Moderate riskFDA DeviceClass II

Vesco Q 1000mL feed sets recalled for possible delivery inaccuracy

Vesco Medical is recalling Vesco Q 1000mL Bag Feed Sets (LL), model VED-051, which are used with enteral feeding pumps. The feeding pump may deliver fluids outside its labeled accuracy of plus or minus 5%, which can cause over- or under-delivery compared with the prescribed dose. The agency says this could cause temporary or medically reversible harm, with a remote chance of serious harm.

From the FDA notice · Read the official notice ↗

What to do

  1. Do not change your feeding routine on your own; contact your care team and Vesco Medical about the recall first.
  2. Check whether your feed set has UDI-DI 00850002550467; the recall covers all lots manufactured through 5/20/2026.
  3. The notice does not say how to get a remedy; ask Vesco Medical or your care team what to do next.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Vesco Q 1000mL Bag Feed Set (LL), Model Number VED-051
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 7 products — see all of them.

Details

Units affected
1,650 units
Illnesses reported
None reported
Manufactured in
60 Collegeview Rd Space 144, Westerville, Ohio, United States
Classification
Class II
Recall number
Z-0375-2027

In the FDA's words

Description
Vesco Q 1000mL Bag Feed Set (LL), Model Number VED-051 The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences. UDI-DI 00850002550467; Lot Numbers: all lots manufactured through 5/20/2026 Distribution: US Nationwide distribution. Affected quantity: 1,650 units
Hazard
The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences.

7 products in this recall

Vesco Q pumps, feed sets recalled for possible dosing inaccuracy

Vesco Medical is recalling 7 Vesco Q products: the Vesco Q Enteral Feeding Pump (model VED-701) and six Vesco Q feed sets (models VED-050, VED-051, VED-052, VED-053, VED-054 and VED-059). The pump may deliver enteral fluids outside its labeled accuracy of plus or minus 5%, so feeds may be over- or under-delivered compared with the prescribed dose. The agency says this could cause temporary or medically reversible effects, with a remote chance of serious harm. No injuries have been reported.

The agency lists each product separately. Check every one you might have.