Moderate riskFDA DeviceClass II

RaySearch RayStation 10.1.1 radiation therapy software recalled for dose calculation error

RaySearch's RayStation radiation therapy planning software version 10B SP1 has a flaw where calculated radiation doses may not update correctly when certain regions of interest are modified. This affects regions with no contours that have material overrides or are classified as Bolus, Fixation, or Support types. Clinical staff may unknowingly rely on outdated dose calculations.

From the FDA notice · Read the official notice ↗

What to do

  1. Identify any RayStation version 10.1.1 systems in use at your facility
  2. Contact RaySearch for the software update or patch
  3. Verify dose calculations manually for affected regions of interest before treatment
  4. Update to the corrected software version as soon as possible

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 10.1.1 Software Version: RayStation 10B SP1 Product Description: Radiation Therapy Treatment Planning System Component: No
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 9 products — see all of them.

Details

Units affected
0
Illnesses reported
None reported
Manufactured in
Eugeniavagen 18c, Stockholm, Sweden
Classification
Class II
Recall number
Z-1107-2026

In the FDA's words

Description
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 10.1.1 Software Version: RayStation 10B SP1 Product Description: Radiation Therapy Treatment Planning System Component: No Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended. Lot Code: UDI: 0735000201047120220131 GTIN: 07350002010471 Serial Numbers: 10.1.1.54 Software Revision: RayStation 10B SP1 Expiration Date: 2027-02-01 Distribution: Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS Affected quantity: 0
Hazard
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

9 products in this recall

RayStation treatment planning software recalled for dose invalidation error

RaySearch Laboratories is recalling nine RayStation and RayPlan radiation therapy treatment planning software entries. Certain regions of interest with no contours, which have a material override or are of type Bolus, Fixation, or Support, may not invalidate calculated radiation dose as intended when their geometry is changed or their material override is removed. No injuries have been reported.

The agency lists each product separately. Check every one you might have.