FDA-Device2026-10-27Class IIPROCESSING DEFECT

Philips Azurion S system recalled over unapproved solid-state drives

OTHERNationwide distributionMedical devices

What this means

Philips is recalling Azurion S systems (model 722285) because solid-state drives not specified in the device file may have been installed in the PCs. These drives may perform differently and could become less reliable over time, leading to loss of imaging, a blank or black display, or trouble storing and viewing images. One unit is affected, and no injuries have been reported.

  • Contact Philips or your authorized service provider to check whether the system's PC has an unspecified SSD.
  • Confirm whether your unit's serial number matches the recall notice.
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.

Intended for
All ages (under 100 years)
Illnesses reported
None reported
Sold states
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.
Affected count
1 unit
Manufactured in
Veenpluis 6, Best, Netherlands
Products
Philips Azurion S. Model Number: 722285.

Products in this recall · 9

The agency lists each product in this recall separately. Check every one you might have.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0300-2027