Moderate riskFDA DeviceClass II

Vesco Q 500mL enteral feed sets recalled for inaccurate delivery

Vesco Medical is recalling Vesco Q 500mL bag feed sets used with enteral feeding pumps. The feeding pump may deliver fluids outside its labeled accuracy of +/-5%, which could over-deliver or under-deliver feeding relative to the prescribed dose if proper controls are not in place. The notice says the likely effects are temporary or medically reversible, with a remote chance of serious harm, and no injuries have been reported.

From the FDA notice · Read the official notice ↗

What to do

  1. Contact your care team or home nutrition provider before changing or stopping your feeding.
  2. Check whether your feed set is model VED-050 with UDI-DI 00850002550450 and was manufactured through 09/01/2026.
  3. Contact Vesco Medical LLC about the recall; the notice does not list a remedy, so ask them how to get a replacement or correction.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Vesco Q 500mL Bag Feed Set (PVC), Model Number VED-050
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 7 products — see all of them.

Details

Units affected
121,889 untis
Illnesses reported
None reported
Manufactured in
60 Collegeview Rd Space 144, Westerville, Ohio, United States
Classification
Class II
Recall number
Z-0374-2027

In the FDA's words

Description
Vesco Q 500mL Bag Feed Set (PVC), Model Number VED-050 The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences. UDI-DI 00850002550450; Lot Numbers: all lots manufactured through 09/01/2026 Distribution: US Nationwide distribution. Affected quantity: 121,889 untis
Hazard
The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences.

7 products in this recall

Vesco Q pumps, feed sets recalled for possible dosing inaccuracy

Vesco Medical is recalling 7 Vesco Q products: the Vesco Q Enteral Feeding Pump (model VED-701) and six Vesco Q feed sets (models VED-050, VED-051, VED-052, VED-053, VED-054 and VED-059). The pump may deliver enteral fluids outside its labeled accuracy of plus or minus 5%, so feeds may be over- or under-delivered compared with the prescribed dose. The agency says this could cause temporary or medically reversible effects, with a remote chance of serious harm. No injuries have been reported.

The agency lists each product separately. Check every one you might have.