- 2021-06-09FDA-DeviceDraeger Medical, Inc.Class IICritical Care Ventilator, Catalog Number(s): 8422400: Babylog VN800 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons.
Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning
- 2021-06-09FDA-DeviceDraeger Medical, Inc.Class IICritical Care Ventilator, Catalog Number(s): 8422200: Babylog VN600 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons.
Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning
- 2021-06-09FDA-DeviceDraeger Medical, Inc.Class IICritical Care Ventilator, Catalog Number(s): 8422500: Evita V800 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons.
Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning
- 2021-06-09FDA-DeviceDraeger Medical, Inc.Class IICritical Care Ventilator, Catalog Number(s): 8422300: Evita V600 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons.
Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning
- 2021-06-09FDA-DeviceVygon U.S.A.Class IIVygon Dressing Change w/ Maxiswab - Product Usage: Custom Dressing Change kit, Product Code: AMS-8465CS.
Compromised sterile barriers due to the Tyvek Lid not properly adhering to the Tray and has the the potential to cause an infection
- 2021-06-09FDA-DeviceSiemens Medical Solutions USA, IncClass IIMultitom RAX with software version VF10
Siemens identified two issues, In some cases, it is possible to perform an image acquisition although more than 20% of the beam does not hit the detector. This may only occur when an organ program for the wall mode without top alignment is selected and the system is moved to the centred position. If afterwards the operator activates the top alignment, increases the collimation, and lifts the tube, the system allows examination even with the tube being misaligned to the detector position. Hence, it is possible to overshoot the detector by more than 20%. From a clinical point of view, it is highly unlikely for the operator to trigger the examination as the misalignment is visible due to the light field being out of range. AND In very rare cases it may occur that the calculated dose value exceeds the limit of 2 Gy. This may potentially occur only when an organ program for free exposure is selected, and the system uses incorrect (too small) source-to-image distance (SID) for calculating applied dose. Hence, the calculated dose is much higher than the actual applied dose causing the buzzer, which normally notifies the operator about the exceeded limit of 2 Gy, to get activated erroneously.
- 2021-06-09FDA-DeviceThe Biomed GuysClass IAlaris Infusion Pump Module, Model 8100, serviced/repaired by The Biomed Guys, with the impacted bezel repair part, TIPA-8100-4410 / Alaris 8100 Bezel.
Bezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.
- 2021-06-09FDA-DeviceLEPU MEDICAL TECHNOLOGY (BEIJING) CO., LTDClass ILeccurate SARS-CoV-2, Antigen Rapid Test Kit (Colloidal Gold Immunochromatography), 1 test, IVD. for the qualitative detection of antigen SARS-CoV-2 in clinical samples (nasal swab).
Antigen Rapid Test Kit is not authorized, cleared, or approved for marketing and/or distribution in the U.S.
- 2021-06-09FDA-DeviceLEPU MEDICAL TECHNOLOGY (BEIJING) CO., LTDClass ILeccurate SARS-CoV-2, Antibody Rapid Test Kit (Colloidal Gold Immunochromatography), 1 test, IVD. for qualitative detection of IgM and/or IgG antibodies to SARS-CoV-2 in human serum, plasma and venipuncture whole blood for use as an aid in identifying individuals with an adaptive immune response to SARS-CoV-2, indicating recent or prior infection.
Leccurate SARS-CoV-2, Antibody Rapid Test Kit (Colloidal Gold Immunochromatography) is not authorized, cleared, or approved for marketing and/or distribution in the U.S.
- 2021-06-02FDA-DeviceOrtho-Clinical Diagnostics, INc.Class IIVITROS CoV2 Ag calibrator (Product Code 6199950) - Product Usage: is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) and anterior nasal swab specimens collected in CDCs formulation of VTM.
Ortho investigated and confirmed the potential for unexpected failed calibration, and condition codes associated with the calibration, when using VITROS SARS-CoV-2 Ag Reagent Pack along with VITROS Immunodiagnostic Products SARS-CoV-2 Ag Calibrator, Lot 20.
- 2021-06-02FDA-DeviceJames Leckey Design LtdClass IIMyWay Walker Chassis - Product Usage: is intended for use by children already using a supportive walking device.
