Vesco Medical Q feed sets recalled for feeding pump accuracy risk
Vesco Medical is recalling Vesco Q 1000mL bag feed sets (Model VED-054) because the feeding pump may deliver enteral fluids outside its labeled accuracy of plus or minus 5%. This could lead to giving more or less feed than the prescribed dose. The FDA classified the recall as Class II and the notice reports no injuries.
From the FDA notice · Read the official notice ↗
What to do
- Talk to your care team before changing or stopping feeding with this product.
- Check for Model VED-054 with UDI-DI 00810096150639 and any lot manufactured through 09/01/2026.
- The notice doesn't say how to get a replacement or refund; contact Vesco Medical or where you got the feed set.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Vesco Q 1000mL Bag Feed Set (PVC), Model Number VED-054
- Sold in
- Nationwide
- Made for
- All ages (under 100 years)
This recall covers 7 products — see all of them.
Details
- Units affected
- 208,650 units
- Illnesses reported
- None reported
- Manufactured in
- 60 Collegeview Rd Space 144, Westerville, Ohio, United States
- Classification
- Class II
- Recall number
- Z-0378-2027
In the FDA's words
- Description
- Vesco Q 1000mL Bag Feed Set (PVC), Model Number VED-054 The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences. UDI-DI 00810096150639; Lot Numbers: all lots manufactured through 09/01/2026 Distribution: US Nationwide distribution. Affected quantity: 208,650 units
- Hazard
- The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences.
7 products in this recall
Vesco Q pumps, feed sets recalled for possible dosing inaccuracy
Vesco Medical is recalling 7 Vesco Q products: the Vesco Q Enteral Feeding Pump (model VED-701) and six Vesco Q feed sets (models VED-050, VED-051, VED-052, VED-053, VED-054 and VED-059). The pump may deliver enteral fluids outside its labeled accuracy of plus or minus 5%, so feeds may be over- or under-delivered compared with the prescribed dose. The agency says this could cause temporary or medically reversible effects, with a remote chance of serious harm. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- Vesco Q Enteral Feeding Pump, Model Number VED-701, Software Versions 6.0 - 7.8Z-0373-2027 · 3,493 units
- Vesco Q 500mL Bag Feed Set (PVC), Model Number VED-050Z-0374-2027 · 121,889 untis
- Vesco Q 1000mL Bag Feed Set (LL), Model Number VED-051Z-0375-2027 · 1,650 units
- Vesco Q 40mm Screw Cap Feed Set, Model Number VED-052Z-0376-2027 · 2,880 units
- Vesco Q ENPlus Spike Feed Set, Model Numer VED-053Z-0377-2027 · 6,060 units
- Vesco Q 1000mL Bag Feed Set (PVC), Model Number VED-054 (this page)Z-0378-2027 · 208,650 units
- Vesco Q 100mL Small Bore Bag Feed Set, Model Number VED-059Z-0379-2027 · 2,220 units