Moderate riskFDA DeviceClass II

RayStation 2024A SP3 software recalled for possible dose loss

RaySearch Laboratories AB is recalling RayStation 2024A SP3, build 15.1.3.10, a radiation therapy treatment planning system. After a deformable registration that expands outward near the boundary of the original dose grid, the clinically relevant dose may be lost. No injuries have been reported, and 60 units are affected.

From the FDA notice · Read the official notice ↗ · All Raysearch Laboratories recalls

What to do

  1. Check whether your facility uses RayStation 2024A SP3, build 15.1.3.10 (UDI 0735000201088420250129).
  2. Contact RaySearch Laboratories AB about affected units. The notice does not state a remedy.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
RayStation 2024A SP3. Model/Build Number: 15.1.3.10. Radiation Therapy Treatment Planning System
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 9 products — see all of them.

Details

Units affected
60 units
Illnesses reported
None reported
Manufactured in
Eugeniavagen 18c, Stockholm, Sweden
Classification
Class II
Recall number
Z-0491-2027

In the FDA's words

Description
RayStation 2024A SP3. Model/Build Number: 15.1.3.10. Radiation Therapy Treatment Planning System Potential for dose deformation after using a deformable registration that expands outwards near the boundary of the original dose grid, possibly causing the clinically relevant dose to be lost. Model/Build Number: 15.1.3.10. UDI: 0735000201088420250129. GTIN: 07350002010884. Serial Numbers: 15.1.3.10. Software Revision: RayStation 2024A SP3. Expiration Date: 2030-02-21. Distribution: Worldwide - US Nationwide distribution. Affected quantity: 60 units
Hazard
Potential for dose deformation after using a deformable registration that expands outwards near the boundary of the original dose grid, possibly causing the clinically relevant dose to be lost.

9 products in this recall

RayStation software recalled for potential dose deformation

RaySearch Laboratories is recalling 9 product entries of its RayStation radiation therapy treatment planning software, from RayStation 6.1.1 through RayStation 2024A SP3, distributed nationwide in the US. The agency says a deformable registration that expands outward near the boundary of the original dose grid may deform the dose, possibly causing the clinically relevant dose to be lost. The notice reports no injuries.

The agency lists each product separately. Check every one you might have.