Medical device recalls
39137 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2021-05-26FDA-DeviceMedtronic Inc.Class IIMedtronic REF: MMT-7020GA, Guardian Sensor (3) for the following packaging: GS3 5PK US GC/ GTIN: 20763000179629 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
- 2021-05-26FDA-DeviceMedtronic Inc.Class IIMedtronic REF: MMT-7020B, Guardian Sensor (3) for the following packaging: GS3 1PK 2L US/GTIN: 00763000140779 GS3 1PK 2L US/ GTIN: 00763000337049 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
- 2021-05-26FDA-DeviceMedtronic Inc.Class IIMedtronic REF: MMT-7020A, Guardian Sensor (3) for the following packaging: GS3 5PK 2L US/ GTIN: 20763000140940 GS3 5PK 2L US/ GTIN: 20763000337005 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
- 2021-05-26FDA-DeviceMedtronic Inc.Class IIMedtronic REF: MMT-7008B, Enlite Glucose Sensor for the following packaging: ENLITE PLUS 23L 1CL / GTIN: 00643169541719 ENLITE PLUS 2L 1CL PR/ GTIN: 00643169786738 ENLITE PLUS 1PK 23L 1CL/ GTIN: 00763000421304 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
- 2021-05-26FDA-DeviceMedtronic Inc.Class IIMedtronic, REF: MMT-7008A, Enlite Glucose Sensor for the following packaging: ENLITE PLUS 5PK 23L 1CL /GTIN: 20643169541706, ENLT PLUS 5PK 2L 1CL PR/ GTIN: 20643169786725, ENLITE PLUS 5PK 23L 1CL/ GTIN: 20763000421292 - Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
- 2021-05-26FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation standalone software radiation therapy treatment planning system. Model Number: 8A, 8B, 9A, 9B, 10A, 10B including some service packs
When using more than one plan in a treatment course, the Total dose display in Dose tracking may be incorrect. There may be a mismatch between the fractions that appear as selected in the Treatment course list in Dose tracking and the fractions that are included in the dose summation. The issue only affects RayTreat users and only the Total dose option. Accumulation and other options in Dose tracking are correct.
- 2021-05-26FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, 10A, 10B, RayPlan 1, 2, 7, 8A, 8B, 9A, 10A, 10B including some of the Service Packs. Standalone software treatment planning system
For some LINAC types, merging clinical beams with beams of approximate dose may lead to the approximate dose erroneously labeled as clinical dose. Merge beams can be used manually, in scripting, or as part of the Automated breast planning feature. In some cases, when merging two beams where the first beam has clinical dose and the second beam has approximate dose, the dose of the resulting beam will be labeled as Clinical, although dose for some of the beam segments is still calculated with the SVD dose engine and it should be labeled Approximate: Mixed dose. The difference between approximate and final Clinical dose is in most cases small, but there can be body sites such as lung where the difference can be significant.
- 2021-05-26FDA-DeviceRoche Diagnostics Operations, Inc.Class IIElecsys Troponin I Immunoassay: Catalog# 05094810160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.
Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.
- 2021-05-26FDA-DeviceRoche Diagnostics Operations, Inc.Class IIElecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.
Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.
- 2021-05-26FDA-DeviceWillowWood Global LlcClass IIAlpha Interlock Suspension System Lanyard Body: An external prosthetic device. The Alpha Lanyard includes suction with a one-way air expulsion valve plus a locking mechanism. Product design features an easy release mechanism. Available with or without a titanium pyramid. The lanyard is part of the Alpha Interlock" Suspension System. This system allows you to interchange suspension methods without remaking a socket Model, 3 part numbers - 700-AIS100, Alpha Interlock Suspension System with Alpha
the firm discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over time, the shorter pin may not engage in either the upper or lower body, causing the cord to come loose or slip which could result in the loss of suspension.
- 2021-05-26FDA-DeviceW.L. Gore & Associates, Inc.Class IIGORE VIBIL Biliary Endoprosthesis: stents, drains and dilators for the biliary ducts. Catalogue numbers: VH1008040, VH1010040, VN1010040, VN0808200 - Product Usage: intended for palliation of malignant strictures in the biliary tree.
The Gore post market surveillance team received a complaint that a GORE¿ VIABIL¿ Biliary Endoprosthesis appeared shorter than the labeled length.
- 2021-05-26FDA-DeviceBoston Scientific CorporationClass IVICI RDS VENOUS STENT System Product Usage: is indicated for improving luminal diameter in the iliofemoral veins for the treatment of symptomatic venous outflow obstruction.
The firm has received reports of stent migration after implantation
- 2021-05-26FDA-DeviceBoston Scientific CorporationClass IVICI VENOUS STENT System Product Usage: is intended for the treatment of obstructions and occlusions in the venous vasculature.
The firm has received reports of stent migration after implantation
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF K7040-001 Spinning Spiros Closed Male Luer - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF CH3762 35" (89 cm) Appx 3.9 ml, 10 Drop Admin Set w/Integrated Clave" Drip Chamber, Clave", Spiros" w/Red Cap, Bag Hanger - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF CH3235 30" (76 cm) Appx 4.1 ml, Yellow 20 Drop Admin Set w/15 Micron Filter, Spiros" - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF CH3034 5" (13 cm) Bag Spike Adapter w/Spiros" w/Red Cap, Vented Cap - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF 20130-01 Spinning Spiros¿ Closed Male Luer, Red Cap REF CH2000S-PC Spinning Spiros¿ Closed Male Luer, Purple Cap - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF 20120-01 Spiros" Closed Male Luer - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF 034-H2629 Appx 1.0 ml, Adaptador universal para sistemas IV con Spiros" - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF 011-H3597 42 cm (16.5 IN) APPX 2.6ml, PUR TRANSFER SET, Clave ADDITIVE PORT, BCV, Spiros - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF 011-H2864-HSR 180 cm (71 IN)APPX 11.5ml, PUR AMBR, Spiros, CAMERA, REG. FLUSSO, BCV, Y-Clave - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF 011-CH3967 Admin Set w/ClaveTM, SpirosTM - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF CH3625 5" (13 cm) Appx 0.50 ml, Smallbore Trifuse Ext Set w/3 ChemoClave (Purple, Green Rings), Spiros w/Red Cap, 3 Clamps - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF CH3020 40" (102 cm) Appx. 4.8 ml, 20 Drop Admin Set w/In-Line Chamber, 2 Spiros", Hanger - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF CH3011 30" (76 cm) Appx 3.3 mL, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Red Cap, Bag Hanger - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF CH2060 60mL Syringe w/Spinning Spiros, Red Cap - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF CH2020 20mL Syringe w/Spinning Spiros, Red Cap - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF CL4151 45" (114 cm) Appx 1.7 ml, Smallbore Ext Set w/Spiros w/Red Cap, Anti-Siphon Valve, Clamp, Y-Connector w/ChemoLock Port, Rotating Luer - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- 2021-05-19FDA-DeviceICU Medical, Inc.Class IIREF CL3960 Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros ; w/ChemoLock; Syringe Transfer Set w/MicroClave, w/ChemoLock Port - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.