Medical device recalls
39180 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2021-02-10FDA-DeviceHeartware, Inc.Class IMedtronic HVAD Pump Implant Kits, labeled as: a. PUMP 1103, b. PUMP 1104, c. PUMP 1104JP, d. PUMP 1100, e. Model 1101, f. Model 1102, g. Model 1104CA-CLIN, h. Model 1205, i. Model MCS1705PU
Pumps manufactured with impellers from a subset of lots from a single supplier are failing to initially start, restart, or experiencing a delay in restart within the maximum number of restart attempts at a rate substantially higher than pump in the overall population.
- 2021-02-10FDA-DeviceMar-Med CoClass IITourni-Cot - Universal, UDI 00855364004085, Model Number TCU, Non-pneumatic Tourniquet - Product Usage: The device is intended to be used in surgery on fingers or toes to provide a blood-less field.
A failure can occur if the user attempts to remove the device by partially snipping the edge and pulling on it. The device may tear into two pieces with a portion remaining in place, which can continue to constrict the vessels. This type of failure can result in digit ischemia.
- 2021-02-10FDA-DevicePhilips Ultrasound IncClass IIPhilips EPIQ Diagnostic Ultrasound System; Model Numbers: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 7G, EPIQ 7C, EPIQ CVx & EPIQ CVxi. MOD EPIQ Elite DIAGNOSTIC ULTRASOUND SYSTEM; Rx Only 100-240V~50/60 Hz, 600 VA MOD EPIQ 5 DIAGNOSTIC ULTRASOUND SYSTEM; GMDN 40761, Rx Only, 100-240V~50/60 Hz, 600VA MOD EPIQ 7 DIAGNOSTIC ULTRASOUND SYSTEM; GMDN 40761, Rx Only, 100-240V~50/60 Hz, 600VA MOD EPIQ DIAGNOSTIC ULTRASOUND SYSTEM; Rx Only,100-240V~50/60 Hz, 600VA
The manufacturer discovered an issue associated with the EPIQ Image Boost with xPlane and Color Flow or Doppler while using the X8-2t TEE Transducer. If Image Boost is enabled the color box shows flow visualization not accurately represent the fluid flow. Similarly, if Image Boost is enabled, both CW and PW Doppler traces will not accurately represent the fluid flow.
- 2021-02-10FDA-DeviceThe Binding Site Group, Ltd.Class IIOptilite Freelite Mx Kappa Free Kit REF LK016.M.OPT.A
Due to customer complaints of positive bias impacting quality control. The high and low level controls are demonstrating a positive bias compared to assigned values.
- 2021-02-10FDA-DeviceFresenius Medical Care Holdings, Inc.Class IIFresenius Medical Care NaturaLyte Liquid Acid concentrate-Liquid Acid Concentrate for Bicarbonate Hemodialyis Part Number: 08-3251-9
Bottles Missing Product Label
- 2021-02-10FDA-DeviceQiagen Sciences LLCClass IISARS-CoV-2 Antigen Test
QIAGEN has become aware of the potential for false positive results to occur with some patient samples.
- 2021-02-10FDA-DeviceBiomeme, Inc.Class IIINucleic acid-based in vitro diagnostic devices for the detection of 2019-novel coronavirus (2019-nCoV) and/or diagnosis of 2019-nCoV infection in human clinical specimens. The RT-PCR test runs on a Franklin Real-Time PCR Thermocycler Device (Reference/Catalog Number 1000003,1000018) with the Software Controller version 6.4.3. The cycler has a "baseboard" unit with firmware and associated software component that control motorized movements of the optical filter carriage used for PCR detection.
Users cannot complete testing due to a sign-error in the software component that controls the filter movement and results in an instrument failure and assay failure before results are generated.
- 2021-02-10FDA-DeviceBaxter Healthcare CorporationClass IIFLOW COUPLER Monitor Flow Coupler devices are sold in 5 sizes (2.0 mm, 2.5 mm, 3.0 mm, 3.5 mm, 4.0 mm). Implantable device for vascular use.
Potential for the battery to lose its ability to be recharged.
- 2021-02-10FDA-DeviceAlung Technologies IncClass IIALung Technologies Hemolung Respiratory Assist System (RAS) with CME BodyGuard 323 Infusion Pump(EUA) - Product Usage: an off-the-shelf infusion pump and administration set to provide a 30 mL/hr continuous saline flush to the integrated blood pump bearing.
