Medical device recalls
39180 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2020-10-28FDA-DeviceMedtronic VascularClass IRashkind Balloon Septostomy Catheter, 6F, Recess Balloon, Closed end, 50 cm, Pediatric, Single Lumen Product number: 008764 GTIN: 00613994760289 The Rashkind balloon septostomy catheters are designed for creating an atrial septal defect or enlarging an existing atrial septal defect as a treatment option for patients with a Cyanotic Congenital heart defect. Cyanotic congenital heart defects are rare conditions that are associated with high morbidity and mortality.
Subsequent failure of catheters in the field and failed quality testing
- 2020-10-21FDA-DeviceSkytron, Div. The KMW Group, IncClass IISkyVision SDS System - Video Integration Model: GC-SDS-BSCB-200 - Product Usage: is a high quality, self-contained video switching and control unit for use in the operating room.
Intermittent failure of a signal being sent to a monitor from the touch panel display causing the displaying monitor to go Blank and the touch panel to display a Blank image when reviewing the target display. If this issue were to happen during a case, cause increase risk to patient due to hospital losing the visual image.
- 2020-10-21FDA-DeviceSiemens Medical Solutions USA, IncClass IIArtis zee or Artis Q systems (listed below) - Product Usage: for single and biplane diagnostic imaging and interventional procedures. System Name/Material Number: Artis Q BIPLANE 10848282 ARTIS Q CEILING 10848281 Artis Q floor 10848280 ARTIS Q ZEEGO 10848283 ARTIS Q.ZEN BIPLANE 10848355 ARTIS Q.zen ceiling 10848354 Artis Q.zen floor 10848353 Artis zee Biplane 10094141 Artis zee ceiling 10094137 Artis zee floor 10094135 Artis zee floor 10094135 Artis zee multi-purpose 10094139 Art
Potential exists for the collision supervision not to work properly following loss of the individual room configuration data settings causing danger of collision with fix mounted room equipment, walls or floor may occur. This might cause components of the system to fall or tilt which could result in crushing of patients, operators or staff, collision of system parts with patients, as well as result in delay or interruption of the clinical procedure
- 2020-10-21FDA-DeviceBEYOND WHITE SPA LLC D.B.A. EZYWIPE OF AMERCIAClass IIEzycare Face Mask, Disposable Non Medical Face Masks - Product Usage: marketed under face mask umbrella EUA as source control.
Disposable face masks labeled as non-medical were distributed to customers expecting medical use face masks.
- 2020-10-21FDA-DeviceCook Inc.Class IIPediatric Nephrostomy Stent Set, Reference Part Numbers 080208-S6 and 080208-S7, Order Numbers G15034 and G15035. The set contains a radiopaque pigtail catheter and stent, retention disc with pull tie, and a 10 French, 30 cm polyvinylchloride connecting tube with one-way stopcock.
The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. The current IFU is in process of being updated and will be provided with orders placed following implementation.
- 2020-10-21FDA-DeviceCook Inc.Class IIPercutaneous Neonatal Pigtail Nephrostomy Set, Reference Part Number 080106, Order Number G14508. The set contains a 22 gage, 5 cm Chiba Needle; an 18 gage, 5 cm trocar needle; an .038-inch (.97 mm) diameter, 30 cm PTFE-coated stainless steel; a 5, 6, and 7 French, 10 cm radiopaque dilator; a 6 French, 7 cm radiopaque pigtail catheter; a retention disc with pull tie; and a 10 French, 10 cm clear, polyvinylchloride connecting tube with one-way stopcock.
The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. The current IFU is in process of being updated and will be provided with orders placed following implementation.
- 2020-10-21FDA-DeviceRoche Diagnostics Operations, Inc.Class IIIAlbumin Gen.2, Catalog 05166861190
Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.
- 2020-10-21FDA-DeviceRoche Diagnostics Operations, Inc.Class IIITina-quant Complement C4 ver.2, Catalog 05991994190
Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.
- 2020-10-21FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAtellica CH Systems Fructosamine (FRUC)-For in vitro diagnostic use in the quantitative measurement of glycated protein (fructosamine) in human serum or plasma on the ADVIA¿ Chemistry systems Siemens Material Number (SMN): 11097637
Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to a potential adjustment of hyperglycemic therapy. Positively biased fructosamine results may prompt changes in glucose management that may affect glycemia. In extreme situations, more aggressive management of glycemia may occur and contribute to hypoglycemia.
