FDA-Device2026-01-28Class II

RayStation/RayPlan. Model/Catalog Numbers: 12.0.0, 12.1.0, 12.0.3, 12.0.4, 12.3.0. Software Version: RayStation 11B, RayStation 11B SP1, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SPT1. Radiation Therapy Treatment Planning System

RAYSEARCH LABORATORIES AB
Hazard

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Sold states
Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS
Affected count
22 units
Manufactured in
Eugeniavagen 18c, Stockholm, Sweden
Products
RayStation/RayPlan. Model/Catalog Numbers: 12.0.0, 12.1.0, 12.0.3, 12.0.4, 12.3.0. Software Version: RayStation 11B, RayStation 11B SP1, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SPT1. Radiation Therapy Treatment Planning System

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1104-2026