- 2019-12-25FDA-DeviceSmith & Nephew, Inc.Class IIsmith&nephew PROFIX Mallet, 750 GRAMS, REF 71512452 - Product Usage: The PROFIX Mallet is a reusable surgical instrument used to tap instruments or implants during knee or hip surgery.
There is a a potential failure mode associated with the use of the mallet, wherein the poly tip of the mallet may become loose and/or come off during use. If the potential failure mode occurs, particles generated in the poly thread potentially could enter the surgical field or a delay in surgery could occur if the tip disengages during the use of the mallet.
- 2019-12-25FDA-DeviceSiemens Medical Solutions USA, IncClass IIArtis zee/ zeego Interventional Fluoroscopic X-Ray System, Model Number: 10094142, Software Version: VC21C
In affected Artis zee floor Interventional Fluoroscopic X-Ray Systems, the activation of a collision sensor will cause a block the movement of the floating tabletop and could result in a delay or interruption of a procedure.
- 2019-12-25FDA-DeviceTekia, Inc.Class IITEKIA Model 872 Single Piece Hydrophilic Acrylic Intraocular Lenses, Diopter 15.0, Sterile. The firm name on the label is TEKIA, INC., Irvine, CA.
A limited number of lenses may be mislabeled with the incorrect diopter power.
- 2019-12-25FDA-DeviceMedtronic Perfusion SystemsClass IIMedtronic Affinity, 38um Arterial Filter, Catalog number REF 351
Medtronic has identified an out of-specification condition exhibiting excess plastic (flash) in the arterial filter directly above the outlet port.
- 2019-12-25FDA-DeviceAmeditech IncClass IIIQuickTox 5 Panel Drug Screen DipCard
Mixed Products containing QuickTox 5 Panel Drugs Screen Dipcard (P/N QT11, LOT #188787) and Alere iScreen Drugs of Abuse Dip Test (P/N X11-IS10-22015-DX LOT #188744)
- 2019-12-25FDA-DeviceIntegra Lifesciences Sales LlcClass IIAscension¿ Silicone PIP-Silicone PIP Sz 1, Model Number: SPIP5201WW - Product Usage: The Ascension Silicone PIP implant is intended for cementless replacement of the proximal interphalangeal joint in patients with advanced osteoarthritis, post-traumatic arthritis and rheumatoid arthritis.
Silicone PIP Implant mislabeled as a size 1 implant on the outer packaging when in fact the package contained a size 0 implant
- 2019-12-25FDA-DeviceTeleflex MedicalClass IIArrow EZ-IO Intraosseous Vascular Access System 45mm Needle+Stabilizer Kit, Product Code 9079P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides intraosseous access in the proximal humerus, proximal tibia, and distal tibia of adult and pediatric patients, and the distal femur in pediatric patients when vascular access is difficult to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.
- 2019-12-25FDA-DeviceTeleflex MedicalClass IIArrow EZ-IO Intraosseous Vascular Access System 25mm Needle+Stabilizer Kit, Product Code 9001P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides intraosseous access in the proximal humerus, proximal tibia, and distal tibia of adult and pediatric patients, and the distal femur in pediatric patients when vascular access is difficult to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.
- 2019-12-25FDA-DeviceTeleflex MedicalClass IIArrow EZ-IO Intraosseous Vascular Access System 15mm Needle+Stabilizer Kit, Product Code 9018P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides intraosseous access in the proximal humerus, proximal tibia, and distal tibia of adult and pediatric patients, and the distal femur in pediatric patients when vascular access is difficult to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.
- 2019-12-25FDA-DeviceSouthwest Technologies IncClass IIElasto-Gel Absorbent Wound Dressing, Item DR8000. 5 Dressings 4" x 4". Sterile. UPC 45713-58000 Bacteriostatic
Elasto-Gel Wound Dressing may be contaminated with mold.
