RayStation 2023B planning software recalled for potential dose loss
RaySearch Laboratories AB is recalling RayStation 2023B, a radiation therapy treatment planning system, build 14.0.0.3338. When a deformable registration expands outward near the edge of the original dose grid, the clinically relevant dose may be deformed and possibly lost. The FDA classifies this as Class II, and no injuries have been reported.
From the FDA notice · Read the official notice ↗ · All Raysearch Laboratories recalls
What to do
- Check whether your system runs RayStation 2023B build 14.0.0.3338 (UDI 0735000201055620230630, GTIN 07350002010556).
- Review deformable registrations on this build for dose near the boundary of the original dose grid before relying on them.
- The notice does not specify a remedy; contact RaySearch Laboratories AB for next steps.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- RayStation 2023B. Model/Build Number: 14.0.0.3338. Radiation Therapy Treatment Planning System
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
This recall covers 9 products — see all of them.
Details
- Units affected
- 61 units
- Illnesses reported
- None reported
- Manufactured in
- Eugeniavagen 18c, Stockholm, Sweden
- Classification
- Class II
- Recall number
- Z-0490-2027
In the FDA's words
- Description
- RayStation 2023B. Model/Build Number: 14.0.0.3338. Radiation Therapy Treatment Planning System Potential for dose deformation after using a deformable registration that expands outwards near the boundary of the original dose grid, possibly causing the clinically relevant dose to be lost. Model/Build Number: 14.0.0.3338. UDI: 0735000201055620230630. GTIN: 07350002010556. Serial Numbers: 14.0.0.3338. Software Revision: RayStation 2023B. Expiration Date: 2028-07-07. Distribution: Worldwide - US Nationwide distribution. Affected quantity: 61 units
- Hazard
- Potential for dose deformation after using a deformable registration that expands outwards near the boundary of the original dose grid, possibly causing the clinically relevant dose to be lost.
9 products in this recall
RayStation software recalled for potential dose deformation
RaySearch Laboratories is recalling 9 product entries of its RayStation radiation therapy treatment planning software, from RayStation 6.1.1 through RayStation 2024A SP3, distributed nationwide in the US. The agency says a deformable registration that expands outward near the boundary of the original dose grid may deform the dose, possibly causing the clinically relevant dose to be lost. The notice reports no injuries.
The agency lists each product separately. Check every one you might have.
- RayStation Systems: RayStation 6.1.1, RayStation 6 SP2. Model/Build Numbers: 6.1.1.2, 6.2.0.7. Software
Radiation Therapy Treatment Planning SystemZ-0483-2027 · 3 units
- RayStation Systems: RayStation 7, RayStation 8A, RayStation 8A SP1, RayStation 8B SP1, RayStation 8B SP2, RayStation 9A. Model/Build Numbers: 7.0.0.19, 8.0.0…Z-0484-2027 · 6 units
- RayStation Systems: RayStation 9B SP1, RayStation 10A SP1. Model/Build Numbers: 9.2.0.483, 10.0.1.52.
Radiation Therapy Treatment Planning SystemZ-0485-2027 · 8 units
- RayStation 10B. Model/Build Number: 10.1.0.613
Radiation Therapy Treatment Planning SystemZ-0486-2027 · 0
- RayStation Systems: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2.
Model/Build Numbers: 11.0.0.951, 11.0.1.29, 11.0.3.116
Radiation Therapy Tre…Z-0487-2027 · 7 units
- RayStation Systems: RayStation 11B, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SP1, RayStation 11B SPT1.
Model/Build Numbers: 12.0.0.932, 12.0.…Z-0488-2027 · 15 units
- RayStation Systems: RayStation 12A, RayStation 12A SP1.
Model/Build Numbers: 13.0.0.1547, 13.1.0.144.
Radiation Therapy Treatment Planning SystemZ-0489-2027 · 7 units
- RayStation 2023B.
Model/Build Number: 14.0.0.3338.
Radiation Therapy Treatment Planning System (this page)Z-0490-2027 · 61 units
- RayStation 2024A SP3.
Model/Build Number: 15.1.3.10.
Radiation Therapy Treatment Planning SystemZ-0491-2027 · 60 units