- 2018-11-07FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIBrilliance Big Bore Oncology CT (Model No. 728243) software version 4.2.0 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.
It was reported that scanners running certain software versions provide confusing information in the "Impressions" section of the Calcium Score report. The generated report provides multiple analyses of the amount of plaque detected, as if they all apply to the patient at the same time. This misleading guidance can be potentially misleading to the user, and may lead to mistreatment of the patient.
- 2018-11-07FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIVereos PET/CT, Software version 2.0.1. Model Number: 882446 Product Usage: The Philips Vereos PET/CT system is a diagnostic imaging device that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem images anatomical cross-sections by computer reconstruction of X-ray transmission data. The PET subsystem images the distribution of PET anatomy-specific radiopharmaceuticals in the patient. The PET/CT system is used for the purposes of detecti
Software Correction: Software errors that may result in incorrect 4D CT images, PET images with SUV quantification error, or disruption of interventional CT procedures; and Software errors that may result in partial set of images, inability to generate CT or PET images, incorrect scan parameters, or scanning the incorrect portion of the body.
- 2018-11-07FDA-DeviceRoche Diagnostics CorporationClass IIRoche Diagnostics URISYS 1100 Urine Analyzer Part Number: 03617556001 Product Usage: The URISYS 1100 Urine Analyzer is a reflectance photometer designed to automatically read and evaluate the results of Chemstrip 5 OB, Chemstrip 7, Chemstrip 10 MD, and Chemstrip 10 UA urine test strips for various urine analytes.: specific gravity, pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin and blood
Limits of Detection (LoD) for protein, nitrite, leukocytes, and erythrocytes on the Urisys 1100 urine analyzer with Chemstrip 5 OB, Chemstrip 7, Chemstrip 10 MD, and Chemstrip 10 UA test strips were higher than what is listed in the test strip method sheets may lead to false negative results for the affected parameters
- 2018-11-07FDA-DeviceSiemens Medical Solutions USA, IncClass IIYsio MAX (model number 10762470) with Tube stand 3D V (model number 7042018 or 7042034), radiographic systems used in radiographic and tomographic exposures of the whole body.
In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.
- 2018-11-07FDA-DeviceSiemens Medical Solutions USA, IncClass IIYsio (model number 10281013 or 10281163) with Tube stand 3D V (model number 7042018 or 7042034), radiographic systems used for radiographic and tomographic exposures of the whole body.
In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.
- 2018-11-07FDA-DeviceSiemens Medical Solutions USA, IncClass IIAXIOM Multix MT (model number 8395415 or 8395399 or 8395381), a patient table used in conjunction with the Siemens Multix PRO- and TOP Systems radiographic exposures of the entire body.
In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.
- 2018-11-07FDA-DeviceSiemens Medical Solutions USA, IncClass IIMultix Fusion system is a radiographic system: Models: Multix Fusion Analog (model number 10746665) with Tube stand 3D V (model number 7042042 or 7042059); Multix Fusion Digital Portable (model number 10746666) with Tube stand 3D V (model number 7042042); Multix Fusion Digital Integrated (model number 10746667) with Tube stand 3D V (model number 7042042); Multix Fusion Digital Wireless (model number 10893300) with Tube stand 3D V (model number 7042042)
Steel cables inside the lifting column of the overhead tube, may rapture without triggering the safety lock, leading to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning
- 2018-11-07FDA-DeviceBioFire Diagnostics, LLCClass IIFilmArray Blood Culture Identification (BCID), IVD, Rx Only, BioFire Diagnostics, LLC, Panel Catalog Numbers/ UDI: RFIT-ASY-0126 (30 test kit)/ 00815381020086, RFIT-ASY-0127 (6 test kit)/ 00815381020093, when used with the following blood culture bottles: Bottle Description/Part Number: BACT/ALERT FA Plus/ 410851, BACT/ALERT FN Plus/ 410852, BACT/ALERT PF Plus/ 410853,
Increased risk of false positive results for Proteus when the product is used with a specific blood culture bottles.
- 2018-11-07FDA-DeviceArthrex, Inc.Class IILi-ion Battery Housing for AR-400 DrillSaw Sports 400 System Product Usage: The battery housing is used to house the battery that powers the AR-400 DrillSaw Sports 400 device.