CME America BodyGuard 323 Infusion Pumps (recalled by manufacturer) is a component to the Hemolung for EUA, and the infusion pump may have a delivery inaccuracy
- 2021-02-10FDA-DeviceBrainreader ApSClass IINeuroreader Medical Image Processing Software - Product Usage: intended to automate the current manual process of identifying, labeling, and quantifying the volume of segmentable brain structures identified on MR images.
The Neuroreader has been distributed with a reference normative database that is different from the version reviewed and cleared by FDA.
- 2021-02-10FDA-DeviceSmith & Nephew, Inc.Class IIVERSABOND AB 40 GRAMS FORMULATION 2, antibiotic infused bone cement, Product Number: 71271440 - Product Usage: VERSABOND is a medium viscosity bone cement which can be used for hip, knee, and shoulder implant fixation. VERSABOND AB includes the antibiotic- gentamycin.
Potential heat and humidity exposure. The affected lot contained an atypical sandy, flaky or grainy consistency when mixing.
- 2021-02-10FDA-DeviceHill-Rom, Inc.Class ILiko Multirall 200 overhead lift- overhead lift is a general-purpose lift with the intended use in healthcare, intensive care and rehabilitation. Affected devices: Liko Multirall 200 (product number 3130001) Universal SlingBar 450 R2R (product number 3156095) Universal SlingBar 350 R2R (product number 3156094) Carriage D45 with Double Hook (product number 3136100) Extension belt 300-400 mm (product number 3136226) Extension belt 400-600 mm (product number 3136227) Extension belt 600-100
Multirall Q-link strap lock has not been fully engaged in the S65 rail carriage hook by user, leading to the Multirall lift becoming detached from the rail system, potentially resulting in the fall of both the lift motor and the patient
- 2021-02-03FDA-DeviceStryker Sustainability SolutionsClass IICardiovascular Reprocessed ViewFlex Xtra ICE Catheter, REF D087031, OM St. Jude Medical, Size 9F (3.2mm), Length 90 cm, Sterile EO, Rx Only, Quantity 1, UDI: 07613327376067
A subset of temporary intracardiac ultrasound catheters are labeled with the incorrect expiration date. They are mislabeled as having a shelf life of two years, while the actual shelf life is one year. The earliest affected product will expire in May 2020. Use of mislabeled devices may lead to hazards sterility breach, component breakdown or delay in treatment.
- 2021-02-03FDA-DevicePhilips North America, LLCClass IITRx4841A 1.4 GHz IntelliVue Tele TRX, Model Number 862439; Includes System Codes 453564007261, 453564007271, 453564007281, 989803196951
The ECG signal from patients being monitored using a Philips TRx4841A and TRx4851A Telemetry Transceiver may not be properly analyzed when it is used with a Philips Patient Information Center iX Release C.02.xx or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms.
- 2021-02-03FDA-DevicePhilips North America, LLCClass IITRx4851A 2.4 GHz IntelliVue Tele TRX, Model Number 862231; Includes Service Numbers 453564052401, 453564052411, 453564052441, 453564052451, 453564166851, 453564166861
The ECG signal from patients being monitored using a Philips TRx4841A and TRx4851A Telemetry Transceiver may not be properly analyzed when it is used with a Philips Patient Information Center iX Release C.02.xx or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms.
- 2021-02-03FDA-DeviceQiagen Sciences LLCClass IItherascreen PIK3CA RGQ PCR Kit (23) US IVD product- aid clinicians in identifying breast cancer patients who may be eligible for treatment with PIQRAY (alpelisib) based on a PIK3CA Mutation Detected result. REF 873121
Kit may generate false Q546R mutation positive results caused by nonspecific molecular interactions within the Q546R higher than previously observed and described within the Instructions For Use.
- 2021-02-03FDA-DeviceBoston Scientific CorporationClass IIEMBLEM MRI S-ICD pulse generators, as part of Boston Scientific S-ICD System; Model A219
Accelerated battery depletion may result in a need for device replacement earlier than expected.
- 2021-02-03FDA-DeviceBoston Scientific CorporationClass IIEMBLEM S-ICD pulse generators, as part of Boston Scientific S-ICD System; Model A209
Accelerated battery depletion may result in a need for device replacement earlier than expected.
- 2021-02-03FDA-DeviceFujifilm Medical Systems U.S.A., Inc.Class IIIWako HDL-C/LDL-C Calibrator - Product Usage: designed to be used with Wako L-Type HDL-C or L-Type LDL-C for the determination of high density lipoprotein cholesterol (HDL-C) or low density lipoprotein cholesterol (LDL-C) in serum or plasma.
