- 2017-05-24FDA-DeviceDel Mar Reynolds Medical, Ltd.Class IISpacelabs Healthcare Arkon Anesthesia Delivery System is intended for use in the hospital environment and operating room. Absorber Adapter Assembly Service Kit, part number: 050-9045-00. Absorber Adapter Assembly, part number: 650-1648-01. Arkon Absorber Canister Carrier, part number: 380-0916-01.
Reports of the Canister Carrier breaking during absorbent exchange or during anesthesia machine transit, causing a leak in the breathing circuit, which could cause hypoventilation in the patient.
- 2017-05-24FDA-DeviceVortran Medical Technology 1, IncClass IIVORTRAN Automatic Resuscitator, VARPlus with Extension Kit; Model PCE-5012; Qty: 10 each. Anesthesiology: This device is to be used by properly trained personnel to deliver emergency, short term, constant flow, pressure cycled ventilatory support.
Vortran received a customer complaint that the spin-nut-DSS connector was assembled in the reverse direction on the oxygen tubing; making the connector unable to be screwed on due to the wrong thread direction.
- 2017-05-24FDA-DeviceVortran Medical Technology 1, IncClass IIVORTRAN Automatic Resuscitator, VARPlus; Model PTE-5002; Qty: 10 each Anesthesiology: This device is to be used by properly trained personnel to deliver emergency, short term, constant flow, pressure cycled ventilatory support.
Vortran received a customer complaint that the spin-nut-DSS connector was assembled in the reverse direction on the oxygen tubing; making the connector unable to be screwed on due to the wrong thread direction.
- 2017-05-24FDA-DeviceHitachi Medical Systems America IncClass IIArietta 70 Ultrasound System Intended for use by trained personnel (doctor, sonographer. etc.) for the diagnostic Ultrasound evaluation.
The strength of the arm mounting mechanism was insufficient and had to be redesigned.
- 2017-05-24FDA-DeviceAesculap Implant Systems LLCClass IIGE426R; HI-LINE XS DIAMOND BURR X-CRS.I D4.5MM; GE426SU; HI-LINE XS DISP.DIAM.BURR X-CRS.I D4.5MM; GE526R; HI-LINE XS DIAMOND BURR X-CRS.II D4.5MM; GE526SU; HI-LINE XS DISP.DIAM.BURR X-CRS.II D4.5; GE626R; HI-LINE XS DIAMOND BURR X-CRS.III D4.5MM; GE626SU; HI-LINE XS DISP.DIAM.BURR X-CRS.III D4.5; Product Usage: The ELAN 4 motor system and the Hi-Line XS handpiece are intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of S
Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction may occur. These particulates could also lead to mechanically induced soft tissue damage.
- 2017-05-24FDA-DeviceAesculap Implant Systems LLCClass IIGP321R; ELAN 4 2-RING DIAMOND BURR COARSE D2.3; GP322R; ELAN 4 2-RING DIAMOND BURR COARSE D3.0; GP323R; ELAN 4 2-RING DIAMOND BURR COARSE D4.0; GP324R; ELAN 4 2-RING DIAMOND BURR COARSE D5.0; GP325R; ELAN 4 2-RING DIAMOND BURR COARSE D6.0; Product Usage: The ELAN 4 motor system and the Hi-Line XS handpiece are intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.
Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction may occur. These particulates could also lead to mechanically induced soft tissue damage.
- 2017-05-24FDA-DeviceAesculap Implant Systems LLCClass IIGP161R; ELAN 4 1-RING DIAMOND BURR COARSE D2.0; GP162R; ELAN 4 1-RING DIAMOND BURR+ COARSE D2.0; GP163R; ELAN 4 1-RING DIAMOND BURR COARSE D2,3; GP164R; ELAN 4 1-RING DIAMOND BURR+ COARSE D2,3; GP165R; ELAN 4 1-RING DIAMOND BURR COARSE D3,0; GP166R; ELAN 4 1-RING DIAMOND BURR+ COARSE D3,0; GP167R; ELAN 4 1-RING DIAMOND BURR++ COARSE D3,0; GP168R; ELAN 4 1-RING DIAMOND BURR COARSE D4,0; GP169R; ELAN 4 1-RING DIAMOND BURR COARSE D5,0; GP170R; ELAN 4 1-RING DIAMOND BURR
Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction may occur. These particulates could also lead to mechanically induced soft tissue damage.
- 2017-05-24FDA-DeviceSiemens Healthcare Diagnostics, IncClass IIADVIA Centaur Systems Insulin ReadyPack; Test Code: IRI; Catalog Number: 02230141, for in vitro diagnostic use in the quantitative determination of insulin in serum.
Current lots of ADVIA Centaur/XP/XPT/CP Insulin (IRI) assay recover approximately 40% higher than the World Health Organization (WHO) 1st IRP 66/304 standardization based on slope values. Other performance characteristics are being met per the assay Instructions for Use. Overall risk to health is negligible and Siemens is not recommending a review of previously generated results.
