Philips Allura Xper FD20 X-ray system recalled for missing cooling drip tray
What this means
Philips identified that some Allura Xper FD20 X-ray systems may be missing a drip tray beneath the cooling unit from previous service work. Without this tray, coolant could leak onto electrical components and cause a short circuit, potentially shutting down the system during use.
- Check if your Allura Xper FD20 has the drip tray installed beneath the cooling unit
- Contact Philips Medical Systems if you are unsure or if the tray is missing
- Do not operate the system if the drip tray is not present
The agency's original wording
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1066-2026More Philips recalls
Want us to check the next one for you?
We read every CPSC, FDA, USDA, and NHTSA notice and email you a plain-English Sunday brief covering your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.