Moderate riskFDA DeviceClass II

RaySearch RayStation 10B recalled for possible radiation dose loss

RaySearch Laboratories AB is recalling RayStation 10B, a radiation therapy treatment planning software, build 10.1.0.613. After a deformable registration expands outward near the edge of the original dose grid, the clinically relevant dose may be lost. The notice reports no injuries.

From the FDA notice · Read the official notice ↗ · All Raysearch Laboratories recalls

What to do

  1. Identify the software by model RayStation 10B, build 10.1.0.613, UDI 0735000201031020201216, and expiration date 2025-12-23.
  2. The notice does not give a fix or workaround. Contact RaySearch Laboratories AB about this version before relying on deformable registration results for this build.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
RayStation 10B. Model/Build Number: 10.1.0.613 Radiation Therapy Treatment Planning System
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 9 products — see all of them.

Details

Units affected
0
Illnesses reported
None reported
Manufactured in
Eugeniavagen 18c, Stockholm, Sweden
Classification
Class II
Recall number
Z-0486-2027

In the FDA's words

Description
RayStation 10B. Model/Build Number: 10.1.0.613 Radiation Therapy Treatment Planning System Potential for dose deformation after using a deformable registration that expands outwards near the boundary of the original dose grid, possibly causing the clinically relevant dose to be lost. Model/Build Number: 10.1.0.613. UDI: 0735000201031020201216. GTIN: 07350002010310. Serial Numbers: 10.1.0.613. Software Revision: RayStation 10B. Expiration Date: 2025-12-23. No products are currently placed on the US market Distribution: Worldwide - US Nationwide distribution. Affected quantity: 0
Hazard
Potential for dose deformation after using a deformable registration that expands outwards near the boundary of the original dose grid, possibly causing the clinically relevant dose to be lost.

9 products in this recall

RayStation software recalled for potential dose deformation

RaySearch Laboratories is recalling 9 product entries of its RayStation radiation therapy treatment planning software, from RayStation 6.1.1 through RayStation 2024A SP3, distributed nationwide in the US. The agency says a deformable registration that expands outward near the boundary of the original dose grid may deform the dose, possibly causing the clinically relevant dose to be lost. The notice reports no injuries.

The agency lists each product separately. Check every one you might have.