- 2020-10-21FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIADVIA Chemistry Systems-Fructosamine (FRUC)-For in vitro diagnostic use in the quantitative measurement of glycated protein (fructosamine) in human serum or plasma on the ADVIA Chemistry systems Siemens Material Number (SMN): 10361941
Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to a potential adjustment of hyperglycemic therapy. Positively biased fructosamine results may prompt changes in glucose management that may affect glycemia. In extreme situations, more aggressive management of glycemia may occur and contribute to hypoglycemia.
- 2020-10-21FDA-DeviceCooperSurgical, Inc.Class IIMediCult Vitrification Cooling Media, Model Number 12284001F
There is a discrepancy between the printing on the Product Tyvek label of the individual component packaging. The Chinese characters on the label indicates that it contains an Equilibration Medium (P/N: 12220001) rather than the Vitrification Medium (P/N: 12230001) as indicated on the remainder of the label.
- 2020-10-21FDA-DeviceGentian ASClass IIGENTIAN Cystatin C Calibrator and Calibrator Kit, for the following models: 1. Gentian C E, Cystatin C Calibrator, 1 x 1 mL, REF 1012, UDI: (01)07090038460092 2. Gentian C E Calibrator Kit, 6 x 1 mL, REF 1051, UDI: (01)07090038460108 3. Gentian C E, BCI Ref: A52763, Cystatin C calibrator Kit, 6 x 1 mL, REF 1051, (01)17090038460037 4. Gentian C E, Cystatin C Calibrator, 1 x 1 mL, REF 8012 (Not distributed in US)
The firm has identified lots of calibrators that have shown to measure incorrect high values over time compared to the assigned values when released for distribution. This could result in higher Cystatin C values measured in patients.
- 2020-10-21FDA-DeviceFresenius Medical Care Holdings, Inc.Class IIAquaBplus and AquaBplus B2 Water Purification System. For use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies. AquaBplus 1500 - Typ USA - 3 x 208V / 60Hz -Fresenius Part Number: 24-1500-1 AquaBplus B2 2000 / B1 2500 (Bplus DUO 2000) - 3 x 208V / 60Hz- Fresenius Part Number: 24-2000-0
Incorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections could become loose and may adversely affect the electrical safety of the products, potentially resulting in electrical shock to the Biomed Technician or clinic staff.
- 2020-10-21FDA-DeviceBiomet, Inc.Class II(1) RingLoc Bi-Polar Hip System Acetabular Cup ArCom: Acetabular Bi-Polar Cup, E1 Antioxidant Infused, 28 MM, 41 MM: Item No. 110010458; Lot No. 710930; UDI No. (01) 00880304568747 (17) 230614 (10) 710930 (2) RingLoc Hip system Acetabular Bi-Polar Cup: Acetabular Cup, ArCom, 41 MM OD, 28 MM ID: Item No. 11-165206, Lot No. 649750; UDI No. (01) 00880304001923 (17) 230614 (10) 649750 (3) RingLoc Bi-Polar Hip System Acetabular Cup ArCom: Acetabular Cup, ArCom, 52 MM OD, 28 MM ID: It
Zimmer Biomet is conducting a medical device recall for three lots of the RingLoc Bi-Polar Hip system Acetabular Cup. The affected lots were gamma sterilized twice. Sufficient data does not exist to support the functionality, shelf life, or package integrity for more than one-time sterilization.
- 2020-10-21FDA-DeviceHandicare ABClass IISwiftHook sold as the following: a. Individually as SwiftHook b. Included with Vega505EE mobile lifts Product Usage: The mobile lifts are intended to assist people with functional impairments in scenarios involving physical positioning.
Carry bar may detach during use of mobile lifts and ceiling lifts equipped with the Swifthook component.
- 2020-10-21FDA-DeviceAlcon Research, LLCClass IIAlcon Custom-Pak Surgical Procedure Packs containing Cardinal Health" PROTEXIS" Latex Micro Surgical Gloves - Product Usage: used by ophthalmic surgeons in a variety of ophthalmic surgeries.
The third-party brand latex micro surgical gloves packaged within specific custom packs may have a degradation defect that could lead to holes and donning tears at the folding area around the cuffs. This could increase the potential risk of infection or cross-patient (clinician) exposure to body fluids.
- 2020-10-21FDA-DeviceGenicon, Inc.Class IIAquas PowerFlow Tube Set - Product Usage: The device delivers sterile irrigation fluids to surgical sites during laparoscopic procedures and to evacuate blood, tissue debris, and smoke from the surgical site
The product design incorporating a motor driven impeller pump powered by one 9V alkaline battery inside the hand-piece did not receive proper premarket clearance from US FDA and lacks a 510(k).
