Serious riskFDA DeviceClass I

Arrow International PSI kits recalled for potential sterility issues

Arrow International included injectable anesthetic and saline solutions from a supplier with quality issues in multiple central venous catheter kits and trays. The supplier's products may not be sterile or may not work as intended. If sterility is compromised, these injectable products could cause serious infection or other life-threatening complications.

From the FDA notice · Read the official notice ↗

What to do

  1. Stop using affected Arrow International PSI and MAC catheter kits immediately.
  2. Contact your hospital or clinic's medical device or procurement department to identify whether you have affected lots.
  3. Do not use the injectable lidocaine, bupivacaine, or saline from these kits.
  4. Return affected kits to Arrow International or your supplier for a replacement.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Description/REF: ACCESS TRAY/GU-04020; PSI KIT: 8.5 FR/AK-09803-CDC; PSI KIT: 9 FR/AK-09903-CDC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-09810-PCMH1; PSI KIT: 9 FR X 10 CM/ASK-09903-KH; PSI KIT: 9 FR X 10 CM/ASK-09903-PHP1; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-UPA; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-YNH; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09907-SL; PSI KIT: 8.5 FR X 4-1/8IN (10 CM)/NA-09803-CDC; PSI KIT: 9 FR X 4-1/8IN (10 CM)/NA-09903-CDC; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-ECL4; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-MGH3; MULTI-LUMEN/PSI KIT/AK-11142-SPCS; MAC KIT: 2-L 9 FR DISTAL X 11.5 CM/CDC-11242-1A; MULTI-LUMEN/PSI KIT/AK-11142-D; MULTILUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-11142-MGH4; MULTI-LUMEN/PSI KIT/NA-11242-CDC; MULTI-LUMEN/PSI KIT/AK-21142-SK; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-LMDTG; MULTI LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-RH; MULTILUMEN/PSI KIT 9 FR X 11.5 CM/ASK-21242-IMC; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/ASK-21242-LMDTG; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMD; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UU3; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-VCU; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-1A; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-X1A; MULTI-LUMEN/PSI KIT/AK-21242-CDC; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-FH4; MULTI LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-GH1; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-SUU; 2-L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-UPMC4; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UKL; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMF; MULTILUMEN/PSI KIT: 9 FR X 4 1/2 IN/ASK-21242-UR1; PSI KIT: 8.5 FR/AK-29803-CDC; PSI KIT: 9 FR/AK-29903-CDC; PSI KIT: 9 FR/AK-29907-CDC; PSI KIT: 8.5 FR X 4IN (10 CM)/ASK-09804-UN1; PSI KIT: 9 FR/ASK-09903-UM; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-LMDT; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-LMDTG; PSI KIT: 9 FR X 4 1/8 (10 CM)/ASK-29903-BID1; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-LMDTG; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-UMD; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-29903-UR1; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/CDC-29803-X1A; PSI KIT: 8.5 FR X 10 CM/ASK-29803-ECL; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-UPM3; PSI KIT: 9 FR X 10 CM/ASK-29903-ECL; PSI KIT: 9FR X 4IN (10 CM) ANTIMICROBIAL/ASK-29903-UPM3; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09907-TG1
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 35 products — see all of them.

Details

Units affected
426,041
Illnesses reported
None reported
Manufactured in
3015 Carrington Mill Blvd, Morrisville, NC, United States
Classification
Class I
Recall number
Z-2385-2026

