Moderate riskFDA DeviceClass II

RayStation 12A software recalled for potential dose deformation

RaySearch Laboratories AB is recalling RayStation 12A and RayStation 12A SP1 radiation therapy treatment planning software. Using a deformable registration that expands outward near the boundary of the original dose grid may deform the dose, possibly causing the clinically relevant dose to be lost. No injuries have been reported.

From the FDA notice · Read the official notice ↗ · All Raysearch Laboratories recalls

What to do

  1. Contact RaySearch Laboratories AB about the affected builds; the notice does not specify a remedy.
  2. Check whether your system runs RayStation 12A build 13.0.0.1547 or RayStation 12A SP1 build 13.1.0.144.
  3. Compare your software against the UDI, GTIN and expiration dates listed in the notice.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
RayStation Systems: RayStation 12A, RayStation 12A SP1. Model/Build Numbers: 13.0.0.1547, 13.1.0.144. Radiation Therapy Treatment Planning System
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 9 products — see all of them.

Details

Units affected
7 units
Illnesses reported
None reported
Manufactured in
Eugeniavagen 18c, Stockholm, Sweden
Classification
Class II
Recall number
Z-0489-2027

In the FDA's words

Description
RayStation Systems: RayStation 12A, RayStation 12A SP1. Model/Build Numbers: 13.0.0.1547, 13.1.0.144. Radiation Therapy Treatment Planning System Potential for dose deformation after using a deformable registration that expands outwards near the boundary of the original dose grid, possibly causing the clinically relevant dose to be lost. Model/Build Numbers: 13.0.0.1547, 13.1.0.144. UDI: 0735000201054920220616, 0735000201067920221007. GTIN: 07350002010549, 07350002010679. Serial Numbers: 13.0.0.1547, 13.1.0.144. Software Revision: RayStation 12A, RayStation 12A SP1. Expiration Date: 2027-06-29, 2027-10-24. Distribution: Worldwide - US Nationwide distribution. Affected quantity: 7 units
Hazard
Potential for dose deformation after using a deformable registration that expands outwards near the boundary of the original dose grid, possibly causing the clinically relevant dose to be lost.

9 products in this recall

RayStation software recalled for potential dose deformation

RaySearch Laboratories is recalling 9 product entries of its RayStation radiation therapy treatment planning software, from RayStation 6.1.1 through RayStation 2024A SP3, distributed nationwide in the US. The agency says a deformable registration that expands outward near the boundary of the original dose grid may deform the dose, possibly causing the clinically relevant dose to be lost. The notice reports no injuries.

The agency lists each product separately. Check every one you might have.