Beef Extract Powder are recommended for use in bacteriological examination of water, milk, and other materials, where uniform composition of media is important. Beef Extract Powder is relied upon for biochemical studies, particularly fermentation reactions because of its
Incorrect expiration date on label. Correct expiration date was March 25, 2019. Product was labeled with expiration date of March 28,2019.
- 2017-06-07FDA-DeviceAcumedia Manufacturers, Inc.Class IIINutrient Gelatin, Acumedia Part Number 7471 A Product Usage: Nutrient Gelatin is used for the differentiation of microorganisms on the basis of gelatinase production. Identifying fermentative and non-fermentative Gram-negative bacilli include testing for gelatin liquefaction. If the proteolytic enzyme gelatinase is present, gelatin is hydrolyzed and loses its gelling characteristic.
Incorrect expiration date was listed on the label. Correct expiration date was November 30, 2016. Product was labeled with expiration date of April 30, 2018.
- 2017-06-07FDA-DeviceAcumedia Manufacturers, Inc.Class IIISabouraud Dextrose Agar w /Lecithin and Tween 20 Acumedia PN 91081 Product Usage: Used for the isolation of fungi from surfaces sanitized with quaternary ammonium compounds
Storage temperature listed on the label was incorrect. Label stated 2-30¿C. Actual storage temperature is 2-8¿C.
- 2017-06-07FDA-DeviceMadison Polymeric EngineeringClass IIDraco Enzymatic Deep-Cleaning Pad with Enzymatic Detergent Catalogue Number: HY0305Z
Product is contaminated with Cladosporium cladosporioides/herbarum/phaenocomae.
- 2017-06-07FDA-DeviceCell Marque CorporationClass IIHHV-8 (13810); 1.0 ml catalog number 265M-16; 7.0 ml catalog number-265M-18; Microbiology - Analyte Specific Reagent In vitro diagnostic use - Analyte Specific Reagent
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- 2017-06-07FDA-DeviceCell Marque CorporationClass IICD117, c-kit (YR145) 1.0 ml catalog number 117R-16 7.0 ml catalog number-18 Microbiology - Analyte Specific Reagent In vitro diagnostic use - Analyte Specific Reagent
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- 2017-06-07FDA-DeviceCell Marque CorporationClass IIVaricella Zoster Virus 7.0 ml catalog number 364M-18 Microbiology - Analyte Specific Reagent In vitro diagnostic use - Analyte Specific Reagent
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- 2017-06-07FDA-DeviceCell Marque CorporationClass IISV40 (MRQ-4) 0.5 ml catalog number 351 M-15. Microbiology - Analyte Specific Reagent In vitro diagnostic use - Analyte Specific Reagent
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- 2017-06-07FDA-DeviceCell Marque CorporationClass IIHerpes Simplex Virus II 0.5 ml catalog number 362A-15 7.0 mL Catalog number362A-18 . Microbiology - Analyte Specific Reagent In vitro diagnostic use - Analyte Specific Reagent
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- 2017-06-07FDA-DeviceCell Marque CorporationClass IIHerpes Simplex Virus I; 7.0 mL Catalog number 361A-18. Microbiology - Analyte Specific Reagent In vitro diagnostic use - Analyte Specific Reagent
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- 2017-06-07FDA-DeviceCell Marque CorporationClass IIHer2/Neu (EP3); 0.5 mL Catalog number 237R-25; 1.0 mL Catalog number 237R-26; 7.0 mL Catalog number 237R-28. Immunology - Analyte Specific Reagent In vitro diagnostic use - Analyte Specific Reagent
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- 2017-06-07FDA-DeviceCell Marque CorporationClass IIHer2/Neu (c-erbB-2) (CB-1 1); 7.0 mL Catalog number 237M-18. In vitro diagnostic use - Analyte Specific Reagent Immunology - Analyte Specific Reagent
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- 2017-06-07FDA-DeviceCell Marque CorporationClass IIH. pylori (polyclonal} 1.0 mL Catalog number 215A-76; Hematology - Analyte Specific Reagent In vitro diagnostic use - Analyte Specific Reagent
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- 2017-06-07FDA-DeviceCell Marque CorporationClass IIEstrogen Receptor (SP1) 7.0 mL Catalog number 249R-18; Hematology - Analyte Specific Reagent In vitro diagnostic use - Analyte Specific Reagent
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- 2017-06-07FDA-DeviceCell Marque CorporationClass IIEstrogen Receptor (EP1) 7.0 mL Catalog number 249R-28; 1.0 mL Catalog number 249R-16 Microbiology - Analyte Specific Reagent In vitro diagnostic use - Analyte Specific Reagent
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.