FDA-Device2026-10-27Class IIPROCESSING DEFECT

Philips Azurion 5 M20 systems recalled over unapproved SSD drives

OTHERNationwide distributionMedical devices

What this means

Philips recalled certain Azurion 5 M20 systems, models 722228 and 722281, because unspecified solid-state drives may have been installed in the PCs. These drives may perform differently and could become less reliable over time, which may affect imaging, the display, or the ability to store and retrieve images. No injuries have been reported.

  • Check whether your facility's Azurion 5 M20 is a listed model number and UDI.
  • Contact Philips for the recall remedy and instructions for the affected systems.
  • Report any display, imaging, or image storage problems to the manufacturer and your facility's biomedical team.
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.

Intended for
All ages (under 100 years)
Illnesses reported
None reported
Sold states
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.
Affected count
465 units
Manufactured in
Veenpluis 6, Best, Netherlands
Products
Philips Azurion 5 M20. Model Numbers: 722228, 722281.

Products in this recall · 9

The agency lists each product in this recall separately. Check every one you might have.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0295-2027

Want us to check the next one for you?

We read every CPSC, FDA, USDA, and NHTSA notice and email you a plain-English Sunday brief covering your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Free · One email a week · Unsubscribe anytime
By signing up you agree to our Terms and Privacy Policy.