Medical device recalls
38685 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2023-02-08FDA-DeviceMedtronic Perfusion SystemsClass IIAffinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF CB811. Used in Cardiopulmonary bypass.
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
- 2023-02-08FDA-DeviceMedtronic Perfusion SystemsClass IIAffinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, OXY/CVR COMBO BB841 FUSION W/BALANCE 26L, REF BB841. Used in Cardiopulmonary bypass.
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
- 2023-02-08FDA-DeviceMedtronic Perfusion SystemsClass IIAffinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF BB811. Used in Cardiopulmonary bypass.
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
- 2023-02-08FDA-DeviceSynCardia Systems LLCClass IISynCardia Total Artificial Heart System (70CC): 500101-001 SynCardia Total Artificial Heart System (50CC): 570500-000
Due to tears reported tears in the cannula due to wear and tear stress.
- 2023-02-08FDA-DeviceSmiths Medical ASD Inc.Class ICADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24, 3) REF 21-7302-24, 4) REF 21-7308-24, 5) REF 21-7309-24, 6) REF 21-7310-24, 7) REF 21-7322-24, 8) REF 21-7323-24, 9) REF 21-7324-24, 10) REF 21-7333-24, 11) REF 21-7343-24, 12) REF 21-7346-24, 13) REF 21-7349-24, 14) REF 21-7359-24, 15) REF 21-7363-24, 16) REF 21-7383-24, 17) REF 21-7391-24, 18) REF 21-7394-24, 19) REF 21-7600-24, 20) REF 21-7609-24, 21) REF 21-7624-24, 22) REF 21-7649-24, 23) REF
There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms
- 2023-02-08FDA-DeviceSmiths Medical ASD Inc.Class ICADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d) REF 21-7649-24, e) REF 21-7600-24JP, f) REF 21-7609-24JP, g) REF 21-7624-24JP, h) REF 21-7649-24JP Computerized Ambulatory Delivery Device
There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms
- 2023-02-08FDA-DeviceSmiths Medical ASD Inc.Class ICADD Administration Set, List Numbers: a) REF 21-7308-24, b) REF 21-7309-24, c) REF 21-7310-24, d) REF 21-7308-24JP, e) REF 21-7309-24JP. Computerized Ambulatory Delivery Device
There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms
- 2023-02-08FDA-DeviceSmiths Medical ASD Inc.Class ICADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24 3) REF 21-7302-24, 4) REF 21-7321-24, 5) REF 21-7322-24, 6) REF 21-7323-24, 7) REF 21-7324-24, 8) REF 21-7333-24, 9) REF 21-7336-24, 10) REF 21-7339-24, 11) REF 21-7343-24, 12) REF 21-7349-24, 13) REF 21-7359-24, 14) REF 21-7383-24, 15) REF 21-7390-24, 16) REF 21-7391-24, 17) REF 21-7394-24, 18) REF 21-7395-24, 19) REF 21-7301-24JP, 20) REF 21-7302-24JP, 21) REF 21-7321-24JP, 22) REF 21-7322-24JP, 2
There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms
- 2023-02-08FDA-DeviceSmiths Medical ASD Inc.Class ICADD Administration Set, List Numbers: a) REF 21-7346-24; b) REF 21-7363-24
There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms
- 2023-02-08FDA-DeviceUniversal Meditech Inc.Class IProducts are labeled as: Skippack Medical Lab SARS-CoV-2 Antigen Rapid Test Kit (Colloidal Gold) 20 TESTS/1 KIT Green and white box under the brand DiagnosUS, and, White box without brand name labeled as "SARS -COV-2 ANTIGEN DEVELOPMENT MADE IN USA ORAL SWAB TESTING
Distribution of SARS-CoV-2 Antigen Rapid Test Kits without FDA market approval, clearance or Emergency Use Authorization.
- 2023-02-01FDA-DeviceBiokit, S.A.Class IIbiokitHSV-2 Rapid Test, REF: 300028001
HSV-2 rapid test may report false positive results.
- 2023-02-01FDA-DeviceSunMed Holdings, LLCClass IIAirQ3 Self Pressurizing ILA with Automatic Cuff Inflation and GA Size REF Number 0.0 60005 0.5 60055 1.0 60105 1.5 60155 2.0 60205 3.0 60305 4.0 60405
There is an orogastric (OG) tube size discrepancy between the labeling in the IFU and the printed text on the device. If the clinician is unable to pass a gastric catheter through one of the channels, the clinician may use a smaller-diameter gastric catheter.
- 2023-02-01FDA-DeviceSunMed Holdings, LLCClass IIAir-Q3 AirQ3 Standard ILA with Manual Cuff Inflation and GA Size REF Number 0.0 50005 0.5 50055 1.0 50105 1.5 50155 2.0 50205 3.0 50305 4.0 50405
There is an orogastric (OG) tube size discrepancy between the labeling in the IFU and the printed text on the device. If the clinician is unable to pass a gastric catheter through one of the channels, the clinician may use a smaller-diameter gastric catheter.
