Moderate riskFDA DeviceClass II

Philips Allura Xper FD20 X-ray system recalled for missing coolant drip tray

Philips identified that some Allura Xper FD20 X-ray systems may be missing a drip tray beneath the cooling unit due to incomplete previous service work. If the tray is not installed, coolant can leak onto electrical components and cause a short circuit, potentially shutting down the system.

From the FDA notice · Read the official notice ↗ · All Philips recalls

What to do

  1. Check if your system's serial number matches the recall list (79 affected units worldwide)
  2. Verify that the drip tray is properly installed beneath the cooling unit
  3. Contact Philips Medical Systems if you are unsure or if your system is on the recall list
  4. Do not operate the system if the drip tray is missing or damaged

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Allura Xper FD20; Model Number: 722012;
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 16 products — see all of them.

Details

Units affected
79 units
Illnesses reported
None reported
Manufactured in
Veenpluis 4-6, Best, Netherlands
Classification
Class II
Recall number
Z-1070-2026

In the FDA's words

Description
Allura Xper FD20; Model Number: 722012; Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown. Model Number: 722012; System Product Name: Allura Xper FD20; UDI-DI: 00884838059054; Serial Numbers: 2942, 1515, 1189, 1594, 535, 1964, 2156, 2742, 2092, 1673, 1803, 1655, 2020, 1973, 2037, 1853, 2098, 1248, 1082, 1576, 1057, 1129, 646, 930, 1421, 462, 203, 627, 2005, 1428, 353, 288, 191, 1994, 1802, 156, 853, 1283, 1956, 2361, 1980, 872, 931, 1295, 2513, 1090, 1374, 468, 1822, 812, 1379, 1817, 1580, 1050, 309, 144, 1841, 1845, 1468, 558, 1968, 1959, 2168, 1168, 1562, 967, 2114, 47, 167, 2091, 2053, 604, 603, 217, 2918, 373, 850, 1820, 1821, 1619; Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen. Affected quantity: 79 units
Hazard
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

16 products in this recall

Philips Allura Xper X-ray systems recalled for coolant short-circuit risk

Philips is recalling 16 Allura Xper imaging system models because a drip tray under the cooling unit may not have been installed during earlier field corrections, or the installation was not properly documented. Without the tray, coolant could reach electrical components, possibly causing short-circuits, blown fuses, or system shutdown. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

More from Philips

See all

Related topics

Electrical shockFireOtherMedical devicesSold nationwideMore from FDA Device