Philips Allura Xper FD20 Biplane X-ray system recalled for missing coolant drip tray
Philips identified that some Allura Xper FD20 Biplane X-ray systems may be missing a drip tray under the cooling unit, either from incomplete past repairs or undocumented installation. Without this tray, coolant liquid could leak onto electrical components and cause short-circuits that shut down the system.
From the FDA notice · Read the official notice ↗ · All Philips recalls
What to do
- Check your system serial number against the list of 23 affected units (serial numbers 601, 230, 370, 27, 320, 193, 60, 373, 560, 400, 549, 506, 381, 382, 132, 169, 233, 236, 243, 526, 447, 163, 403)
- If your system is affected, contact Philips Medical Systems to verify the drip tray is installed and properly documented
- Do not operate the system if you suspect the drip tray is missing
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Allura Xper FD20 Biplane; Model Number: 722013;
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
This recall covers 16 products — see all of them.
Details
- Units affected
- 23 units
- Illnesses reported
- None reported
- Manufactured in
- Veenpluis 4-6, Best, Netherlands
- Classification
- Class II
- Recall number
- Z-1071-2026
In the FDA's words
- Description
- Allura Xper FD20 Biplane; Model Number: 722013; Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown. Model Number: 722013; System Product Name: Allura Xper FD20 Biplane; UDI-DI: 00884838059061; Serial Numbers: 601, 230, 370, 27, 320, 193, 60, 373, 560, 400, 549, 506, 381, 382, 132, 169, 233, 236, 243, 526, 447, 163, 403; Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen. Affected quantity: 23 units
- Hazard
- Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
16 products in this recall
Philips Allura Xper X-ray systems recalled for coolant short-circuit risk
Philips is recalling 16 Allura Xper imaging system models because a drip tray under the cooling unit may not have been installed during earlier field corrections, or the installation was not properly documented. Without the tray, coolant could reach electrical components, possibly causing short-circuits, blown fuses, or system shutdown. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- Product Name: Allura Xper FD10C; Model number: 722001;Z-1062-2026 · 38 units
- Allura Xper FD10F; Model Number: 722002;Z-1063-2026 · 12 units
- Allura Xper FD10; Model Number: 722003;Z-1064-2026 · 93 units
- Allura Xper FD10/10; Model Number: 722005;Z-1065-2026 · 14 units
- Allura Xper FD20; Model Number: 722006;Z-1066-2026 · 143 units
- Allura Xper FD20 Biplane; Model Number: 722008;Z-1067-2026 · 18 units
- Allura Xper FD10; Model Number: 722010;Z-1068-2026 · 56 units
- Allura Xper FD10/10; Model Number: 722011;Z-1069-2026 · 21 units
- Allura Xper FD20; Model Number: 722012;Z-1070-2026 · 79 units
- Allura Xper FD20 Biplane; Model Number: 722013; (this page)Z-1071-2026 · 23 units
- Allura Xper FD20 OR Table; Model Number: 722023;Z-1072-2026 · 4 units
- Allura Xper FD10; Model Number: 722026;Z-1073-2026 · 95 units
- Allura Xper FD10/10; Model Number: 722027;Z-1074-2026 · 22 units
- Allura Xper FD20; Model Number: 722028;Z-1075-2026 · 2 units
- Allura Xper FD20/10; Model Number: 722029;Z-1076-2026 · 4 units
- Allura Xper FD20/20; Model Number: 722038;Z-1077-2026 · 8 units