RaySearch RayStation software recalled for potential dose deformation
RaySearch Laboratories AB is recalling certain RayStation 11B radiation therapy treatment planning software versions. The agency says that after a deformable registration that expands outward near the boundary of the original dose grid, the dose may be deformed, possibly causing the clinically relevant dose to be lost. No injuries have been reported.
From the FDA notice · Read the official notice ↗ · All Raysearch Laboratories recalls
What to do
- Check whether your installation uses one of the listed versions: 12.0.0.932, 12.0.3.68, 12.0.4.12, 12.1.0.1221, or 12.3.0.119.
- Contact RaySearch Laboratories AB and your care team about the affected software; the notice does not say what corrective action is available.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- RayStation Systems: RayStation 11B, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SP1, RayStation 11B SPT1. Model/Build Numbers: 12.0.0.932, 12.0.3.68, 12.0.4.12, 12.1.0.1221, 12.3.0.119. Radiation Therapy Treatment Planning System
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
This recall covers 9 products — see all of them.
Details
- Units affected
- 15 units
- Illnesses reported
- None reported
- Manufactured in
- Eugeniavagen 18c, Stockholm, Sweden
- Classification
- Class II
- Recall number
- Z-0488-2027
In the FDA's words
- Description
- RayStation Systems: RayStation 11B, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SP1, RayStation 11B SPT1. Model/Build Numbers: 12.0.0.932, 12.0.3.68, 12.0.4.12, 12.1.0.1221, 12.3.0.119. Radiation Therapy Treatment Planning System Potential for dose deformation after using a deformable registration that expands outwards near the boundary of the original dose grid, possibly causing the clinically relevant dose to be lost. Model/Build Numbers: 12.0.0.932, 12.0.3.68, 12.0.4.12, 12.1.0.1221, 12.3.0.119. UDI: 0735000201042620211208, 0735000201050120220422, 0735000201060020220620, 07350002010495, 0735000201057020221222. GTIN: 07350002010426, 07350002010501, 07350002010600, 07350002010495, 0735000201057. Serial Numbers: 12.0.0.932, 12.0.3.68, 12.0.4.12, 12.1.0.1221, 12.3.0.119. Software Revision: RayStation 11B, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SP1, RayStation 11B SPT1. Expiration Date: 2026-12-16, 2027-04-27, 2027-06-23, 2027-03-18, 2028-03-20. Distribution: Worldwide - US Nationwide distribution. Affected quantity: 15 units
- Hazard
- Potential for dose deformation after using a deformable registration that expands outwards near the boundary of the original dose grid, possibly causing the clinically relevant dose to be lost.
9 products in this recall
RayStation software recalled for potential dose deformation
RaySearch Laboratories is recalling 9 product entries of its RayStation radiation therapy treatment planning software, from RayStation 6.1.1 through RayStation 2024A SP3, distributed nationwide in the US. The agency says a deformable registration that expands outward near the boundary of the original dose grid may deform the dose, possibly causing the clinically relevant dose to be lost. The notice reports no injuries.
The agency lists each product separately. Check every one you might have.
- RayStation Systems: RayStation 6.1.1, RayStation 6 SP2. Model/Build Numbers: 6.1.1.2, 6.2.0.7. Software
Radiation Therapy Treatment Planning SystemZ-0483-2027 · 3 units
- RayStation Systems: RayStation 7, RayStation 8A, RayStation 8A SP1, RayStation 8B SP1, RayStation 8B SP2, RayStation 9A. Model/Build Numbers: 7.0.0.19, 8.0.0…Z-0484-2027 · 6 units
- RayStation Systems: RayStation 9B SP1, RayStation 10A SP1. Model/Build Numbers: 9.2.0.483, 10.0.1.52.
Radiation Therapy Treatment Planning SystemZ-0485-2027 · 8 units
- RayStation 10B. Model/Build Number: 10.1.0.613
Radiation Therapy Treatment Planning SystemZ-0486-2027 · 0
- RayStation Systems: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2.
Model/Build Numbers: 11.0.0.951, 11.0.1.29, 11.0.3.116
Radiation Therapy Tre…Z-0487-2027 · 7 units
- RayStation Systems: RayStation 11B, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SP1, RayStation 11B SPT1.
Model/Build Numbers: 12.0.0.932, 12.0.… (this page)Z-0488-2027 · 15 units
- RayStation Systems: RayStation 12A, RayStation 12A SP1.
Model/Build Numbers: 13.0.0.1547, 13.1.0.144.
Radiation Therapy Treatment Planning SystemZ-0489-2027 · 7 units
- RayStation 2023B.
Model/Build Number: 14.0.0.3338.
Radiation Therapy Treatment Planning SystemZ-0490-2027 · 61 units
- RayStation 2024A SP3.
Model/Build Number: 15.1.3.10.
Radiation Therapy Treatment Planning SystemZ-0491-2027 · 60 units