Medical device recalls
39229 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2017-08-30FDA-DeviceMedtronic Navigation, Inc.Class IIMedtronic Instrument Kit 9733908 Fusion ENT, reusable, RX. This kit includes Kit 9733452 (containing 9733449, Straight Suction, 9733450, 70 deg Curved Suction, and 9733451, 90 deg Curved Suction). The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. T
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
- 2017-08-30FDA-DeviceMedtronic Navigation, Inc.Class IIMedtronic Instrument Set EM ENT, Ref. #9733452 and #9733452-G02, reusable, RX. This kit contains 9733449, Straight Suction, 9733450, 70 deg Curved Suction, and 9733451, 90 deg Curved Suction. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
- 2017-08-30FDA-DeviceMedtronic Navigation, Inc.Class IIMedtronic Suction 9734308 Small AxiEM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical to
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
- 2017-08-30FDA-DeviceMedtronic Navigation, Inc.Class IIMedtronic Curved Suction 90 9733451 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
- 2017-08-30FDA-DeviceMedtronic Navigation, Inc.Class IIMedtronic Curved Suction 70 9733450 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical t
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
- 2017-08-30FDA-DeviceMedtronic Navigation, Inc.Class IIMedtronic Straight Suction 9733449 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical to
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
- 2017-08-30FDA-DeviceAlcon Research, Ltd.Class IIORA System with VerifEye, Catalog Number 8065998300 For use during intraocular lens surgery
Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens coefficients for an IOL model being downloaded from the Alcon server in an incorrect order.
- 2017-08-30FDA-DeviceAlcon Research, Ltd.Class IIORA System with VerifEye+ Cart, Catalog Number 8065998307 For use during intraocular lens surgery
Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens coefficients for an IOL model being downloaded from the Alcon server in an incorrect order.
- 2017-08-30FDA-DeviceSolta Medical IncClass IIVASER PRO Amplifier, VASER 2.2 VASERlipo System stem, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring (modification of the anatomy).
Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017.
- 2017-08-30FDA-DeviceSolta Medical IncClass IIVASER Amplifier, VASER 2.2 VASERlipo System, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring (modification of the anatomy).
Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017.
- 2017-08-30FDA-DeviceCardinal Health 200, LLCClass IIConvertors Robotic Drape with Attached Leggings; Product is packaged as Single Sterile Drape/8 per Case
Product 29474 LAVH drape was packaged as 29574 Robotic Drape with Leggings.
- 2017-08-23FDA-DeviceBoston Scientific CorporationClass IIEncore(TM) 26 Inflation Device, REF Catalog Nos.: (a) H74904526011, (b) M001151050, (c) M00566670, (d) M0067101140
Boston Scientific (BSC) has received complaints for cracked trays on the Encore device.
- 2017-08-23FDA-DeviceC.R. Bard, Inc.Class IIArctic Sun 5000 Temperature Management System, 100-120 V North America and 100-120 V US
Potential failure of the Arctic Sun 5000 Temperature Management System may occur when the continued use is in excess of 2000 system hours without preventive maintenance.
- 2017-08-23FDA-DeviceBaxter Healthcare CorporationClass IIHomeChoice Pro Automated PD system Model numbers: 5C8310, 5C8310R Product Usage: The HomeChoice Automated Personal Cycler is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.
Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis (PD) therapy, damaging the cassettes for the HomeChoice or HomeChoice PRO cyclers. If damaged cassettes are used, the cyclers may not consistently detect very small holes/cuts in the sheeting of the cassette in the patient valve region and the cycler may deliver air into the patient.
- 2017-08-23FDA-DeviceBaxter Healthcare CorporationClass IIHomeChoice Automated PD system Model numbers: 5C4471, 5C4471R Product Usage: The HomeChoice Automated Personal Cycler is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.
Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis (PD) therapy, damaging the cassettes for the HomeChoice or HomeChoice PRO cyclers. If damaged cassettes are used, the cyclers may not consistently detect very small holes/cuts in the sheeting of the cassette in the patient valve region and the cycler may deliver air into the patient.
- 2017-08-23FDA-DeviceBoston Scientific CorporationClass IISQ-RX 1010 Pulse Generator, Rx. Product Usage: The S-ICD system is intended to provide defibrillation therapy for the treatment of life-threatening ventricular tachyarrhythmias in patients who do not have symptomatic bradycardia, incessant ventricular tachycardia, or spontaneous, frequently recurring ventricular tachycardia that is reliably terminated with anti-tachycardia pacing
The device can deliver an atypical amount of energy due to memory corruption inside the device.
