RayStation 2024B radiation therapy software recalled for dose calculation error
Update RayStation 2024B software immediately
RayStation version 2024B has a software issue where calculated radiation doses may not update correctly when certain regions of interest (ROIs) are modified. This affects ROIs without contours that have material overrides or are classified as Bolus, Fixation, or Support types. Clinics using this version should verify their dose calculations are accurate.
- Contact RaySearch Laboratories for a software update or patch
- Review recent treatment plans using version 2024B to ensure dose calculations are correct
- Do not proceed with treatment if dose recalculation appears incorrect
- Check your facility's distribution list to confirm if your system is affected
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.