Moderate riskFDA DeviceClass II

Vesco Q ENPlus Spike Feed Set recalled for possible dosing inaccuracy

Vesco Medical LLC is recalling Vesco Q ENPlus Spike Feed Sets, model VED-053, because the feeding pump may deliver enteral fluids outside its labeled accuracy of plus or minus 5%. This could cause over- or under-delivery of feeding compared with the prescribed dose. The notice says any effects would likely be temporary or medically reversible, with a remote chance of serious harm, and no injuries have been reported.

From the FDA notice · Read the official notice ↗

What to do

  1. Contact your care team before changing how you use this feed set.
  2. Check whether your feed set has UDI-DI 00850002550481 and a lot manufactured through 08/12/2026.
  3. The notice doesn't say how to get a remedy; contact Vesco Medical LLC.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Vesco Q ENPlus Spike Feed Set, Model Numer VED-053
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 7 products — see all of them.

Details

Units affected
6,060 units
Illnesses reported
None reported
Manufactured in
60 Collegeview Rd Space 144, Westerville, Ohio, United States
Classification
Class II
Recall number
Z-0377-2027

In the FDA's words

Description
Vesco Q ENPlus Spike Feed Set, Model Numer VED-053 The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences. UDI-DI 00850002550481; Lot Numbers: all lots manufactured through 08/12/2026 Distribution: US Nationwide distribution. Affected quantity: 6,060 units
Hazard
The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences.

7 products in this recall

Vesco Q pumps, feed sets recalled for possible dosing inaccuracy

Vesco Medical is recalling 7 Vesco Q products: the Vesco Q Enteral Feeding Pump (model VED-701) and six Vesco Q feed sets (models VED-050, VED-051, VED-052, VED-053, VED-054 and VED-059). The pump may deliver enteral fluids outside its labeled accuracy of plus or minus 5%, so feeds may be over- or under-delivered compared with the prescribed dose. The agency says this could cause temporary or medically reversible effects, with a remote chance of serious harm. No injuries have been reported.

The agency lists each product separately. Check every one you might have.