- 2026-04-08FDA-DevicePhilipsClass IINationwideCheck your Philips Allura Xper FD20/20 X-ray system
Philips has identified a problem with the wired foot switch on certain Allura Xper FD20/20 X-ray imaging systems. The foot switch may fail to start X-ray imaging or cause imaging to work only intermittently, which could affect clinical procedures.
- 2026-04-08FDA-DevicePhilipsClass IINationwideCheck your Philips Allura Xper X-ray system
Philips has identified a problem with the wired foot switch on certain Allura Xper FD20/10 X-ray systems that may prevent imaging from starting or cause it to work intermittently. This affects 85 systems worldwide, including 20 in the United States. No injuries have been reported.
- 2026-04-08FDA-DevicePhilipsClass IINationwideContact Philips about your ALLURA Xper FD10 table
Philips has identified a problem with the wired foot switch on certain ALLURA Xper FD10 operating room tables where X-ray imaging may not start or may work only intermittently. Two units worldwide are affected. No injuries have been reported.
- 2026-04-08FDA-DevicePhilipsClass IINationwideStop using Philips ALLURA Xper FD20 foot switch
Philips has found that the wired foot switch on certain ALLURA Xper FD20 operating room tables may fail to start X-ray imaging or may work only intermittently. This affects 4 units worldwide. No injuries have been reported.
- 2026-04-08FDA-DevicePhilipsClass IINationwideStop using affected Philips ALLURA X-ray tables
Philips has identified a problem with the wired foot switch on certain ALLURA Xper FD20 surgical X-ray tables. The switch may fail to start X-ray imaging or may cause imaging to work only intermittently, which could disrupt surgical procedures. This affects 196 units worldwide.
- 2026-04-08FDA-DevicePhilipsClass IINationwideStop using affected Philips ALLURA X-ray systems
Philips has identified a problem with the wired foot switch on certain ALLURA Xper FD20 Biplane X-ray systems where imaging may not start or may work intermittently. This affects 461 units worldwide, including 198 in the US. No injuries have been reported.
- 2026-04-08FDA-DevicePhilipsClass IINationwideStop using affected Philips X-ray foot switches
Philips has identified a problem with the wired foot switch on certain ALLURA Xper FD20 X-ray systems where imaging may not start or may start intermittently. This affects specific serial numbers across three system codes. No injuries have been reported.
- 2026-04-08FDA-DevicePhilipsClass IINationwideCheck your Philips ALLURA Xper system serial number
Philips has identified a problem with the wired foot switch on certain ALLURA Xper FD10/10 X-ray systems where imaging may not start or may work on and off intermittently. This affects 112 systems distributed worldwide, identified by specific serial numbers.
- 2026-04-08FDA-DevicePhilipsClass IINationwideStop using Philips ALLURA Xper FD10 foot switch
Philips has identified a defect in the wired foot switch on certain ALLURA Xper FD10 X-ray imaging systems that may prevent X-ray imaging from starting or cause it to work intermittently. This could delay or disrupt imaging procedures. Philips is working with healthcare facilities to resolve the issue.
- 2026-04-08FDA-DevicePhilipsClass IINationwideCheck your Philips ALLURA X-ray foot switch
Philips has identified a problem with the wired foot switch on certain ALLURA Xper FD10F X-ray imaging systems. The foot switch may fail to start X-ray imaging or may cause imaging to work intermittently, which could affect clinical procedures. The issue affects 10 systems total, with no injuries reported.
- 2026-04-08FDA-DevicePhilipsClass IINationwideContact Philips about Allura Xper FD10C foot switch issue
Philips has identified a problem with the wired foot switch on certain Allura Xper FD10C X-ray imaging systems. When using the foot switch, X-ray imaging may not start or may start intermittently, which could affect clinical workflow. This affects a small number of systems worldwide.
- 2026-04-08FDA-DeviceSiemensClass IINationwideCheck Siemens Atellica CH A1c_E test results
Siemens has recalled certain lots of its Atellica CH A1c_E hemoglobin test kit because when used with another reagent (RCRP) on the same analyzer, the A1c test may give lower-than-accurate results. Lower results could delay diabetes diagnosis or lead to inadequate treatment. This issue does not affect labs using separate analyzers for each test.
- 2026-04-08FDA-DeviceReCor MedicalClass IINationwideDo not use ReCor Medical Paradise catheter
ReCor Medical distributed 24 units of Paradise ultrasound renal denervation catheters (Model PRDS-068-02) that did not meet quality standards and were supposed to be destroyed, not sold. The devices were distributed in nine states: California, Colorado, Florida, Kansas, Missouri, North Carolina, Ohio, Texas, and Virginia. No injuries have been reported.
- 2026-04-08FDA-DeviceReCor MedicalClass IINationwideCheck Paradise renal denervation catheter lot M4907
ReCor Medical distributed one Paradise ultrasound renal denervation catheter (Model PRDS-068-02, Lot M4907) that did not meet manufacturing standards and was supposed to be destroyed. The device was distributed in nine U.S. states. No injuries have been reported.
- 2026-04-08FDA-DevicePhilipsClass INationwideCheck your Philips Respironics ventilator settings
Philips Respironics has recalled certain Trilogy Evo Universal home ventilators (Software Version 1.05.15.00) because the obstruction alarm may not trigger as quickly as safety standards require. In some cases, the alarm could be delayed by up to four breaths, which may prevent timely notification of a breathing tube blockage.