- 2017-05-24FDA-DeviceSiemens Healthcare Diagnostics, IncClass IIADVIA Centaur Systems Insulin Calibrator; Test Code: Calibrator IRI; Catalog Number: 04618899
Current lots of ADVIA Centaur/XP/XPT/CP Insulin (IRI) assay recover approximately 40% higher than the World Health Organization (WHO) 1st IRP 66/304 standardization based on slope values. Other performance characteristics are being met per the assay Instructions for Use. Overall risk to health is negligible and Siemens is not recommending a review of previously generated results.
- 2017-05-24FDA-DeviceBeckman Coulter Inc.Class IISalicylate Reagent (SALY 2 x 45), Catalog No. 378194. Intended for quantitative determination of salicylate concentration in human serum or plasma.
Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.
- 2017-05-24FDA-DeviceBeckman Coulter Inc.Class IITriglycerides Reagent (TG 2 x 300), Catalog No. 445850. Intended for quantitative determination of triglycerides concentration in human serum or plasma.
Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.
- 2017-05-24FDA-DeviceBeckman Coulter Inc.Class IIHDL Cholesterol Reagent (HDL 2 x 200), Catalog No. A15625. Intended for quantitative determination of HDL cholesterol in the high density lipoprotein fraction of human serum or plasma.
Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.
- 2017-05-24FDA-DeviceBeckman Coulter Inc.Class IICholesterol Reagent(CHOL 2 x 300), Catalog No. 467825 Intended for quantitative determination of cholesterol concentration in human serum or plasma.
Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.
- 2017-05-24FDA-DeviceBeckman Coulter Inc.Class IIEnzymatic Creatinine Reagent (CR-E 2 x 200), Catalog No. A60298. Intended for the quantitative determination of creatinine (CR-E) concentration in human serum, plasma or urine.
Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.
- 2017-05-24FDA-DeviceBeckman Coulter Inc.Class IIUric Acid Reagent (URIC 2 x 300), Catalog No. 442785. Intended for the quantitative determination of uric acid concentration in human serum, plasma or urine.
Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.
- 2017-05-24FDA-DeviceBeckman Coulter Inc.Class IILactate Reagent (LACT 2 x 50), Catalog No. A95550 Intended for the quantitative determination of Lactate concentration in human plasma and cerebrospinal fluid (CSF).
Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.
- 2017-05-24FDA-DeviceOrtho-Clinical DiagnosticsClass IIenGen (TM) Laboratory Automation Systems using all TCAutomation(TM) Software Versions with the InOut Communication Interface, IVD
Software anomaly; Thermo-Fisher Scientific initially discovered and Ortho-Clinical Diagnostics, subsequently, confirmed a software anomaly that may potentially result in miss-associated sample IDs involving the communication between enGen TCAutomation (TCA) Bypass modules (manufactured by Thermo-Fisher Scientific) and the VITROS Analyzers. To date, no occurrences of this issue have been observed on Orthos enGen(TM) Systems. Sample IDs may become miss-associated with the actual sample being processed due to discrepant information passed to the analyzer through InOut communication device, triggered by a specific set of events occurring between the Bypass module and the analyzer. --- All of the following conditions must be present for the issue to potentially occur: (1.) The Bypass module is in the process of positioning Sample A at the aspiration position, and (2.) At least two additional sample tubes (Samples B and C) are waiting to be aspirated in the Bypass module queue, and (3.) A Radio Frequency Identification (RFID) Tag Read failure occurs on the Sample A carrier, and (4.) The VITROS System is not available to aspirate on the track (for example, the analyzer is initializing, or the user is making entries on the diagnostic screen, or samples are being aspirated from the front of the analyzer, etc.). --- If the software anomaly occurs and the issue is present, the Sample ID in the INOUT communication device software queue becomes asynchronous with the actual Sample ID physical queue, and the VITROS assay result will be reported for the wrong Sample ID. All subsequent samples will have miss-associated results until the Bypass module is reset or the queue is empty. --- This issue could potentially miss-associate a sample/patient ID with a test result of a different sample. Depending on the test involved, patients health condition, etc., the potential health risk posed to the affected patient could range from no risk to serious risk, and potentially could lead to misdiagnosis or inappropriate medical intervention. To date, no patient harm was reported related to this product issue.
- 2017-05-24FDA-DeviceOrthosoft, Inc. dba Zimmer CASClass IINavitrack System - OS Knee Universal, Stereotaxic instrument
Zimmer CAS voluntarily conducted a retrospective recall of the Navitrack System OS Knee Universal software versions 2.2.0.82, 2.2.1.3, 2.1.6.0.2, D.54 and A.42 due to software issues resulting in erros and discrepancies .