FUJIFILM has become aware that incorrect concentration values were printed on the bottle label for all product manufactured as part of Lot EL629. This may result incorrect results when used in testing.
- 2021-02-03FDA-DeviceK2M, IncClass IIYukon Polyaxial Screw Size 4.0x26mm (catalog number 7601-04026), component of YUKON Spinal System
The Yukon polyaxial screw used in the Yukon Spinal System was incorrectly laser marked with size 5.0 mm (diameter) & 24 mm (length).
- 2021-02-03FDA-DeviceBraxton Medical CorporationClass IIB. Braun Introcan IV Safety Catheters, sterile - Product Usage: It is a single-use device to generate intravascular and tissue access to sample blood, monitor blood pressure, or administer fluids and blood intravascular.
Recalled products do not have FDA approval for sale in the United States.
- 2021-02-03FDA-DeviceNeocis, Inc.Class IIDRW-0739-01:Gen 2 Anterior Chairside Splint LRUL (lower right-upper left); accessories to the parent system, the Neocis Guidance System (NGS)
The LRUL and LLUR posterior chairside splints and the LLUR anterior chairside splints may be mislabeled.
- 2021-02-03FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIMedtronic Cobalt DR MRI SureScan DDPB3D1, MR Conditional implantable dual chamber cardioverter defibrillator with SureScan technology and Bluetooth wireless telemetry (DDE-DDDR)
Under rare circumstances, the component may be susceptible to a failure mechanism that could impact device performance, including potential loss of function.
- 2021-02-03FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIMedtronic Cobalt HF CRT-D MRI SureScan DTPB2D1, MR Conditional implantable cardioverter defibrillator with cardiac resynchronization therapy, SureScan technology, and Bluetooth wireless telemetry (DDE-DDDR)
Under rare circumstances, the component may be susceptible to a failure mechanism that could impact device performance, including potential loss of function.
- 2021-02-03FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIMedtronic COBALT HF QUAD CRT-D MRI SureScan, Model Number DTPB2QQ, MR Conditional implantable cardioverter defibrillator with cardiac resynchronization therapy, SureScan technology, and Bluetooth wireless telemetry (DDE-DDDR)
Under rare circumstances, the component may be susceptible to a failure mechanism that could impact device performance, including potential loss of function.
- 2021-02-03FDA-DeviceBecton Dickinson & CompanyClass IIBD Regional Block Needles 22G x 1.50in (50 units/cs.). Indicated for administration of medications to affect anesthesia or analgesia to a specific anatomical location (region). Catalog No. 408348
BD was advised by the European Notified Body, NSAI, that the instructions for use (IFU) needs to be updated to include the general contraindications
- 2021-02-03FDA-DeviceBecton Dickinson & CompanyClass IIBD Stimex Insulated Needle 22G x 2.13in ( 50 units/cs.) Indicated for administration of medications to affect anesthesia or analgesia to a specific anatomical location (region). Catalog No. 404204
BD was advised by the European Notified Body, NSAI, that the instructions for use (IFU) needs to be updated to include the general contraindications
- 2021-02-03FDA-DeviceOMNIlife science Inc.Class IISERF Novae Dual Mobility Cup Insert 49/28-ApeX-LNK Poly Liner. Intended for use with the Apex Modular Acetabular Cup for total hip replacement procedures. Product Code: H6-24928
Acetabular Liners have a higher than acceptable oxidation limit, potential for patients to experience a higher wear rate for acetabular liners and require a revision surgery
- 2021-02-03FDA-DeviceOMNIlife science Inc.Class IISERF Novae Dual Mobility Cup Insert 47/28- ApeX-LNK Poly Liner. Intended for use with the Apex Modular Acetabular Cup for total hip replacement procedures. Product Code: H6-24728
Acetabular Liners have a higher than acceptable oxidation limit, potential for patients to experience a higher wear rate for acetabular liners and require a revision surgery
- 2021-02-03FDA-DeviceOMNIlife science Inc.Class IISERF Novae Dual Mobility Cup Insert 51/28- ApeX-LNK Poly Liner. Intended for use with the Apex Modular Acetabular Cup for total hip replacement procedures. Product Code: H6-25128
Acetabular Liners have a higher than acceptable oxidation limit, potential for patients to experience a higher wear rate for acetabular liners and require a revision surgery