- 2019-12-25FDA-DevicePolymer Technology SystemsClass IICVS Health At Home A1C Test Kit, Software Version Revision D, REF Number 3059
A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
- 2019-12-25FDA-DevicePolymer Technology SystemsClass IIReliOn FastA1C Test, Software Version Revision D, REF Number 3055
A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
- 2019-12-25FDA-DevicePolymer Technology SystemsClass IIWalgreens At-Home A1C Test Kit, Software Version Revision D, REF Number 3056
A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
- 2019-12-25FDA-DevicePolymer Technology SystemsClass IIA1CNow SELF CHECK (European), Software Version Revision D, REF Numbers 3054, 3062
A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
- 2019-12-25FDA-DevicePolymer Technology SystemsClass IIA1CNow+ Systems (professional use product), Software Version Revision D, REF Numbers US: 3021, 3024; EU: 3037, 3038, 3057, 3058
A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
- 2019-12-25FDA-DeviceCarl Zeiss Meditec, Inc.Class IIVisuMax Software Version 2.10.13 with activated Software-Module ReLEx (SMILE) option used in conjunction with the VisuMax Femtosecond Laser System for Refractive Correction. Catalog Number 0000000-1345-518 Product Usage: The VisuMax Femtosecond Laser is a precision ophthalmic surgical laser designed for the creation of incisions in the cornea. The VisuMax Femtosecond Laser is indicated for use in small incision lenticule extraction (SMILE) for the reduction or elimination of myopia. This d
VisuMax devices with software version 2.10.13 and activated SW-Module ReLEx (SMILE) option may experience a software issue. In case of a suction loss during treatment the software allows the user to choose the option immediate restart or restart treatment. Suction loss can occur in phase 1 through phase 5. The software defect refers to Phase 2 (between 10% and 100% of lower lenticule cut) only. The software offers a flap cut, but due to the software issue it performs a cap cut instead of a flap cut if the user proceeds.
- 2019-12-25FDA-DeviceRoche Diagnostics Operations, Inc.Class IIRoche Homocysteine-In vitro test for the quantitative determination of total L-homocysteine in human serum and plasma on Roche/Hitachi cobas c systems Catalog Number: 06542921190 - - Product Usage: The assay can assist in the diagnosis of patients suspected of having hyperhomocysteinemia or homocystinuria.
Homocysteine Reagent, Calibration Failures and Quality Control Recovery Issues on cobas c 701 and 702 modules
- 2019-12-25FDA-DeviceGE Healthcare, LLCClass IIAestiva 7100, Model Numbers: 1. 1006-9320-000 2. 1006-9321-000 3. 1006-9320-000-001101 4. 1006-9021-000 5. 1006-9012-000 6. 1006-9111-000 7. 1006-9321-000-007639 8. 1006-9321-000-000023 9. 1006-9321-000-015180 10. 1006-9320-000-010369 11. 1006-9321-000-019486 12. 1006-9321-000-007601 13. 1006-9321-000-009266 14. 1006-9321-000-008526 15. 1006-9321-000-116507 16. 1006-9321-000-009244 17. 1006-9321-000-012769 18. 1006-9320-000-004497 19. 1006-9320-000-009697 20. 1006-9320-000
Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.
- 2019-12-25FDA-DeviceGE Healthcare, LLCClass IIAespire 7100, Model Numbers: 1. 1009-9000-000-009469 2. 1009-9000-000 3. 1009-9000-000-013913 4. 1009-9000-000-001366 5. 1009-9000-000-009857 6. 1009-9003-000 7. 1009-9000-000-008120 8. 1009-9000-000-009600 9. 1009-9000-000-014290 10. 1009-9000-000-010585 11. 1009-9000-000-004993 12. 1009-9011-000-004795 13. 1009-9000-000-013407 14. 1009-9000-000-004604 15. 1009-9000-000-017674 16. 1009-9000-000-008763 17. 1009-9000-000-008306 18. 1009-9000-000-003277 19. 1009-9000-000-00416
Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.
- 2019-12-25FDA-DeviceGE Healthcare, LLCClass IIAestiva 7900, Model Numbers: a) 1006-9300-000 b) 1006-9305-000 c) 1006-9118-000 d) 1006-9112-000 e) 1006-9113-000 f) 1006-9020-000 g) 1006-9033-000 h) 1006-9117-000 i) 1006-9119-000 j) 1006-9030-000 k) 1006-9032-000 l) 1006-9022-000 m) To be provided n) 1006-9120-000 o) 1006-9027-000 p) 1006-9003-000 q) 1006-9025-000
Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.