- 2020-10-21FDA-DeviceBiomet, Inc.Class IIBiomet Comprehensive Reverse Shoulder Glenosphere Mini Baseplate with Taper Adapter-Shoulder Prosthesis item: 010000589 - Product Usage: intended to serve as the foundation for the reverse shoulder prosthesis head (glenosphere).
Product potentially being packaged without a taper adapter or baseplate. The package should contain one (1) baseplate and one (1) taper adapter, product may be packaged with two (2) baseplates and no (0) adapters or with no (0) baseplates and two (2) adapters
- 2020-10-21FDA-DeviceGetinge Group Logistics America, LLCClass IIServo-i Ventilator, Part No. 6487800, UDI Code 07325710000823 - Product Usage: intended for general and critical ventilator care, and it can be used for low end to high end intensive care, in different configurations. It supports a variety of breathing modes and can be used for Neonates, Pediatric and Adults.
A potentially shorter than specified nebulizer connector may result in difficulty in installing the Aeroneb Nebulizer module to the Servo-i device successfully.
- 2020-10-21FDA-DeviceMizuho OSIClass IILevo Arm Label: LEVO ARM REF 7887-050 MIZUHO OSI Refurbished Levo Arm Label: LEVO ARM REF 7887-050R MIZUHO OSI - Product Usage: used with Mizuho OSI spinal surgery tables to provide patient head support and positioning during spinal surgery.
Due to the potential for fluid ingress into the Arm which could affect the device's electrical system and lead to a failure of the locking mechanism
- 2020-10-14FDA-DeviceSiemens Medical Solutions USA, Inc.Class IIACUSON AcuNav Volume Intracardiac Echocardiography Catheter 12.5F
Due to adverse events received by the firm for left atrial perforations and Atrioventricular Node block.
- 2020-10-14FDA-DeviceRoche Diagnostics Operations, Inc.Class IIDISCOVERY ULTRA Advanced Staining System (RUO-Research Use Only) Material Number: 05987750001
Potential for Fluid leaks within the instruments, creating an electrical short inside the power socket, which generates heat, smoke, and eventually burning and melting of the electrical cord connector and socket
- 2020-10-14FDA-DeviceRoche Diagnostics Operations, Inc.Class IIBenchMark ULTRA Advanced Staining System (IVD)- Material Number: 05342716001
Potential for Fluid leaks within the instruments, creating an electrical short inside the power socket, which generates heat, smoke, and eventually burning and melting of the electrical cord connector and socket
- 2020-10-14FDA-DeviceBiomet, Inc.Class IIBiomet ARTHROSIMPLICITY KIT EU Item Number: 24-4055
Product not properly being aligned with the adequate gamma sterilization group and result in the product not being properly sterilized
- 2020-10-14FDA-DeviceBiomet, Inc.Class IIBiomet ONPOINT SCOPE PROCEDURE KIT-EU Item Number: 24-3055
Product not properly being aligned with the adequate gamma sterilization group and result in the product not being properly sterilized
- 2020-10-14FDA-DeviceBiomet, Inc.Class IIBiomet OnPoint SCOPE PROCEDURE KIT Item Number: 24-3050
Product not properly being aligned with the adequate gamma sterilization group and result in the product not being properly sterilized
- 2020-10-14FDA-DeviceBiomet, Inc.Class IIBiomet Regenerex Primary Taper Cap Item Number 141269
Product not properly being aligned with the adequate gamma sterilization group and result in the product not being properly sterilized
- 2020-10-14FDA-DeviceBard Peripheral Vascular IncClass IIBard Conquest PTA Balloon dilation Catheter, 6mm x 40mm, 75 cm, Cat. No. CQ7564, UDI: (01) 00801741062902 - Product Usage: use in Percutaneous Transluminal Angioplasty of the femoral, iliac, and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
Dilation catheter packaged with the wrong size balloon. The labeling states that the balloon size is 6mm x 40mm, however included is a 5mm x 40mm balloon. Use of a balloon that has a smaller diameter than expected could result in inconsequential prolongation of an existing procedure.
- 2020-10-14FDA-DeviceBiomet 3i, LLCClass IICERTAIN BELLATEK TITANIUM ABUTMENT 6.0MM Reference Number: IEDAT6
Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may cause misalignment, the definitive abutment will not seat in the correct orientation upon attempted placement with the patient
- 2020-10-14FDA-DeviceBiomet 3i, LLCClass IICERTAIN BELLATEK TITANIUM ABUTMENT 5.0MM Reference Number: IEDAT5
Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may cause misalignment, the definitive abutment will not seat in the correct orientation upon attempted placement with the patient