Moderate riskFDA DeviceClass II

RaySearch RayStation software recalled for potential dose deformation

RaySearch Laboratories AB is recalling RayStation radiation therapy treatment planning software in several versions and builds. The software may deform dose after a deformable registration that expands outward near the edge of the original dose grid, possibly causing clinically relevant dose to be lost. No injuries have been reported.

From the FDA notice · Read the official notice ↗ · All Raysearch Laboratories recalls

What to do

  1. Before relying on deformable registration results, check whether your facility runs an affected version: 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.1.8, 8.1.2.5 or 9.0.0.113.
  2. Identify affected installations using the build numbers, UDI and GTIN codes, and software revisions listed in the notice.
  3. Contact RaySearch Laboratories AB about affected versions. The notice does not specify a remedy, so ask the manufacturer how to get a fix or instructions.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
RayStation Systems: RayStation 7, RayStation 8A, RayStation 8A SP1, RayStation 8B SP1, RayStation 8B SP2, RayStation 9A. Model/Build Numbers: 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.1.8, 8.1.2.5, 9.0.0.113. Radiation Therapy Treatment Planning System
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 9 products — see all of them.

Details

Units affected
6 units
Illnesses reported
None reported
Manufactured in
Eugeniavagen 18c, Stockholm, Sweden
Classification
Class II
Recall number
Z-0484-2027

In the FDA's words

Description
RayStation Systems: RayStation 7, RayStation 8A, RayStation 8A SP1, RayStation 8B SP1, RayStation 8B SP2, RayStation 9A. Model/Build Numbers: 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.1.8, 8.1.2.5, 9.0.0.113. Radiation Therapy Treatment Planning System Potential for dose deformation after using a deformable registration that expands outwards near the boundary of the original dose grid, possibly causing the clinically relevant dose to be lost. Model/Build Numbers: 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.1.8, 8.1.2.5, 9.0.0.113. UDI: 0735000201006820171130, 0735000201011220180608, 0735000201013620180918, 0735000201020420190214, 0735000201023520190524, 0735000201017420190612 GTIN: 07350002010068, 07350002010112, 07350002010136, 07350002010204, 07350002010235, 07350002010174. Serial Numbers: 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.1.8, 8.1.2.5, 9.0.0.113. Software Revision: RayStation 7, RayStation 8A, RayStation 8A SP1, RayStation 8B SP1, RayStation 8B SP2, RayStation 9A. Expiration Date: 2022-12-08, 2023-06-20, 2023-10-04, 2024-02-18, 2023-06-05, 2023-06-25 Distribution: Worldwide - US Nationwide distribution. Affected quantity: 6 units
Hazard
Potential for dose deformation after using a deformable registration that expands outwards near the boundary of the original dose grid, possibly causing the clinically relevant dose to be lost.

9 products in this recall

RayStation software recalled for potential dose deformation

RaySearch Laboratories is recalling 9 product entries of its RayStation radiation therapy treatment planning software, from RayStation 6.1.1 through RayStation 2024A SP3, distributed nationwide in the US. The agency says a deformable registration that expands outward near the boundary of the original dose grid may deform the dose, possibly causing the clinically relevant dose to be lost. The notice reports no injuries.

The agency lists each product separately. Check every one you might have.

Related topics

OtherSoftwareMedical devicesSold nationwideMore from FDA Device