- 2017-08-23FDA-DeviceBoston Scientific CorporationClass IIEMBLEM MRI S-ICD Model 219, Subcutaneous-Implantable Cardioverter Defibrillator, Rx. Product Usage: The S-ICD system is intended to provide defibrillation therapy for the treatment of life-threatening ventricular tachyarrhythmias in patients who do not have symptomatic bradycardia, incessant ventricular tachycardia, or spontaneous, frequently recurring ventricular tachycardia that is reliably terminated with anti-tachycardia pacing
The device can deliver an atypical amount of energy due to memory corruption inside the device.
- 2017-08-23FDA-DeviceBoston Scientific CorporationClass IIEMBLEM S-ICD, Model A209, Subcutaneous Implantable Cardioverter Defibrillator, RX. Product Usage: The S-ICD system is intended to provide defibrillation therapy for the treatment of life-threatening ventricular tachyarrhythmias in patients who do not have symptomatic bradycardia, incessant ventricular tachycardia, or spontaneous, frequently recurring ventricular tachycardia that is reliably terminated with anti-tachycardia pacing.
The device can deliver an atypical amount of energy due to memory corruption inside the device.
- 2017-08-23FDA-DeviceMagellan Diagnostics, Inc.Class IILeadCare II Blood Lead Test Kit, containing LeadCare II Lead Controls Product Usage: IVD test kit for the detection of lead in blood. The kit consists of sensors, treatment reagent, lead controls (two levels), heparinized capillary tubes and plungers, transfer droppers and a calibration button. The LeadCare II Lead Controls packaged in the kit lots listed in #2 below are the subject of this recall (Field Correction) action.
LeadCare II Test Kits controls out of range
- 2017-08-23FDA-DeviceCenorin, LLCClass IIThermaSure(TM), a CENORIN(TM) product, SERIES 1000 MEDICAL DEVICE DRYER: (a) Model 1009; (b) Model 1010 Forced air dryer for medical devices
Dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn.
- 2017-08-23FDA-DeviceCenorin, LLCClass IIThermaSure(TM), a CENORIN(TM) product, SERIES 300 MEDICAL DEVICE DRYER: (a) Model 330; (b) Model 340 Forced air dryer for medical devices
Dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn.
- 2017-08-23FDA-DeviceCenorin, LLCClass IIThermaSure(TM), a CENORIN(TM) product, SERIES 130 MEDICAL DEVICE DRYER: (a) Model 135; (b) Model 136 Forced air dryer for medical devices
Dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn.
- 2017-08-23FDA-DeviceICU Medical IncClass IIPlum 360 Infusion System
Potential for the Connectivity Engine Module of the affected unit to disengage from the main chassis, causing the unit to shut down without an alarm notification. This would delay initiation of therapy or interrupt an infusion.
- 2017-08-23FDA-DeviceCareFusion 303, Inc.Class IIVersaSafe Plastic Cannula - 11 Gauge, Model 9391-0200 Product Usage: Used with secondary IV sets or syringes with a female luer lock adapter to administer fluid and medications through a needle or catheter inserted into a patients artery or vein.
Punctures in blister packaging that was detected during the packaging process.
- 2017-08-23FDA-DeviceGE Medical Systems, LLCClass IISigna PET/MR (K163619, K142098) Product Usage: The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information, acquired simultaneously and isocentrically. The combined system maintains independent functionality of the MR and PET devices, allowing for single modality MR and/or PET imaging. These systems are intended to be utilized by appro
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
- 2017-08-23FDA-DeviceGE Medical Systems, LLCClass II1.5T Signa HDxt, 3.0T Signa HDxt, Signa HDi, Signa Vibrant Product Usage: The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, i
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
- 2017-08-23FDA-DeviceGE Medical Systems, LLCClass II1.5T Signa HDx, 3.0T Signa HDx, 1.5T Signa HDxt, 3.0T Signa HDxt, Signa HDi, Signa Vibrant Product Usage: The GE Signa¿ HDx MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa¿ HDx MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the en
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
- 2017-08-23FDA-DeviceGE Medical Systems, LLCClass IISIGNA Creator, SIGNA ExplorerProduct Usage: 1.5T SIGNA Creator and 1.5T SIGNA Explorer is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TM
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
- 2017-08-23FDA-DeviceGE Medical Systems, LLCClass IIDiscovery MR750 3.0T Product Usage: Intended use: The Discovery MR750 3.0Tsystem is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ,
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
- 2017-08-23FDA-DeviceGE Medical Systems, LLCClass IIOptima MR450w 1.5T Product Usage: The Optima MR450w 1.5T system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, he
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.