The Li-ion Battery Housing for the AR-400 DrillSaw Sports 400 System may not provide a complete seal around the battery, allowing for potential fluid ingress.
- 2018-11-07FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7, a stand-alone software treatment planning system.
Robust optimization for plans using beam set + background dose is not supported in RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7.
- 2018-11-07FDA-DeviceCMP Industries, LlcClass IITechnoflux IMPAK Resina lfquida acrilica elastic Curado-Calor Tipo 1, Clase 1 REF: 3748GB. Product Usage - Impak Elastic Acrylic Resin Liquid, Used for Dental Splints and Dental Nightguards
Foreign material in product
- 2018-11-07FDA-DeviceCMP Industries, LlcClass IIIMPAK Elastic Acrylic Resin Liquid Heat Cure Type 1, Class I, 1 Quart (946ml) REF:3306. Product Usage - Impak Elastic Acrylic Resin Liquid, Used for Dental Splints and Dental Nightguards
Foreign material in product
- 2018-11-07FDA-DeviceStryker GmbHClass IIHoffman LRF Hexapod Strut, Extra Short 183mm-295mm, Model Number 4935-0-040. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures Post-traumatic joint contracture which has resulted in loss of range of motion Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction Pseudoarthrosis or non-union of long bones Limb lengt
The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in partial of full displacement or loss of the pin and, accordingly, stability of the strut of the frame cannot be ensured.
- 2018-11-07FDA-DeviceStryker GmbHClass IIHoffman LRF Hexapod Strut, Extra Short 131mm-191mm, Model Number 4935-0-030. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures Post-traumatic joint contracture which has resulted in loss of range of motion Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction Pseudoarthrosis or non-union of long bones Limb lengt
The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in partial of full displacement or loss of the pin and, accordingly, stability of the strut of the frame cannot be ensured.
- 2018-11-07FDA-DeviceStryker GmbHClass IIHoffman LRF Hexapod Strut, Extra Short 105mm-139mm, Model Number 4935-0-020. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures Post-traumatic joint contracture which has resulted in loss of range of motion Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction Pseudoarthrosis or non-union of long bones Limb lengt
The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in partial of full displacement or loss of the pin and, accordingly, stability of the strut of the frame cannot be ensured.
- 2018-11-07FDA-DeviceStryker GmbHClass IIHoffman LRF Hexapod Strut, Extra Short 89mm-109mm, Model Number 4935-0-010. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures Post-traumatic joint contracture which has resulted in loss of range of motion Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction Pseudoarthrosis or non-union of long bones Limb length
The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in partial of full displacement or loss of the pin and, accordingly, stability of the strut of the frame cannot be ensured.
- 2018-11-07FDA-DeviceZimmer Biomet, Inc.Class ISpF PLUS-Mini (60 ¿A/W) Implantable Spinal Fusion Stimulator SpF PLUS-Mini Implantable Fusion Stimulators are indicated as a spinal fusion adjunct to increase the probability of fusion success in 1 or 2 levels
Lack of adequate validation and controls to ensure that product cleanliness parameters were consistently met.
- 2018-11-07FDA-DeviceZimmer Biomet, Inc.Class IEBI Osteogen Implantable Bone Growth Stimulator The OsteoGen Implantable Bone Growth Stimulator is indicated in the treatment of long bone nonunions.
Lack of adequate validation and controls to ensure that product cleanliness parameters were consistently met.
- 2018-11-07FDA-DeviceMagellan Diagnostics, Inc.Class IIMagellan LeadCare¿ II Blood Lead Test Kit. In Vitro Test for the measurement of lead in fresh whole blood. Catalog Number: 70-6762 Primary DI No: 00850355006000 The LeadCare II Blood Lead Test Kit is for in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in fresh whole blood. This product is for professional use only. Sensors and treatment reagent of the test kit are specific for lead only. Contents of the LeadCare II Blood Lead Test Kit must only be us
Defect with the calibration button for lot 1808M. Level 1 and Level 2 Quality Control (QC) values will generate results of HIGH (>65 ¿g/dL). Results for patient testing with this lot will be displayed as HIGH.