- 2023-02-01FDA-DeviceBiomet, Inc.Class IIOxford Partial Knee System, Fixed Lateral Tibial Construct- Intended to replace part of a knee joint when used in conjunction with a femoral and tibial component Item Number: 154341
Packaged in the incorrect shelf carton. Incorrect shelf carton; outer carton is smaller than the required outer carton, which forces the Tyvex pouch to be folded and packaged in the carton causing compound creases and compromise the sterile barrier
- 2023-02-01FDA-DeviceAlign Technology IncClass IIInvisalign Express 10 REF 8512 Invisalign System Comprehensive REF 9000 Invisalign System Moderate REF 9001
3D orthodontic planning software has a defect that leads to an issue where incorrect number of aligners may be produced (less or greater than the doctor requested) and incorrect packaging is provided.
- 2023-02-01FDA-DeviceFuture Diagnostics Solutions B.V.Class IIFuture Diagnostics STAT-IntraOperative-Intact-PTH (STAT-IO-I-PTH) Immunoassay kit REF 4K-IPT-00. in vitro diagnostic.
Due to an issue of high coefficient of variability (CV) problem caused by the accuspheres inside the Immunoassay kits.
- 2023-02-01FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IICENTURION MEDICAL PRODUCTS, SILVER ANTIMICROBIAL VAD DRIVE, REF DT18760
The kit contained an expired component.
- 2023-02-01FDA-DeviceCooperVision, Inc.Class IIMyDay Toric (90-pack) Reveal 1 Day Toric (90-pack) Soft (hydrophilic) Contact Lens (daily wear). Single Use.
Lenses may be of the incorrect power. A patient who uses an affected lens may experience poor visual acuity.
- 2023-02-01FDA-DeviceBaxter Healthcare CorporationClass IIBaxter Electrocardiograph, ELI 380 -DCS21
One ELI 380 Electrocardiograph intended to be used for a pilot program was shipped to a customer by mistake on 26 October 2022.
- 2023-02-01FDA-DeviceTELEFLEX LLCClass IIEndotracheal Tube labeled as the following: a. Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff; b. Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff; c. Flexiset Super Safety Clear Tracheal Tube, oral/nasal with Cuff and Insertion Aid; d. RUSCHELIT Safety Clear Tracheal Tube, oral/nasal, Magill; e. RUSCHELIT Safety Clear Tracheal Tube, oral/nasal, Murphy. Intended for patients requiring endotracheal intubation
Potential for pilot balloon non-inflation or cuff non-deflation.
- 2023-02-01FDA-DeviceTELEFLEX LLCClass IIEndotracheal Tube labeled as the following: a. Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal; b. Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff; c. RUSCHELIT Super Safety Clear Tracheal Tube, oral/nasal, Murphy. Intended for patients requiring endotracheal intubation
Potential for pilot balloon non-inflation or cuff non-deflation.
- 2023-02-01FDA-DeviceTELEFLEX LLCClass IISlick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal. Intended for patients requiring endotracheal intubation
Potential for pilot balloon non-inflation or cuff non-deflation.
- 2023-02-01FDA-DeviceAbbott Molecular, Inc.Class IIAlinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing, amplification, detection, and data reduction.
The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration data file) referenced causing a misconfiguration of the calibration. 2. Inability of Maintenance and Diagnostics 2300 (Amp Detect Optical Calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. Under specific conditions during the ADU manufacturing process, an incorrect calibration may potentially have been configured.
- 2023-02-01FDA-DeviceMeridian Bioscience IncClass IIRevogene SARS-CoV-2_IVD real-time RT-PCR test intended for the qualitative detection of RNA from SARSCoV-2 Catalog Number: 410700
Leakage of the original specimen from the plastic microfluidic cartridge (PIE), possibility that laboratory personnel may be exposed to aerosolized specimen containing SARS-CoV-2
- 2023-02-01FDA-DeviceJohnson & Johnson Surgical Vision, Inc.Class IISingle-Use Packs, REF: OPO80
Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.
- 2023-02-01FDA-DeviceJohnson & Johnson Surgical Vision, Inc.Class IIFUSION Dual Pump Packs, REF: OPO73
Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.
- 2023-02-01FDA-DeviceJohnson & Johnson Surgical Vision, Inc.Class IIVeritas Advanced Infusion Packs, REF: VRT-AI
Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.
- 2023-02-01FDA-DeviceAgfa Healthcare NVClass IIAgfa HealthCare Enterprise Imaging XERO Viewer
There is a software defect that can cause issues with images.
- 2023-02-01FDA-DeviceSiemens Medical Solutions USA, IncClass IISoftware : myNeedle Guide 2D license, myNeedle Guide 2D AWP only Workflow license, or myNeedle Guide 3D license. Used on the following CT (Computed tomography systems): SOMATOM go.Now Model 11061610 NA* SOMATOM go.Now Model 11061618 NA* SOMATOM go.Up Model 11061620 SOMATOM go.Up Model 11061628 NA* SOMATOM go.All Model 11061630 SOMATOM go.All Model 11061638 NA* SOMATOM go.Top Model 11061640 SOMATOM go.Top Model 11061648 NA* SOMATOM X.cite Model 11330001 SOMATOME X.
In the syngo.CT VA40 and syngo.CT VA50 myNeedle Guide application, the order of the orientation labels head , center, and feet in the axial thick slice segments are erroneously swapped
- 2023-02-01FDA-DeviceGetinge Usa Sales IncClass IGetinge Flow-e Anesthesia System, Part No. 6887900
Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.