Moderate riskFDA DeviceClass II

Vesco Q feeding pump recalled for possible inaccurate fluid delivery

Vesco Medical is recalling certain Vesco Q enteral feeding pumps, model VED-701, with software versions 6.0 through 7.8. The pump may deliver fluids outside its labeled accuracy of plus or minus 5%, which could lead to over- or under-delivery compared with the prescribed dose if proper controls are not in place. The notice says the effects could be temporary or medically reversible, with a remote chance of serious harm, and no injuries have been reported.

From the FDA notice · Read the official notice ↗

What to do

  1. Contact your care team and Vesco Medical before changing how you use the pump.
  2. Check the model number (VED-701), software version (6.0 to 7.8) and serial number; affected units were made through 11/05/2025 (UDI-DI 00850002550498).
  3. The notice does not state a remedy; ask Vesco Medical how to get a correction or replacement.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Vesco Q Enteral Feeding Pump, Model Number VED-701, Software Versions 6.0 - 7.8
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 7 products — see all of them.

Details

Units affected
3,493 units
Illnesses reported
None reported
Manufactured in
60 Collegeview Rd Space 144, Westerville, Ohio, United States
Classification
Class II
Recall number
Z-0373-2027

In the FDA's words

Description
Vesco Q Enteral Feeding Pump, Model Number VED-701, Software Versions 6.0 - 7.8 The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences. UDI-DI: 00850002550498; Serial Numbers: all serial numbers manufactured though 11/05/2025, Software revisions through 7.8 Distribution: US Nationwide distribution. Affected quantity: 3,493 units
Hazard
The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences.

7 products in this recall

Vesco Q pumps, feed sets recalled for possible dosing inaccuracy

Vesco Medical is recalling 7 Vesco Q products: the Vesco Q Enteral Feeding Pump (model VED-701) and six Vesco Q feed sets (models VED-050, VED-051, VED-052, VED-053, VED-054 and VED-059). The pump may deliver enteral fluids outside its labeled accuracy of plus or minus 5%, so feeds may be over- or under-delivered compared with the prescribed dose. The agency says this could cause temporary or medically reversible effects, with a remote chance of serious harm. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

Related topics

OtherSoftwareMedical devicesSold nationwideMore from FDA Device