Due to a failure of the welds, the casters detached from the rear legs of the walker.
- 2021-06-02FDA-DeviceCooperSurgical, Inc.Class IIThe Ultem Polymeric Resin Koh Cup is a non- sterile, reusable vaginal fornices delineator. It is made of medical grade plastics and not made with natural rubber latex. Model Number(s): KCP-30-2
The affected Koh Cup were incorrectly packaged with a 3.5cm Ultem Koh cup, P/N 34131-04, instead of a 3.0cm Ultem Koh cup, P/N 34131-03 as required.
- 2021-06-02FDA-DeviceBaxter Healthcare CorporationClass IIGambro Cartridge Blood Set, Product Code 101025 - Product Usage: intended for single use in a hemodialysis treatment using the Phoenix Dialysis Delivery Systems.
There is a potential leak between the venous patient connector and patient's hemodialysis access.
- 2021-06-02FDA-DeviceRadiometer Medical ApSClass IIPIC070 Arterial Blood Sampler PICO70 - Product Usage: Preheparinized, electrolyte-balanced, arterial blood sampler product line for collection of arterial samples for pH, blood gas, oximetry, electrolyte and metabolite analysis. Model Number/Product Description: 956-518 PICO70 w/o needle; 956-519 PICO70 w/o needle; 956-522 PICO70 22G x 25 mm; 956-525 PICO70 22G x 32 mm; 956-529 PICO70 23G x 16 mm; 956-533 PICO70 23G x 25 mm; 956-534 PICO70 23G x 32 mm; 956-547 PICO70 25G x 16 mm;
Radiometer has become aware of an issue with PICO70 Arterial Blood Sampler. The issue relates to the sterile barrier system for the product. During internal stress testing on PICO70 Arterial Blood Samplers a breach in the sterile barrier system was observed in one sample. Risk for the patient: The described error is considered to have a remote possibility of resulting in bacterial bloodstream infection. The bloodstream infection may be asymptomatic but may also progress to sepsis or life-threatening septic shock. Immunosuppressed patients are at particular risk. Due to issues (breaches) with the sterile barrier system for the Arterial Blood Sampler.
- 2021-06-02FDA-DevicePentax of America IncClass IIPentax Medical Video Duodenoscope - Product Usage: intended to be used with endoscopic devices and other ancillary equipment to provide optical visualization of (via a video monitor), and therapeutic access to, Biliary Tract via the Upper Gastrointestinal Tract, Model Number: ED34-i10T2.
IFU updated: The warning section of the IFU for both the OE-A63 distal end cap and ED34-i10T2 duodenoscope has been updated to notify users of the associated risks with the distal end cap (OE-A63) unexpectedly becoming detached during a procedure and updated to notify users of what immediate actions should be taken in case the event occurs. This can result in unforseen events such as mucosal injury, lacerations, or bleeding of the patient. Detachment of the distal end cap (OE-A63) into the oral cavity of the patient may also result in aspiration
- 2021-06-02FDA-DeviceNobel Biocare Procera LLCClass IINPr FCZ¿Imp Crwn ASC CC NP CAD 1-25¿ s6 - Product Usage: Implant Crown intended to be placed in the upper or lower jaw creating a prosthetic device (full contour crown), and to restore the patient s chewing function and esthetics. Material Number: 77387
Not manufactured to specification
- 2021-06-02FDA-DeviceNobel Biocare Procera LLCClass IINPr FCZ¿Imp Crwn ASC CC RP CAD 1-25¿ s4 - Product Usage: Implant Crown intended to be placed in the upper or lower jaw creating a prosthetic device (full contour crown), and to restore the patient s chewing function and esthetics. Material Number: 77364
Not manufactured to specification
- 2021-06-02FDA-DeviceNobel Biocare Procera LLCClass IINPr FCZ¿Imp Crwn ASC CC NP CAD 1-25¿ s4 - Product Usage: Implant Crown intended to be placed in the upper or lower jaw creating a prosthetic device (full contour crown), and to restore the patient s chewing function and esthetics. Material Number: 77363
Not manufactured to specification
- 2021-06-02FDA-DeviceNobel Biocare Procera LLCClass IINPr FCZ¿Imp Crwn ASC CC RP CAD 1-25¿s3 - Product Usage: Implant Crown intended to be placed in the upper or lower jaw creating a prosthetic device (full contour crown), and to restore the patient s chewing function and esthetics. Material Number: 77352
Not manufactured to specification
- 2021-06-02FDA-DeviceNobel Biocare Procera LLCClass IINPr FCZ¿Imp Crwn ASC CC RP CAD 1-25¿ s2 - Product Usage: Implant Crown intended to be placed in the upper or lower jaw creating a prosthetic device (full contour crown), and to restore the patient s chewing function and esthetics Material Number: 77339
Not manufactured to specification
- 2021-06-02FDA-DeviceDentech Corp.Class IIRubber Dam Clamp, KSK - Product Usage: used for prevention of contamination by saliva and moisture prevention during dental treatment.