- 2019-12-25FDA-DeviceGE Healthcare, LLCClass IIAestiva MRI, Model Numbers: a) 1006-9310-000 b) 1006-9110-000 c) 1006-9023-000 d) 1006-9028-000 e) 1006-9310-000-305077 f) 1006-9310-000-015243 g) 1006-9310-000-017602 h) 1006-9310-000-103785 I) 1006-9310-000-025109 j) 1006-9310-000-009650 k) 1006-9310-000-015224 l) 1006-9310-000-031881 m) 1006-9310-000-031854 n) 1006-9310-000-026571
Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.
- 2019-12-18FDA-DeviceMaquet Cardiovascular Us Sales, LlcClass IIMaquet Cardiopulmonary (MCP)/Getinge ROTAFLOW Drive, Model Number 701022161. The device is a component of the ROTAFLOW System Console.
A loose ODU plug / Coaxial cable connection may result in fluctuating flow values on the Console.
- 2019-12-18FDA-DeviceGETINGE US SALES LLCClass IIMaquet Cardiopulmonary GmbH / Getinge Quadrox-iD Pediatric Diffusion Membrane Oxygenator with Bioline Coating, Sterile REF: BEQ-HMOD 30000
Sterile barrier system may be compromised
- 2019-12-18FDA-DeviceViewray, Inc.Class IIMRIdian Linac Radiation Therapy System Model 10000 and 20000
Issue with transferring treatment regiment from one system to another. Adequate instructions for transferring treatment plan(s) are not provided within the operator's manual or within the software operating system.
- 2019-12-18FDA-DeviceNeuronetrix Solutions, LLC dba CognisionClass IICognision, Subject Kit (Clinical Grade), Model 3000- containing Hydro Dot brand Wave Prep cups-electroconductive cream for use with external electrodes. Order No. D5000, Part No. PN-1720 - Product Usage: An electroconductive cream for use with external electrodes to reduce the impedance of the contact between the electrode surface and the skin. The cream is not intended for use with stimulating electrode. The Wave Prep¿ cups are a component in the Subject Kit.
Product has the potential to be contaminated with Burkholderia cepacia (B. cepacia
- 2019-12-18FDA-DeviceVolcano CorporationClass IIPhilips Volcano FFR software used in the following systems connected to the McKesson, Schwarzer, and some Philips Xper Hemodynamic Systems (HMS): (1) Volcano Imaging System s5i, Part #807400001; (2) s5ix Imaging System with s5iz PC Workstation Refurbished, Part#807400001 R and 807401200; (3) s5i Imaging System, Part#806300003; (4) s5i Imaging System, Part#5322466; (5) s5x Imaging System, Part#807300001; (6) s5x Imaging System Refurbished, Pat#807300001 R; (7) s5 Imaging System Refurbished, Part#
Interoperability issue that affects certain systems that use the recalled software when it is connected to specific Hemodynamic Systems (HMS).
- 2019-12-18FDA-DeviceLeica Microsystems, Inc.Class IIBOND Polymer Refine Detection Kit, Catalog No. DS9800. This detection system is for in vitro diagnostic use. Bond Polymer Refine Detection is a biotin-free, polymeric horseradish peroxidase (HRP)-linker antibody conjugate system for the detection of tissue-bound mouse and rabbit IgG and some mouse IgM primary antibodies. It is intended for staining sections of formalin-fixed, paraffin-embedded tissue on the BOND automated system.
BOND Polymer Refine Detection kits (DS9800) were held out of refrigerated storage longer than specified which may exhibit suboptimal staining.
- 2019-12-18FDA-DeviceConformis, Inc.Class IIiTotal CR Knee Replacement System Model Number: TCR-111-1111-010102 - Product Usage: The iTotal CR Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.
Medial and lateral polyethylene inserts did not securely fit into the cobalt chrome tray implant.
- 2019-12-18FDA-DeviceThe Binding Site Group, Ltd.Class IIOptilite C-Reactive Protein Kit - Product Usage: The Optilite C-Reactive Protein Reagent is intended for the quantitative in vitro determination of C-reactive protein (CRP) concentration in serum using the Binding Site Optilite analyser. Measurement of C-Reactive Protein aids in evaluation of the amount of injury to body tissues and for evaluation of infection, tissue injury, and inflammatory disorders. This test should be used in conjunction with other laboratory and clinical findings. The Opti
Deterioration of C-reactive protein (CRP) reagent in Optilite C-Reactive Protein Kits
- 2019-12-18FDA-DeviceElekta IncClass IIMonaco Radiation Treatment Planning System (RTP) System
Monaco is using the incorrect energy when optimizing and calculating dose.