In the FDA's words

Description
Description/REF: ACCESS TRAY/GU-04020; PSI KIT: 8.5 FR/AK-09803-CDC; PSI KIT: 9 FR/AK-09903-CDC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-09810-PCMH1; PSI KIT: 9 FR X 10 CM/ASK-09903-KH; PSI KIT: 9 FR X 10 CM/ASK-09903-PHP1; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-UPA; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-YNH; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09907-SL; PSI KIT: 8.5 FR X 4-1/8IN (10 CM)/NA-09803-CDC; PSI KIT: 9 FR X 4-1/8IN (10 CM)/NA-09903-CDC; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-ECL4; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-MGH3; MULTI-LUMEN/PSI KIT/AK-11142-SPCS; MAC KIT: 2-L 9 FR DISTAL X 11.5 CM/CDC-11242-1A; MULTI-LUMEN/PSI KIT/AK-11142-D; MULTILUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-11142-MGH4; MULTI-LUMEN/PSI KIT/NA-11242-CDC; MULTI-LUMEN/PSI KIT/AK-21142-SK; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-LMDTG; MULTI LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-RH; MULTILUMEN/PSI KIT 9 FR X 11.5 CM/ASK-21242-IMC; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/ASK-21242-LMDTG; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMD; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UU3; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-VCU; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-1A; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-X1A; MULTI-LUMEN/PSI KIT/AK-21242-CDC; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-FH4; MULTI LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-GH1; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-SUU; 2-L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-UPMC4; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UKL; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMF; MULTILUMEN/PSI KIT: 9 FR X 4 1/2 IN/ASK-21242-UR1; PSI KIT: 8.5 FR/AK-29803-CDC; PSI KIT: 9 FR/AK-29903-CDC; PSI KIT: 9 FR/AK-29907-CDC; PSI KIT: 8.5 FR X 4IN (10 CM)/ASK-09804-UN1; PSI KIT: 9 FR/ASK-09903-UM; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-LMDT; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-LMDTG; PSI KIT: 9 FR X 4 1/8 (10 CM)/ASK-29903-BID1; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-LMDTG; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-UMD; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-29903-UR1; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/CDC-29803-X1A; PSI KIT: 8.5 FR X 10 CM/ASK-29803-ECL; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-UPM3; PSI KIT: 9 FR X 10 CM/ASK-29903-ECL; PSI KIT: 9FR X 4IN (10 CM) ANTIMICROBIAL/ASK-29903-UPM3; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09907-TG1 Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. REF(Material)/UDI-DI/Lot(Batch): GU-04020/30801902118807, 10801902193275, 30801902118807, 30801902118807/33F24C0753, 33F24E0736, 33F24G0154, 33F24G0648, 33F24H0377, 33F25A0059, 33F25A0072, 33F25A0395, 33F25F0260, 33F25G0736, 33F25H0298, 33F25J0603, 33F26B0084, 33F26B0314, 33F26C0578, 33F24D0036; AK-09803-CDC/20801902121756, 10801902193275, 10801902117417/33F24B0126, 33F24B0127, 33F24C0021, 33F24D0021, 33F24E0072, 33F24G0338, 33F24G0587, 33F24J0506, 33F24K0105, 33F24L0722, 33F25A0468, 33F25C0107, 33F25C0953, 33F25E0054, 33F25E0921, 33F25G0086, 33F25H0023, 33F25J0828, 33F25K0255, 33F25K0364, 33F25L0328, 33F24A0239; AK-09903-CDC/10801902117417, 10801902193275, 20801902121916/33F24A0356, 33F24B0128, 33F24B0129,33F24C0022, 33F24D0022, 33F24E0074, 33F24E0075, 33F24G0340, 33F24G0588, 33F24J0507, 33F24K0107, 33F24L0676, 33F24L0759, 33F25A0470, 33F25C0108, 33F25C0730, 33F25E0056, 33F25E0923, 33F25H0025, 33F25J0471, 33F25K0031, 33F25L0237, 33F25M0920, 33F26A1053, 33F24A0386; ASK-09810-PCMH1/10801902193275/33F25K0051, 33F24J0109; ASK-09903-KH/10801902119350, 10801902193275/33F24D0713, 33F24K0385, 33F24L1060, 33F25D0340, 33F25E0778, 33F25H0567, 33F25M0163, 33F26A1407, 33F25E1184; ASK-09903-PHP1/10801902193275, 10801902115222/33F24J0115, 33F25H0497, 33F26A0410, 33F24F0625; ASK-09903-UPA/10801902193275, 10801902118193/33F24K0387, 33F24L0855, 33F24A0787; ASK-09903-YNH/10801902118193, 10801902193275, 10801902118209/33F24E0527, 33F24H0478, 33F24L0682, 33F25H0135, 33F25J0915, 33F24A0110; ASK-09907-SL/10801902118209, 10801902193275/33F24D0053, 33F24D0967, 33F24G0265, 33F25E0215, 33F25E0454, 33F25G0680, 33F25J0551, 33F25K0328, 33F25L0206, 33F26C0312, 33F25B0188; NA-09803-CDC/10801902119930, 10801902193275, 10801902119947/33F24C0756, 33F24E0739, 33F25A0342, 33F25J0615, 33F26B0028, 33F24A1327; NA-09903-CDC/10801902119947, 10801902193275, 10801902119954/33F24E0740, 33F24H0381, 33F24J0199, 33F24J0361, 33F24L0020, 33F25B0278, 33F25C0551, 33F25E0340, 33F25E0400, 33F25E0468, 33F25F0419, 33F25H0708, 33F26A1655, 33F26B0023, 33F24A0113; ASK-09903-ECL4/10801902193275, 10801902224344/33F26A1582, 33F26B0399, 33F25E0534; ASK-09903-MGH3/10801902193275, 10801902189841, 10801902212006/33F25C0264, 33F25G0338, 33F25J0360, 33F25K0130, 33F26B0126, 33F26C0004, 33F
Hazard
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

35 products in this recall

Arrow kits with lidocaine and bupivacaine recalled for sterility risk

Arrow is recalling 35 kits and sets that contain lidocaine, bupivacaine, or 0.9% sodium chloride from a supplier that was itself recalled for quality issues. If the sterility of these water-based injectable drugs is compromised, it could lead to serious or life-threatening injuries. No injuries have been reported, and the products were distributed nationwide.

The agency lists each product separately. Check every one you might have.