Philips Allura Xper FD20/10 X-ray system recalled for missing cooling drip tray
Check your Philips Allura Xper FD20/10 system
Philips identified that a small number of Allura Xper FD20/10 X-ray systems may be missing a drip tray that should be installed beneath the cooling unit. Without this tray, coolant liquid could contact electrical components and cause short-circuits, potentially shutting down the system.
- Check if your system has a drip tray installed beneath the cooling unit
- Contact Philips if you are unsure whether the tray is present or was previously installed
- Do not operate the system if the drip tray is confirmed missing
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1076-2026More PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls
- FDA-Device2026-06-03Philips Azurion imaging systems recalled for hard drive failure risk affecting imaging functionality
- FDA-Device2026-06-03Philips Allura medical imaging systems recalled for hard drive failure risk
- FDA-Device2026-05-27Philips Azurion X-ray systems recalled for cooling hose degradation risk
- FDA-Device2026-05-27Philips Integris-Allura X-ray systems recalled for cooling unit oil leakage risk
- FDA-Device2026-05-27Philips Azurion X-ray systems recalled for impaired table movement due to control wear
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief