- 2020-01-15FDA-DeviceAbbott MedicalClass IIFast-Cath Hemostasis Introducer, HEMOSTASIS INTRODUCER, 8.5F, 2.83 mm, REF 406143 - Product Usage: The Fast-Cath Hemostasis Introducers are indicated for the introduction of angiographic catheters, closed-end catheters, balloon catheters, and electrodes into a vessel where minimizing blood loss is essential.
Three lots of Fast-Cath Hemostasis Introducer devices contained 8F sized dilators and sheaths instead of the expected 8.5F devices.
- 2020-01-15FDA-DeviceAvanos Medical, Inc.Class IIHaylard Closed Suction System for Neonates/Pediatrics, 5 Fr, 1. REF 195, Y-Adapter, SAP Finished Product Code 109382303; 2. REF 201, with BALLARD Technology, SAP Finished Product Code 109382803; 3. REF 202, Elbow, SAP Finished Product Code 109382903; 4. REF 195-5 Y-Adapter, SAP Finished Product Code 109838302.
Avanos Medical has received reports stating that the central lumen of some 5 Fr Neonate/Pediatric Closed Suction Catheters may be fully blocked or partially blocked. This may cause the closed suction catheters (CSCs) to inadequately suction secretions and/or saline fluids form the patient's airway to prevent oxygen desaturation.
- 2020-01-15FDA-DeviceIntersurgical IncClass IISuperset with 22MM ID, Double Swivel Elbow 22MM OD/15MM ID and Bronchoscopy Port, REF 3521000. Airway connector.
The port may crack in the swivel elbow where the flip cap is inserted, and these cracks could possibly leak if they reach a significant size.
- 2020-01-15FDA-DeviceIntersurgical IncClass IISuperset with 22MM ID, Double Swivel Elbow 22MM OD/15MM ID and Bronchoscopy Port, REF 3503031. Airway connector.
The port may crack in the swivel elbow where the flip cap is inserted, and these cracks could possibly leak if they reach a significant size.
- 2020-01-15FDA-DeviceIntersurgical IncClass IIDouble Swivel Elbow with Bronchoscopy Port 15MM OD, 22MM OD/15MM ID, REF 1996030. Airway connector.
The port may crack in the swivel elbow where the flip cap is inserted, and these cracks could possibly leak if they reach a significant size.
- 2020-01-15FDA-DeviceIntersurgical IncClass IIDouble Swivel Elbow with Bronchoscopy Port 15MM OD, 22MM OD/15MM ID, REF 1898000. Airway connector.
The port may crack in the swivel elbow where the flip cap is inserted, and these cracks could possibly leak if they reach a significant size.
- 2020-01-15FDA-DeviceAbbott MedicalClass IIFast-Cath Trio HEMOSTASIS INTRODUCER, 14F, 4.6 mm, REF 406303
Fast-Cath Trio Hemostasis Introducer devices contained 12F sized dilators and sheaths instead of the expected 14F devices.
- 2020-01-15FDA-DeviceSiemens Medical Solutions USA, IncClass IICios Alpha, Model Number 10308191; mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical application
During a routine inspection of a system, a crack on the C-arm holder was discovered. In the event the C-arm holder has a crack, the stability of the C-arm may be compromised and could lead to deformations of the C-arm. In very unlikely cases, an extreme impact to the C-arm such as heavy collision with a wall or table, may lead to a detachment of the C-arm and could potentially result in injury to patients and staff.
- 2020-01-15FDA-DeviceStryker GmbHClass IIStryker Pin To Rod Coupling External Fixation System Hoffmann II Compact - Pin To Rod Coupling 5/3-4mm Catalog #: 4940-1-020 - Product Usage: Indications Distal radius fractures (intra and extra) articular Intercarpal fracture dislocations Unstable forearm fractures Unstable elbow dislocation Multiple extremity fractures Radial osteotomies Foot and ankle fractures Paediatric fractures
Pin To Rod Coupling could not clamp the pin and rod as intended
- 2020-01-15FDA-DeviceStryker GmbHClass IIStryker T2 Alpha Delta Strike Plate IMN Instruments - part of the IMN Instruments System and is used to support insertion and withdrawal of nails through the Targeting Device Catalog Number: 2351-0050
Instrument can break at the level of the thread when being exposed to high forces during nail implantation or removal which can lead to complications during nail insertion and removal.
- 2020-01-15FDA-DeviceMedical Action Industries, Inc. 306Class IIMedical convenience trays and kits, labeled as: Medical Action Industries, Inc.: ***CVC*** Clear Sequence Change Kit, REF 77070B ***DIALYSIS*** Clear Sequence Change Kit, REF 77068B ***PEDS CVC*** Clear Sequence Change Kit, REF 77190B ***PICC*** Clear Sequence Change Kit, REF 77069B Adult Central Line Dressing Change Kit, REF 74636C Central Line Dressing Kit, VCU, REF 57647B Central Line Removal Kit, REF 78631B Chest Tube Cleaning Kit, REF 73000 CVC Dressing Change Kit, REF 80643D D
The kits contain a bottle of saline solution which may exceed the USP specified pH requirements of 4.5-7.0.
- 2020-01-15FDA-DeviceNextremity SolutionsClass IIStratum MDS Screw 3.5x50mm ST Part Number: STRM-MDS-3550ST
Components 2mm shorter than indicated on the package label
- 2020-01-15FDA-DeviceNextremity SolutionsClass IIStratum MDS Screw 3.5x48mm ST Part Number: STRM-MDS-3548ST
Components 2mm shorter than indicated on the package label
- 2020-01-15FDA-DeviceNextremity SolutionsClass IIStratum MDS Screw 3.5x46mm ST Part Number: STRM-MDS-3546ST
Components 2mm shorter than indicated on the package label
- 2020-01-15FDA-DeviceMedacta Usa IncClass IIMedacta Shoulder SYSTEM HUMERAL ANATOMICAL METAPHYSIS, REF 04.01.0023 through 04.01.0055, all sizes - Product Usage: The Medacta Anatomic Shoulder Prosthesis is indicated for treatment of humeral fractures and for primary or revision total shoulder replacement in patients with an intact or reparable rotator cuff shoulder joint, severe arthropathy or a previously failed joint replacement.
Certain lot numbers of the Medacta Humeral Anatomical Cementless Metaphysis have a dimensional condition that could limit insertion of the screw used to lock the metaphysis to the humeral diaphysis.
- 2020-01-15FDA-DeviceAerolase CorporationClass IILightPod Neo Lasers
This relates to the standard 21 CFR 1040.10(f)(3) requiring a readily available remote interlock connector for the laser device
- 2020-01-15FDA-DeviceAerolase CorporationClass IILightPod ERA lasers
This relates to the standard 21 CFR 1040.10(f)(3) requiring a readily available remote interlock connector for the laser device
- 2020-01-08FDA-DeviceMedela IncClass IIInvia Foam Dressing Kit with FitPad - X-Large (3 pcs) SKU # 0876224. The Invia Foam Dressing Kit with FitPad consists of a foam pad, a section interface (Invia FitPad) and Transparent Film (one or more pieces depending on the kit size).
Invia Foam Dressing Kit with FitPad could have holes in the sterile packaging.
- 2020-01-08FDA-DeviceNucletron BVClass IIThere is a complaint regarding internal check cable guide mechanism failure for the Nucletron Flexitron brachytherapy afterloader system. Due to an obstruction during the check cable out drive, the internal guiding tube became displaced and the check cable became damaged. This issue could cause treatment interruption or in a very unlikely scenario, incorrect source positioning and radiation treatment errors.
A component failure may lead to treatment interruption or incorrect source positioning.
- 2020-01-08FDA-DeviceNidek, Inc.Class IIINIDEK Specular Microscope, Model CEM-530 - Product Usage: The device provides noncontact, high magnification image capture of endothelium enabling observation of the size and shape of cells. Information such as the number of endothelial cells, cell density, and cell area is analyzed through the captured images.
The electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2)
- 2020-01-08FDA-DeviceCarl Zeiss Meditec, Inc.Class IIThe PLEX Elite 9000, a Swept-Source OCT [SS-OCT] - Product Usage: The PLEX Elite 9000 Swept-Source OCT [SS-OCT] is a non- contact, high resolution, wide field of view tomographic and biomicroscopic imaging device intended for in-vivo viewing, axial cross-sectional and three-dimensional imaging of posterior ocular structures. The device is indicated for visualizing posterior ocular structures including, but not limited to, retina, retinal nerve fiber layer, ganglion cell plus inner plexiform lay
Under certain fault conditions, the existing design may expose the operator to a hazardous voltage when contacting the metal enclosure (including mounting screws) around the up/down switch.
- 2020-01-08FDA-DeviceMedtronic Navigation, Inc.-LittletonClass IIO-arm" O2 Imaging System, Product Number/CFN: BI-700-02000
Due to a manufacturing issue affecting particular O-arm" O2 Imaging Systems, Medtronic has determined that the gantry tractor motor drive belt may loosen over an extended number of uses at a higher frequency in these systems. If the belt loosens, it may result in rotation of 3D images about the gantry isocenter. While the 3D image is anatomically accurate within the image itself and may be used to confirm therapy, its electronic registered location may be rotated relative to the actual physical position of the patient.
- 2020-01-08FDA-DeviceVilla Sistemi Medicali S.P.A.Class IIApollo EZ, Model Number 9784152036 - Product Usage: The Apollo EZ remote-controlled table is X-ray equipment that can be used to perform general radiographic and fluoroscopic examinations. The system is a fixed position type. The movements of the table and tube support are motorized and remote-controlled. Medical purpose is the diagnosis, visualization and documentation of various human diseases or injuries.
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
- 2020-01-08FDA-DeviceVilla Sistemi Medicali S.P.A.Class IIApollo DRF, Model Number 9784220831 - Product Usage: The remote controlled table Apollo DRF is an X-ray equipment that allows to perform general radiographic and fluoroscopic examinations. The system is a fixed position type. The movements of the table and tube support are motorized and remote-controlled. Medical purpose is the diagnosis, visualization and documentation of various human diseases or injuries.
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
- 2020-01-08FDA-DeviceVilla Sistemi Medicali S.P.A.Class IIApollo, Model Numbers 9784100231, 9784120131, 9784130131 - Product Usage: Apollo table is a remote controlled X-ray system suitable for radiography and fluoroscopy examinations. The system is stationary mounted. The movements of the table and the tube stand are motorized and remotely controlled. Medical purpose is the diagnosis, visualization and documentation of various human diseases or injuries.
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
- 2020-01-08FDA-DeviceVilla Sistemi Medicali S.P.A.Class IIJuno DRF, Model Number 709020 - Product Usage: General. multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures.
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
- 2020-01-08FDA-DeviceDeerfield Imaging, Inc.Class IIiMRX MRI and Angiography System, part number 10094137 and 10094141, a surgical/interventional suite with integrated MRI and X-ray angiography. The IMRIS iMRX system is the integration of a Siemens Artis zee family (biplane or floor or ceiling mounted single plane) and the IMRIS Neuro 3T System. The System is a solution providing MRI scanning capabilities in the angiography room while retaining all functionality of Siemens Artis zee system. The System includes a traditional MRI unit that has
There is a potential hardware error for the iMRX System with a generator A100. A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible
- 2020-01-08FDA-DeviceAbbott Point Of Care Inc.Class IIi-STAT cTnI cartridge
i-STAT¿ cTnI blue cartridges may generate a higher than expected number of quality check codes. The observed rate of quality check codes could be in the 3 to 5% range.
- 2020-01-08FDA-DeviceMedicrea InternationalClass IIMedicrea Pass LP, REF B02217545, Non Sterile, Material: Ti-6Al-4V ELI, VIS PEDICULAIRE POLYAXIALE 7.5MM x 45MM, POLYAXIAL PEDICLE SCREW 7.5MM x 45 MM, POLYAXIAL PEDIKELSCHRAUBE 7.5MM X 45MM, TORNILLO PEDICULAR POLIAXIAL 7.5MM x 45MM VITE POLIASSIALE PEDUNCOLARE 7.5MM x 45MM, PARAFUSO PEDICULAR POLIAXIAL 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, WIELOOSIOWA SRUBA TRANSPEDIKULARNA 7.5MM x 45MM, (01)03613720177906(10)19A0630(17)290103,
It has been determined that some Pedicle system screw identified as 45mm long have an actual length of 50mm. Pedicle screws with a diameter of 7.5 mm or more are reserved for the sacrum area. As a result, a screw that is too long could come into contact with the nerves and cause postoperative pain.
- 2020-01-08FDA-DeviceRoche Diabetes Care, Inc.Class IIBlood Glucose Monitoring System, Model Number 08313202001 ACCU CHEK GUIDE HCP SAMPLE LAUNCH BOX - Product Usage: The Accu-Chek Guide Blood Glucose Monitoring System is intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectiveness of glucose control. The Accu-Chek Guide blood Glucose Monitoring System is intended for in vitro diagnostic single-patient use by people with diabetes. The Accu-Chek Guide Blo
Potential power issues with the Accu-Chek¿ Guide Blood Glucose Monitoring System.