- 2026-01-21FDA-DeviceMedlineClass IINationwideStop using affected Medline pre-op kits
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
- 2026-01-21FDA-DeviceMedlineClass IINationwideStop using recalled Medline IV kits
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
- 2026-01-21FDA-DeviceMedlineClass IINationwideMedline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) DRAWER 2 IV SUPPLIES, Medline kit SKU ACC010271B; 2) DRAWER 2 IV FLUIDS & TUBING, Medline kit SKU ACC010416B; 3) OXYGENATION & TUBING TRAY, Medline kit SKU ACC010525C; 4) TOP DRAWER #6, Medline kit SKU ACC010688; 5) RRMC CODE CART DRAWER 3, Medline kit SKU ACC010770; 6) SURGERY BAG, Medline kit SKU DYKM1279C; 7) ADMISSION KIT, Medline kit SKU DYKS1179; 8) PRE OP KIT, Medline kit
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
- 2026-01-21FDA-DeviceMedlineClass IINationwideStop using Medline DR HOLCOMB FACIAL PACK kits
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
- 2026-01-21FDA-DeviceMedlineClass IINationwideStop using Medline IV Administration Sets
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
- 2026-01-21FDA-DeviceMedlineClass IINationwideStop using affected Medline surgical kits
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
- 2026-01-21FDA-DeviceMedlineClass IINationwideStop using affected Medline Pre Op Kits
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
- 2026-01-21FDA-DeviceMedlineClass IINationwideStop using affected Medline cardiac kits
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
- 2026-01-21FDA-DeviceMedlineClass IINationwideStop using Medline AN01 carotid anesthesia kits
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
- 2026-01-21FDA-DeviceFresenius KabiClass IINationwideCheck Fresenius Kabi Ivenix Infusion System software
Emphasizing instructions for LVP duration programming located in the IFU.
- 2026-01-21FDA-DeviceGE HealthcareClass IINationwideCheck GE MAXXUS imaging system detector support
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
- 2026-01-21FDA-DeviceGE HealthcareClass IINationwideCheck GE HealthCare nuclear medicine system detector support
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
- 2026-01-21FDA-DeviceGE HealthcareClass IINationwideCheck GE HealthCare nuclear medicine system detector support
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
- 2026-01-21FDA-DeviceGE HealthcareClass IINationwideCheck GE HealthCare nuclear medicine system support
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
- 2026-01-21FDA-DeviceMedtronic NeuromodulationClass IINationwideStop using Medtronic Vanta Clinician Programmer App
Complaints received that Vanta A71200 CP App does not function as intended during use. There is a potential for prolonged or postponed surgical procedure.
- 2026-01-21FDA-DeviceBoston ScientificClass IINationwideCheck Boston Scientific Coyote catheter batches
Boston Scientific is initiating the removal of certain batches of Coyote" Over-the-Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Balloon Dilatation Catheter devices due to the potential for inflation/deflation difficulty during use in the peripheral vasculature. An internal investigation found that a manufacturing process variation resulted in catheters that may not properly inflate/deflate. This was limited to Coyote OTW batches manufactured from 12Sep2025 through 13Oct2025. The processing variation has since been addressed. No other material numbers or batches are impacted by this removal.
- 2026-01-21FDA-DeviceMobius ImagingClass IINationwideCheck your AIRO CT pendant for safety label
AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.
- 2026-01-21FDA-DeviceMobius ImagingClass IINationwideCheck your AIRO CT pendant replacement kit
AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.
- 2026-01-21FDA-DeviceMedicreaClass IINationwideStop using Medicrea IB3D PL Instruments Set
Due to complaints, orthopedic rotation tool handle may untread from the shaft resulting in the inability to rotate implant.
- 2026-01-21FDA-DeviceMedicreaClass IINationwideStop using Medicrea IB3D implant inserter
Due to complaints, orthopedic rotation tool handle may untread from the shaft resulting in the inability to rotate implant.
- 2026-01-21FDA-DeviceRepligenClass IINationwideStop using Repligen MAVERICK Spectrometer
Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.
- 2026-01-21FDA-DeviceAbbottClass INationwideStop using recalled FreeStyle Libre 3 Plus sensors
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
- 2026-01-21FDA-DeviceAbbottClass INationwideStop using affected FreeStyle Libre 3 Plus sensors
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
- 2026-01-21FDA-DeviceAbbottClass INationwideStop using Abbott FreeStyle Libre 3 Plus sensors
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
- 2026-01-21FDA-DeviceAbbottClass INationwideStop using Abbott FreeStyle Libre 3 sensors
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
- 2026-01-21FDA-DeviceCanonClass IINationwideContact Canon Medical about your VANTAGE TITAN 3T MRI
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
- 2026-01-21FDA-DeviceCanonClass IINationwideVANTAGE GALAN 3T Model MRT-3020/MEXL-3020
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
- 2026-01-21FDA-FoodHerbs ForeverClass INationwideStop using Herbs Forever Hingwastika Churna
Potential undeclared allergen ingredient (Wheat)
- 2026-01-21FDA-FoodHerbs ForeverClass INationwideStop using Herbs Forever Gastro Care supplements
Potential undeclared allergen ingredient (Wheat)
- 2026-01-21FDA-FoodAZ GemsClass IINationwideDo not eat AZ Gems Jumbo Raw Shrimp
elevated levels of chloramphenicol in frozen shrimp