- 2026-04-08FDA-DevicePhilipsClass INationwideCheck your Philips Respironics ventilator settings
Philips Respironics has recalled certain Trilogy EV300 home ventilators (software version 1.05.15.00) because the obstruction alarm may not trigger quickly enough if the breathing tube becomes blocked. In rare cases, the alarm could be delayed by up to four breaths, which is longer than safety standards allow.
- 2026-04-08FDA-DevicePhilipsClass INationwideCheck your Philips Respironics ventilator settings
Some Philips Respironics Trilogy Evo O2 ventilators with software version 1.05.15.00 may not sound an obstruction alarm quickly enough if the breathing tube becomes blocked. The alarm could be delayed by up to four breaths, which could reduce the time available to respond to a blockage.
- 2026-04-08FDA-DevicePhilipsClass INationwideCheck your Philips Respironics Trilogy Evo ventilator
Some Philips Respironics Trilogy Evo home ventilators with Software Version 1.05.15.00 may not sound their obstruction alarm quickly enough if the breathing tube becomes blocked. The alarm could be delayed by up to four breaths, which could reduce the time a user has to respond to an airway obstruction.
- 2026-04-08FDA-DeviceAbiomedClass INationwideCheck your Impella Controller serial number
Abiomed's Impella Controller may display a delayed "Purge System Blocked" alarm when used with first-generation Impella 5.5 pumps. This could prevent timely detection of a blocked purge system. The issue affects 62 controllers with specific serial numbers in software versions 8.2.2 through 8.5.
- 2026-04-08FDA-DeviceAbiomedClass INationwideCheck your Abiomed Impella pump cassette
Abiomed is recalling purge cassettes used in Impella heart pump systems because Generation 1 cassettes have an increased risk of leaking during use. This affects multiple Impella pump models sold globally, with specific batch numbers identified from 2024.
- 2026-04-08FDA-FoodImu-Tek Animal HealthClass IIStop using Imu-Tek Colostrum-5 Powder
Imu-Tek Colostrum-5 Powder (Lot #22, expiration 020528) may not have been processed adequately, which could allow harmful bacteria to survive. The affected product was distributed across 29 states. No illnesses have been reported so far.
- 2026-04-08FDA-FoodImu-Tek Animal HealthClass IIStop using Imu-Tek Colostrum-5 capsules
Imu-Tek Animal Health is recalling Colostrum-5 120-capsule bottles (Lot #013, expiration 092826) because the product may be under-processed, which can affect its safety. The affected bottles were distributed across 33 states.
- 2026-04-08FDA-FoodHey TeaClass IIStop using HEYTEA coconut drink
HEYTEA USA's Coconut Drink 1 (1L cartons) contains milk that is not listed on the label. The product was distributed to food service establishments in multiple states. This poses a serious risk to anyone with a milk allergy or severe intolerance.
- 2026-04-08FDA-FoodSuper World TradingClass IIIStop eating Mei Heong Yuen roasted peanuts
Mei Heong Yuen Garlic Flavor Roasted Peanuts (9.17 oz) contain cyclamates, a sweetener banned in the United States. The product was distributed to multiple states under UPC 6912449124329 with manufacture date July 10, 2025. No injuries have been reported.
- 2026-04-08FDA-FoodSuper World TradingClass IIIDon't eat Mei Heong Yuen roasted peanuts
Mei Heong Yuen walnut-flavor roasted peanuts contain sodium cyclamate, a sweetener banned in the United States. The 9.17 oz packages were distributed to multiple states. No illnesses have been reported.
- 2026-04-08FDA-FoodSuper World TradingClass IIIDon't eat Mei Heong Yuen roasted peanuts
Mei Heong Yuen tangerine flavor roasted peanuts contain sodium cyclamate, a sweetener banned in the United States. The product was distributed to 19 states. No injuries have been reported.
- 2026-04-08FDA-FoodLone Star Dairy ProductsClass INationwideStop using Lone Star Dairy milk powder
Lone Star Dairy Products LLC is recalling spray dried milk powder products because Salmonella bacteria was found during testing of finished products. The recall affects specific lot codes of NFDM Low Heat, NFDM High Heat, and SMP Medium Heat products sold in 25 kg bags and 1 metric ton totes.
- 2026-04-08FDA-FoodPocasClass IIIDon't drink Pocas popping boba RTD tea
Pocas International Corp. is recalling Popping Boba RTD Mango Passion Fruit Green Tea because of a packaging integrity issue that may allow the product to degrade or become contaminated. The recall affects 517 cases distributed across nine states and Canada. No injuries have been reported.
- 2026-04-08FDA-FoodPocasClass IIIDon't consume Pocas popping boba tea
Pocas International Corp. is recalling its Popping Boba RTD Strawberry Dragon Fruit/Oolong Tea because the packaging may not be properly sealed. This could allow the product to spoil or become contaminated. The affected product was distributed to nine states and Canada.
- 2026-04-08FDA-FoodPocasClass IIIDon't drink Pocas popping boba tea
Pocas International Corp. is recalling Popping Boba RTD Mixed Berry Hibiscus Tea because of a packaging integrity issue that may allow the product to spoil or become contaminated. The recall affects 148 cases distributed to New York, Virginia, Florida, Massachusetts, Texas, Louisiana, Utah, California, and Canada.
Get notified about new Massachusetts recalls
Free weekly digest. Add Massachusetts to your profile to filter alerts to your state.
Get the Sunday Brief