- 2017-05-24FDA-DeviceOrthosoft, Inc. dba Zimmer CASClass IIINavitrack¿ System - OS Knee Universal, Orthopedic Stereotaxic Instrument
ZIMMER CAS voluntarily conducted a retrospective recall of the Navitrack System - OS Knee Universal, due to software anomalies which were corrected with a new revision in version 2.2.1.3.
- 2017-05-24FDA-DeviceBeckman Coulter Inc.Class IIAccess BR Monitor, Part No. 387620
Following biases observed in the results of the French National Quality Controls, Beckman Coulter has been informed that a study completed by the French Competent Authority Agence nationale de s¿curit¿ du Medicament et des produits de sant¿ (ANSM) showed diagnosed cancer patients in non-remission of their metastatic disease, may have results below the upper reference limit (URL) of 31.3 U/mL as stated in the Access BR Monitor instructions for use (IFU).
- 2017-05-24FDA-DeviceBiomet Microfixation, LLCClass IIBiomet Microfixation TraumaOne System Surgical Tray - Model # 46-1600
Part of the number on the tray lid differ from the part number on the base.
- 2017-05-24FDA-DeviceEden Spine Europe SAClass IIGIZA Vertebral Body Replacement
Complaints that implants have been disassembled by surgeons because of unscrewing completely the locking screw.
- 2017-05-24FDA-DeviceRespironics California IncClass IPhilips V60 Ventilators, Material No. 850008, 1053613, 1053614, 1053615, 1053616, 1053617, 1076715, 1076716, 1076717, DU1053617, U1053617. Old Material No. 850008, 989805628251, 989805612101, 989805613391, 989805613661, 989805611761, 989805627411, 989805627431, 989805627441, 989805616411, 989805636631.
Over time, low-frequency vibrations can cause the pins within the female connectors on the Motor Controller to Data Acquisition Board ribbon cable to become partially displaced causing momentary high resistance that prevents or restricts data travelling through the cable during the high resistance event.
- 2017-05-24FDA-DeviceAbbott VascularClass INC Tenku RX PTCA Balloon Catheter, Part No. 1012448-08T, 1012448-12T, 1012448-15T, 1012449-08T, 1012449-12T, 1012449-15T, 1012450-08T, 1012450-12T, 1012450-15T, 1012451-08T, 1012451-12T, 1012451-15T, 1012452-08T, 1012452-12T, 1012452-15T, 1012453-08T, 1012453-12T
Products from the identified lots may exhibit difficulty in removing the protective balloon sheath which can result in issues with inflating or deflating the balloon.
- 2017-05-24FDA-DeviceAbbott VascularClass INC Traveler RX Coronary Dilatation Catheter, Part No. 1013152-12, 1013152-15, 1013152-20, 1013153-08, 1013153-12, 1013153-15, 1013153-20, 1013154-08, 1013154-12, 1013154-15, 1013154-20, 1013155-08, 1013155-12 1013155-15, 1013155-20, 1013156-12, 1013156-15, 1013157-08, 1013157-12, 1013157-15
Products from the identified lots may exhibit difficulty in removing the protective balloon sheath which can result in issues with inflating or deflating the balloon.
- 2017-05-24FDA-DeviceAbbott VascularClass INC Trek RX Coronary Dilatation Catheter, Part No. 1012448-06, 1012448-08, 1012448-12, 1012448-15, 1012448-20, 1012449-06, 1012449-08, 1012449-12, 1012449-15, 1012449-20, 1012449-25, 1012450-06, 1012450-08, 1012450-12, 1012450-15, 1012450-20, 1012451-06, 1012451-08, 1012451-12, 1012451-15, 1012451-20, 1012451-25, 1012452-08, 1012452-12, 1012452-15, 1012452-20, 1012453-08, 1012453-12, 1012453-15, 1012453-20
Products from the identified lots may exhibit difficulty in removing the protective balloon sheath which can result in issues with inflating or deflating the balloon.
- 2017-05-24FDA-DeviceMagellan Diagnostics, Inc.Class IMagellan Diagnostics LeadCare Ultra Blood Lead Test Kits Part Number: 80-0010
LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.
- 2017-05-24FDA-DeviceMagellan Diagnostics, Inc.Class IMagellan Diagnostics LeadCare Ultra Blood Analyzer Part Number: 80-0010
LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.
- 2017-05-24FDA-DeviceReFlow MedicalClass IWingman 35 Crossing Catheter, 65cm, Model Numbers: WGM35065US WGM35065CE WGM35090US WGM35090CE WGM35135US WGM35135CE
ReFlow Medical is recalling the Wingman 35 Crossing Catheter because it may be prone to tip detachment.
- 2017-05-24FDA-DeviceZiehm Imaging IncClass IIZiehm Imaging solo FD Mobile Fluoroscopic C-arm
During Ziehm Imaging GmbH FDA inspection, the label for the Collimator and X-ray Control panel was missing the manufacturer date as part of the identification for certified component