Medical device recalls
38685 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2022-07-06FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation 11B
This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.
- 2022-07-06FDA-DeviceHaemonetics CorporationClass IIProduct Name: TEG5000 Analyzer (07-022, 07-033) with TEG Analytical Software (07-030, 07-031) and TEG Platelet Mapping Assay - PlateletMapping ADP & AA (07-014), PlateletMapping ADP (07-015), PlateletMapping AA (07-016). Issue occurs when Platelet Mapping Assay results from a TEG 5000 analyzer are viewed using TEG Manager Software (versions 1.1-4.3.1)
When the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or AA) Assays and TEG Mangement Software, incorrect values of the PlateletMapping ADP or AA % Inhibition and % Aggretagion may be displayed on TEG Manager for the affected assays.
- 2022-07-06FDA-DeviceAtrium Medical CorporationClass IIAtrium Firm PVC Thoracic Catheter 28French-Intended to facilitate the evacuation of air and/or fluid from the chest cavity or mediastinum. Product REF Number: 15028
Mismatch between the expiration date on the sterile barrier pouch and the expiration date on the case/shipper label. Product sterile barrier pouch label expiration date of 16SEP2024 (2024-09-16) while the case and shipper labels both showed 21SEP2024 (2024-09-21), causing product being used for up to five (5) days past the true expiration date
- 2022-07-06FDA-DeviceOlympus Corporation of the AmericasClass IISOLTIVE SuperPulsed Laser Fibers, Model# Labeled TFL-FBX150BS (Carton) - 150 Micron Ball Tip Single Use Fiber Model # Labeled TFL-FBX200S (Pouch) - 200 Micron Single Use Fiber Model # Labeled TFL-FBX200BS (Carton) - 200 Micron Ball Tip Single Use Fiber Model # Labeled TFL-FBX550S (Carton) - 550 Micron Single Use Fiber Model # Labeled TFL-FBX200S (Carton) - 200 Micron Single Use Fiber Model # Labeled TFL-FBX365S (Pouch) - 200 Micron Single Use Fiber
Device model and size for affected lots of Laser Fibers may be mismatched between the product's carton, immediate pouch packaging, and/or the device.
- 2022-07-06FDA-DeviceROi CPS LLCClass IIregard Item Number: 800774001, CV00979A - CV Minor Kit, Sterile.
A nonsterile component was packaged in a sterile convenience kit.
- 2022-07-06FDA-DeviceSiemens Medical Solutions USA, IncClass IIArtis Pheno, Interventional Fluoroscopic X-Ray System, Model No. 10849000
A video signal from one of the transceivers of the Large Display may be lost. In very rare cases, both transceivers of the Large Display could potentially fail which can lead to the loss of both video signals. This issue can sporadically affect the Large Display in the examination room as well as the displays in the control room.
- 2022-07-06FDA-DeviceC.R. Bard IncClass IISensica Urine Output System
Complaints relating to urine output measurement accuracy.
- 2022-07-06FDA-DeviceDain Technology, Inc. (Subsidiary of Soundable Health, Inc.)Class IIProudP Everyday Uroflow Tracker Software Version 1.x.x Software Version 1.2.0 or earlier
Due to interference with the Live Listen feature of hearing aid or AirPods, the user's iPhone may perform automatic processing of the urination sound signal, resulting in lower urination volume and velocity values than expected.
- 2022-07-06FDA-DeviceWright Medical Technology, Inc.Class IIStryker INFINITY Resect Guide for INBONE Talus, Size 4, Model #33620254, non-sterile, an instrument used with the INFINITY Total Ankle System.
The instrument is incorrectly color coded.
- 2022-07-06FDA-DeviceARROW INTERNATIONAL Inc.Class IIArrow Three-Lumen CVC
Gravity flow rates on the affected product lidstock are incorrect.
- 2022-07-06FDA-DeviceCanon Medical System, USA, INC.Class IISystem: Alphenix 4D CT in combination with CAS-930A (C arm Support) Model Number: INFX-8000C/BF, INFX-8000C/BH, INFX-8000C/SW
CT operation may be restricted by an interlock which is a result of an system error and the CT system may stop operating properly after an attempt to cancel the error message has been initiated.
- 2022-07-06FDA-DeviceBaxter Healthcare CorporationClass II#1 MAC F/O LARYNGOSCOPE; F/O Laryngoscope Set-MAC; F/O Laryngoscope Set-MAC W/LED
Packaging error.
- 2022-07-06FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIMedline Pill Splitters; 1 each/box; 144 boxes/carton
Product lid (which houses a blade) is coming unhinged from the base, thus exposing the blade.
- 2022-07-06FDA-DeviceGE Healthcare, LLCClass IIField replacement back up batteries distributed on or after April 1, 2019 for Avance CS2, Avance CS2 Pro, Avance, Amingo, Aespire View anesthesia devices
Backup batteries can fail earlier than their estimated life.
- 2022-07-06FDA-DeviceGE Healthcare, LLCClass IIBack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019
Backup batteries can fail earlier than their estimated life.
- 2022-07-06FDA-DeviceMicrogenics CorporationClass IIQMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3 QMS TACROLIMUS CAL D 10015569-3 QMS TACROLIMUS CAL E 10015570-3 QMS TACROLIMUS CAL F 10015571-3
Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.
