FDA-Device2026-10-27Class IIPROCESSING DEFECT

Philips Azurion 7 M20 imaging systems recalled over unapproved SSDs

OTHERNationwide distributionMedical devices

What this means

Philips recalled 1,896 Azurion 7 M20 systems (model numbers 722079, 722224, and 722282) because some solid-state drives not listed in the device file may have been installed in their PCs. These drives may perform differently and become less reliable over time, which could lead to lost imaging or motorized movement, a blank or black display, or problems storing or accessing images. No injuries have been reported.

  • Contact Philips or your facility's biomedical engineering team to confirm whether your system is included.
  • Follow any instructions from Philips, since this notice does not specify a remedy.
  • Report any display failures, image storage problems, or unexpected motion to Philips and your facility.
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.

Intended for
All ages (under 100 years)
Illnesses reported
None reported
Sold states
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.
Affected count
1,896 units
Manufactured in
Veenpluis 6, Best, Netherlands
Products
Philips Azurion 7 M20. Model Numbers: 722079, 722224, 722282.

Products in this recall · 9

The agency lists each product in this recall separately. Check every one you might have.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0299-2027