Serious riskFDA DeviceClass II

RaySearch RayStation radiation therapy software recalled for dose calculation error

RaySearch RayStation/RayPlan radiation therapy planning software (versions 11B through 11B SPT1) has a bug that prevents dose calculations from updating correctly when changes are made to certain regions of interest with material overrides or specific geometry types. This could result in treatment plans using outdated radiation dose calculations.

From the FDA notice · Read the official notice ↗ · All Raysearch Laboratories recalls

What to do

  1. If your facility uses affected RayStation software versions, contact RaySearch Laboratories immediately
  2. Do not rely on dose calculations for regions of interest without contours until you receive a software update
  3. Verify your software version and check the distribution list to confirm if your facility is affected

Is yours affected?

Compare your label or packaging with the recalled product.

Product
RayStation/RayPlan. Model/Catalog Numbers: 12.0.0, 12.1.0, 12.0.3, 12.0.4, 12.3.0. Software Version: RayStation 11B, RayStation 11B SP1, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SPT1. Radiation Therapy Treatment Planning System
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 9 products — see all of them.

Details

Units affected
22 units
Illnesses reported
None reported
Manufactured in
Eugeniavagen 18c, Stockholm, Sweden
Classification
Class II
Recall number
Z-1104-2026

In the FDA's words

Description
RayStation/RayPlan. Model/Catalog Numbers: 12.0.0, 12.1.0, 12.0.3, 12.0.4, 12.3.0. Software Version: RayStation 11B, RayStation 11B SP1, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SPT1. Radiation Therapy Treatment Planning System Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended. UDI: 0735000201042620211208, 0735000201049520220312, 0735000201050120220422, 0735000201060020220620 and 0735000201057020221222. GTIN: 07350002010426, 07350002010495, 07350002010501, 07350002010600 and 07350002010570. Serial Numbers: 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12, 12.3.0.119. Software Revisions: RayStation 11B, 11B SP1, 11B SP2, 11B SP3, 11B SPT1. Expiration Date: 2028-03-20. Distribution: Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS Affected quantity: 22 units
Hazard
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

9 products in this recall

RayStation treatment planning software recalled for dose invalidation error

RaySearch Laboratories is recalling nine RayStation and RayPlan radiation therapy treatment planning software entries. Certain regions of interest with no contours, which have a material override or are of type Bolus, Fixation, or Support, may not invalidate calculated radiation dose as intended when their geometry is changed or their material override is removed. No injuries have been reported.

The agency lists each product separately. Check every one you might have.