Medtronic MiniMed 770G insulin pump recalled for incorrect insulin delivery based on pump position
Check your MiniMed 770G pump position during use
Medtronic MiniMed insulin pumps may deliver too much or too little insulin depending on whether the pump is held above or below the infusion site on your body. Too much insulin can cause dangerously low blood sugar; too little can cause dangerously high blood sugar. This happens because gravity affects how insulin flows through the pump.
- Keep your pump at approximately the same height as your infusion site (where the needle enters your skin) during normal use
- Contact Medtronic or your healthcare provider for guidance on safe pump positioning for your specific infusion site location
- Monitor your blood sugar more closely than usual and watch for signs of very high or very low blood sugar
- Do not stop using your pump without speaking to your doctor first
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1742-2026More Medtronic MiniMed, Inc. recalls
- FDA-Device2026-04-08Medtronic MiniMed insulin pumps recalled for incorrect insulin delivery based on pump position
- FDA-Device2026-04-08Medtronic MiniMed Paradigm insulin pumps recalled for incorrect insulin delivery based on pump position
- FDA-Device2026-04-08Medtronic MiniMed insulin pumps recalled for incorrect insulin delivery based on pump position
- FDA-Device2026-04-08Medtronic MiniMed insulin pumps recalled for incorrect insulin delivery based on pump position
- FDA-Device2026-04-08Medtronic MiniMed insulin pumps recalled for incorrect insulin delivery based on pump position
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief