Moderate riskFDA DeviceClass II

RayStation radiation therapy planning software recalled for dose calculation error

RayStation versions 17.0.0 and 17.0.1 have a software error that prevents dose calculations from updating properly when certain regions of interest are modified. This can occur when regions have no contours but include material overrides or are marked as Bolus, Fixation, or Support types. Users may not see updated radiation doses after making changes.

From the FDA notice · Read the official notice ↗

What to do

  1. Contact RaySearch Laboratories for a software update or patch
  2. Verify your RayStation version matches the affected versions (17.0.0, 17.0.1, RayStation v2025, or RayStation v2025 SP1)
  3. Review recent treatment plans using the affected software for accuracy

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025, RayStation v2025 SP1 Product Description: Radiation Therapy Treatment Planning System Component: No
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 9 products — see all of them.

Details

Units affected
0
Illnesses reported
None reported
Manufactured in
Eugeniavagen 18c, Stockholm, Sweden
Classification
Class II
Recall number
Z-1110-2026

In the FDA's words

Description
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025, RayStation v2025 SP1 Product Description: Radiation Therapy Treatment Planning System Component: No Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended. Lot Code: UDI: 0735000201082220250423, 0735000201094520250702 GTIN: 07350002010822, 07350002010945 Serial Numbers: 17.0.0.1270, 17.0.1.113 Software Revision: RayStation v2025, RayStation v2025 SP1 Expiration Date: 2030-10-22 Distribution: Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS Affected quantity: 0
Hazard
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

9 products in this recall

RayStation treatment planning software recalled for dose invalidation error

RaySearch Laboratories is recalling nine RayStation and RayPlan radiation therapy treatment planning software entries. Certain regions of interest with no contours, which have a material override or are of type Bolus, Fixation, or Support, may not invalidate calculated radiation dose as intended when their geometry is changed or their material override is removed. No injuries have been reported.

The agency lists each product separately. Check every one you might have.