- 2018-11-07FDA-DeviceAcumedia Manufacturers, Inc.Class IIICampylobacter Blood Free Selective Medium, PN 7527A Used with cefoperazone for the selective isolation of Campylobacter spp. In a laboratory setting. Not intended for use in the diagnosis of disease or other conditions in humans.
The product is labeled with an incorrect expiration date.
- 2018-11-07FDA-DeviceAcumedia Manufacturers, Inc.Class IIIPalcam Agar Base, Acumedia, PN 7669, Product Usage - Dehydrated culture media used with supplements as a selective and differential medium for the detection and isolation of Listeria monocytogenes from foods and environmental samples in a laboratory setting. Not intended for use in the diagnosis of disease or other conditions in humans.
Expiration date on the label is incorrect. Correct expiration date was November 30, 2020 and label on distributed product listed the expiration date as November 30, 2021.
- 2018-11-07FDA-DeviceArrow International IncClass IIArrow AGB + Multi-Lumen CVC Kit, Cat. No. CDC-42703-B1A. Product Usage - The Arrow CVC is intended to provide short-term (< 30 days) central venous access for treatment of diseases or conditions requiring central venous access, including, but not limited to the following: Lack of usable peripheral IV sites, Central venous pressure monitoring, Total parenteral nutrition (TPN), Infusions of fluids, medications, or chemotherapy, and Frequent blood sampling or receiving blood transfusions/blood prod
The product was shipped after its expiry date due to a system error. The product lidstock identifies the correct expiration date however the accompanying purchase order shipping documentation identifies an incorrect expiration date. Per standard clinical practice the lidstock would most likely be checked prior to use and the product would not be used leading to a minor delay while a replacement is located. In the unlikely event that the product lidstock is not checked prior to use, there is potential for use of expired product, and product functionality/or sterility cannot be guaranteed.
- 2018-11-07FDA-DeviceLeica Microsystems, Inc.Class IIBOND Slide Labeler Zebra Printer Model TLP3842 and Model GX4301, for use with BOND-III and BOND-MAX instruments
The power supply unit for the printer used with the device can potentially overheat or cause a fire hazard because it could contain an improperly-mixed phosphorus compound.
- 2018-11-07FDA-DeviceMaquet Datascope Corp - Cardiac Assist DivisionClass ICardiosave Rescue IABP, Part Number 0998-UC-0800-83
There is a potential for interruption and/or inability to start therapy to the patient prior to or during use of Cardiosave IABP for users who are at altitudes above 3200 Ft.
- 2018-11-07FDA-DeviceMaquet Datascope Corp - Cardiac Assist DivisionClass ICardiosave Hybrid IABP, Part Number 0998-00-0800-53
There is a potential for interruption and/or inability to start therapy to the patient prior to or during use of Cardiosave IABP for users who are at altitudes above 3200 Ft.
- 2018-10-31FDA-DeviceGE Medical Systems, LLCClass IIGE Healthcare, 3.0T 6 Channel Phased Array Flex Coil, M0050SS. Product Usage The 3.0T GE 6-Channel Phased Array Flex Coil is indicated for use on the order of the physician in conjunction with a 3.0T MRI scanner, as an accessory to produce 2D and 3D images.
Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue.
- 2018-10-31FDA-DeviceMedtronic Navigation, Inc.-LittletonClass IIO-arm 1000 2nd Edition Imaging System, OARM ASSY Bl70000027GERR SYS PRODUCT REFURB, Model Number Bl70000027GERR Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.
- 2018-10-31FDA-DeviceMedtronic Navigation, Inc.-LittletonClass IIO-arm 1000 2nd Edition Imaging System, OARM ASSY Bl70000027GER SYS PRODUCT GER, Model Number Bl70000027GER Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.
- 2018-10-31FDA-DeviceMedtronic Navigation, Inc.-LittletonClass IIO-arm 1000 2nd Edition Imaging System, BASE OARM Bl70000027230R SYSTEM 230V RWK, Model Number Bl70000027230R Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.
- 2018-10-31FDA-DeviceMedtronic Navigation, Inc.-LittletonClass IIO-arm 1000 2nd Edition Imaging System, BASE OARM Bl70000027230 SYSTEM 230V, Model Number Bl7000002 7230 Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.