- 2022-06-29FDA-DeviceCardioTek BVClass IICardioTek EP-TRACER Software V2.x.
Device did not pass electrical safety testing for adequate insulation.
- 2022-06-29FDA-DeviceMaquet Cardiovascular, LLCClass IIINTERGARD Woven Vascular Graft is a woven polyester graft coated with a bovine collagen. It has an external velour surface and non-velour inner surface. The device is available in straight tubes and bifurcated configurations, as well as hemabridge, aortic arch, thoracic aortic graft, and cardioroot configurations designed for use on the thoracic aorta
Mislabeled: Product labeled as Intergard Woven Straight graft REF IGW0032-30, D: 8 mm, L: 30 cm, contained an Intergard Woven Straight graft of D: 32 mm and L: 30 cm
- 2022-06-29FDA-DeviceBeaver Visitec International, Inc.Class IIBVI Wet -Field Eraser, 18GA Blunt Tip, 45Deg.-Intended to remove tissue and control bleeding by the use of high frequency electrical current Part Number: 221251
Boxes labeled as Wet-Field Eraser, 18GA Blunt Tip, 45Deg., were shipped containing incorrect product a Wet-Field Eraser, 18GA Bevel Tip, Straight, Part Number: 221250
- 2022-06-29FDA-DeviceNew Standard Device IncClass IIMetalogix Slotted 90 Tubular Wrench x 10mm REF 900104 LOT A29618A
Due to manufacturing processing error, wrench causing stripping when torqueing nuts.
- 2022-06-29FDA-DeviceGetinge Usa Sales IncClass IFlow-e Anesthesia System : Intended for use in administering anesthesia while controlling the entire ventilation of patient Product Code/REF Number: 6887900
Potential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction unit from being activated.
- 2022-06-29FDA-DeviceGetinge Usa Sales IncClass IFlow-c Anesthesia System : intended for use in administering anesthesia while controlling the entire ventilation of patient Product Code/REF Number: 6887700
Potential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction unit from being activated.
- 2022-06-29FDA-DeviceOlympus Corporation of the AmericasClass IIEVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model: GIF-HQ190
Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190 which would not meet the intended design specifications
- 2022-06-29FDA-DeviceOlympus Corporation of the AmericasClass IIColonovideoscope Model CF-H180AL
A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out (approximately 1.4 mm) further than normal protrusion,could lead to patient injury; specifically, mucosal damage that could result in hemorrhage or bleeding
- 2022-06-29FDA-DeviceBard Peripheral Vascular IncClass IIREF: PIG1260K, 6Fr, X10, Safe-t-Centesis Catheter drainage Kit, Sterile EO,
There is the potential that a 6 Fr Catheter Drainage kit contains an 8 Fr needle instead of a 6 Fr needle. They can be used in conjunction with the following drainage methods: Universal extension set and drainage bag, vacuum bottle, and wall suction.
- 2022-06-29FDA-DeviceSmiths Medical ASD Inc.Class IMedfusion Syringe Pump Models: 3500, 3500-0600-00, 3500-0600-01, 3500-0600-249, 3500-0600-50, 3500-0600-51, 3500-0600-82, 3500-306, 3500-402, 3500-414, 3500-415, 3500-500, 3500BC, 3500E, 3500G, 3500SD, 3500SD-500, 3500VX, 3500VX-306, 3500VX-414, 3500VX-415, 3500VX-500, 3500ZE, Software Versions 6.0.0, 6.0.1
Multiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) - Infusion Continues after System Failure, 5. Clearing of Program Volume Delivered (PVD), 6. False Alarm for Rate Below Recommended Minimum for Syringe Size, 7. Incorrect Bolus or Loading Dose Time Display, 8. Domain Name Server (DNS) Port 1001
- 2022-06-29FDA-DeviceSmiths Medical ASD Inc.Class IMedfusion Syringe Pump Model: Model 4000-0100-50, Model 4000-0101-249, Model 4000-0101-50, Model 4000-0101-51, Model 4000-0101-78, Model 4000-0105-249, Model 4000-0105-51, Model 4000-0105-78, Model 4000-0106-00, Model 4000-0106-01, Model 4000-0106-231
Multiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) - Infusion Continues after System Failure, 5. Clearing of Program Volume Delivered (PVD), 6. False Alarm for Rate Below Recommended Minimum for Syringe Size, 7. Incorrect Bolus or Loading Dose Time Display, 8. Domain Name Server (DNS) Port 1001
- 2022-06-29FDA-DeviceAlcon Laboratories Ireland, LtdClass IIAlcon Clareon IOL with AutonoMe Delivery System
Due to device (preloaded intraocular lens) mislabeling, associated with partial mixing of product during overlapping manufacturing runs, and attributed to improper manufacturing line clearance.
- 2022-06-29FDA-DeviceMedtronic IncClass IHeartWare Battery, Model #1650DE, a component of the HeartWare Ventricular Assist Device (HVAD) System.
Batteries for the HVAD system have a weld defect affecting internal components within the HVAD battery, resulting in the battery to malfunction and no longer provide power or prevent the battery from holding a complete charge or properly recharging.
- 2022-06-22FDA-DeviceVolcano CorpClass IIPhilips OmniWire Pressure guide wire REF 89185J PN 300000252891
Due to a potential failed sterilization process.