Serious riskFDA DeviceClass II

RayStation 2024B radiation therapy software recalled for dose calculation error

RayStation version 2024B has a software issue where calculated radiation doses may not update correctly when certain regions of interest (ROIs) are modified. This affects ROIs without contours that have material overrides or are classified as Bolus, Fixation, or Support types. Clinics using this version should verify their dose calculations are accurate.

From the FDA notice · Read the official notice ↗ · All Raysearch Laboratories recalls

What to do

  1. Contact RaySearch Laboratories for a software update or patch
  2. Review recent treatment plans using version 2024B to ensure dose calculations are correct
  3. Do not proceed with treatment if dose recalculation appears incorrect
  4. Check your facility's distribution list to confirm if your system is affected

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 16.0.0 Software Version: RayStation 2024B Product Description: Radiation Therapy Treatment Planning System Component: No
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 9 products — see all of them.

Details

Units affected
0
Illnesses reported
None reported
Manufactured in
Eugeniavagen 18c, Stockholm, Sweden
Classification
Class II
Recall number
Z-1109-2026

In the FDA's words

Description
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 16.0.0 Software Version: RayStation 2024B Product Description: Radiation Therapy Treatment Planning System Component: No Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended. Lot Code: UDI: 0735000201077820240625 GTIN: 07350002010778 Serial Numbers: 16.0.0.847 Software Revision: RayStation 2024B Expiration Date: 2029-06-28 Distribution: Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS Affected quantity: 0
Hazard
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

9 products in this recall

RayStation treatment planning software recalled for dose invalidation error

RaySearch Laboratories is recalling nine RayStation and RayPlan radiation therapy treatment planning software entries. Certain regions of interest with no contours, which have a material override or are of type Bolus, Fixation, or Support, may not invalidate calculated radiation dose as intended when their geometry is changed or their material override is removed. No injuries have been reported.

The agency lists each product separately. Check every one you might have.