FDA-Device2026-10-27Class IIPROCESSING DEFECT

Philips Azurion 7 M12 medical systems recalled over unapproved SSDs

OTHERNationwide distributionMedical devices

What this means

Philips is recalling certain Azurion 7 M12 medical imaging systems (model numbers 722078 and 722223) because non-specified solid-state drives may have been installed in the PCs that run them. These drives may perform differently and become less reliable over time, which could cause loss of imaging, a blank or black display, or trouble storing or viewing images. No injuries have been reported.

  • Contact your Philips service representative or the facility that operates this equipment for next steps.
  • Do not make changes to the device yourself; wait for Philips to confirm the fix.
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.

Intended for
All ages (under 100 years)
Illnesses reported
None reported
Sold states
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.
Affected count
602 units
Manufactured in
Veenpluis 6, Best, Netherlands
Products
Philips Azurion 7 M12. Model Numbers: 722078, 722223.

Products in this recall · 9

The agency lists each product in this recall separately. Check every one you might have.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0298-2027

Want us to check the next one for you?

We read every CPSC, FDA, USDA, and NHTSA notice and email you a plain-English Sunday brief covering your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Free · One email a week · Unsubscribe anytime
By signing up you agree to our Terms and Privacy Policy.