Removing parts for safety reasons due to possible breakage.
- 2021-06-02FDA-DeviceRoche Diagnostics Operations, Inc.Class IIIRON GEN 2: ROCHE COBAS INTEGRA ANALYZER, REAGENT CASSETTES, Catalog#: 03183696122 - Product Usage: In vitro test for the quantitative determination of iron in human serum and plasma on Roche/Hitachi cobas c and cobas Integra systems.
Roche received several complaints regarding increased control recovery and discrepant elevated results for the Iron Gen.2 (IRON2) assay on the cobas c311, 501, and 502 analyzers and on the COBAS Integra 400 plus analyzer.
- 2021-06-02FDA-DeviceSiemens Healthcare Diagnostics, IncClass IICentaur Syphilis Assay-SYP Kit, 200 Test - Product Usage: For in vitro diagnostic use in the qualitative determination of antibodies to Treponema pallidum in human serum or plasma using the ADVIA Centaur XP/XPT System and ADVIA Centaur CP Systems, SMN:10492493
Potential for the ADVIA Centaur Syphilis assay to carryover into other commercial assays. This carryover effect only occurs when the impacted assays are used immediately following testing with the Syphilis assay on the ADVIA Centaur XP/XPT or ADVIA Centaur CP systems
- 2021-06-02FDA-DeviceWelch Allyn IncClass IIProBP 3400 Digital Blood Pressure Device ((handheld, desk mount, wall mount models) Model: 34XFHT-B Intended: ProBP 3400 automatically measures systolic and diastolic pressure (excluding neonates) and pulse rate, as well as calculates Mean Arterial Pressure (MAP).
Potential Risk of fluid ingress to the power supplies, resulting in risk of an internal electrical short and possible electrical shock
- 2021-06-02FDA-DeviceWelch Allyn IncClass IIGS777 Wall Transformer Model: 77710 Intended Use: The GS777 Wall Transformer is intended for use in ambulatory and acute care settings.
Potential Risk of fluid ingress to the power supplies, resulting in risk of an internal electrical short and possible electrical shock
- 2021-06-02FDA-DevicePhilips North America LlcClass IIIntelliVue MX40 Patient Monitor, Model Number: 865350 (US) 865352 (OUS) Software version number B.06.59 - Product Usage: Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals.
Manufacturing failed to follow process steps resulting in the distribution of potentially defective devices
- 2021-06-02FDA-DeviceBeckman Coulter Inc.Class IIAccess SARS-CoV-2 Antigen, REF C68668, UDI: (01)15099590742713, 3.30 mL/R1a, 4.70 mL/R1b, 4.20 mL/R1c, 3.10 mL, R1d, 3.10 mL, R1e
specimens collected in transport media may result is false negative results when stored in certain conditions.
- 2021-06-02FDA-DevicePhilips Respironics, Inc.Class IIBiPAP A40, Brazil: Continuous ventilator, non-life-supporting. Material Number: BR1111169
Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.
- 2021-06-02FDA-DevicePhilips Respironics, Inc.Class IIBiPAP A30, Brazil: Continuous ventilator, non-life-supporting. Material Number: BR1111143
Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.
- 2021-06-02FDA-DevicePhilips Respironics, Inc.Class IIBiPAP A40 System Silver, JP: Continuous ventilator, non-life-supporting. Material Number: 1111177
Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.