FDA medical device recalls
Formal FDA enforcement reports for medical devices — implants, monitors, diagnostic tests, and hospital and home-use equipment. 961 recalls in the last 12 months, in plain English.
Recent recalls
Page 9 of 33- Straumann GM Helix Implant recalled for dimension mismatch
Straumann USA LLC is recalling the GM Helix Implant (REF 109.984) because the product label on the packaging does not match the actual dimensions of the implant inside. This mismatch could lead to use of an incorrect implant size during dental procedures. The affected lot (VZPF4) was distributed nationwide to dental providers.
ModerateFDASep 23, 2026Straumann
- SunMed Tracheal Tube Size 2.5 recalled for obstruction risk
The 6Fr suction catheter may not pass smoothly through certain SunMed Tracheal Tube Size 2.5 samples (Model 1-7333-25), creating a potential obstruction during use. This issue affects multiple lot numbers distributed nationwide.
ModerateFDASep 23, 2026Well Lead
- GE Healthcare imaging software recalled for wrong-patient record risk
GE Healthcare is recalling six medical imaging and archive software products, with about 4,000 systems affected. Under certain conditions, historical imaging data moved from an older system can be linked to the wrong patient, so studies or images may appear in another patient's record. No injuries have been reported.
SeriousFDASep 23, 2026GE Healthcare6 products
- NOxBOXi nitric oxide system recalled for tubing leak risk
The oxygen tubing in certain NOxBOXi Nitric Oxide Delivery Systems may leak or disconnect, which could reduce oxygen flow to the patient. This happens because the tubing or its connection can fail. The recall affects 1,547 units with specific serial numbers.
ModerateFDASep 23, 2026Noxbox
- ICU Medical IV sets recalled for particulate contamination risk
ICU Medical is recalling eight IV administration sets with drip chambers. The drip chambers may have black specks, which may come from discolored plastic left behind during manufacturing. The company has not ruled out discoloration or particles in the fluid path. No injuries have been reported.
ModerateFDASep 23, 2026ICU Medical8 products
- Haylard procedure kits recalled for sterility risk from saline ampules
AVID Medical is recalling four Haylard procedure kits, about 2,300 in total, over possible product quality and sterility problems. The kits contain sodium chloride ampules that are also being recalled by Spectra Medical. The kits were distributed in Illinois, North Carolina, Virginia, and Wyoming, and no injuries have been reported.
SeriousFDASep 23, 2026AVID Medical4 products
- Abiomed Impella controllers recalled for purge sensor risk
Abiomed is recalling three groups of Impella controllers, about 10,750 units in total, that are used with heart pump systems. A component in the purge pressure sensor assembly, the purge flag, can fail, which may stop the controller from correctly recognizing the purge cassette. No injuries have been reported.
SeriousFDASep 23, 2026Abiomed3 products
- Bayer MEDRAD Stellant FLEX Syringe Kit recalled for sterilization issue
Bayer Medical Care has recalled MEDRAD Stellant FLEX Syringe Kits (catalog number FLEXS-150-QFT) used with the Stellant FLEX CT Injection System because some devices may not have been properly sterilized during manufacturing. The recall affects approximately 27,800 kits across the United States.
ModerateFDASep 23, 2026Bayer
- Maquet insertion kits recalled for sterile packaging risk
Maquet is recalling three Percutaneous Insertion Kit products because the packaging may not consistently maintain the sterile barrier. No injuries have been reported. About 271,000 units were distributed in the US and many other countries.
ModerateFDASep 23, 2026Maquet3 products
- GE Carescape Central Station recalled for monitoring delay risk
GE's Carescape Central Station patient monitoring system may take up to 3 minutes to resume monitoring after a restart. During this gap, clinically important changes in a patient's condition might not be detected right away.
ModerateFDASep 23, 2026GE
- Philips PIC iX patient monitoring systems recalled for software freeze
Philips is recalling two patient information system products because a software malfunction can cause the system to freeze and stop responding. The recall includes 7,671 affected units across the two products, which were distributed in the US and internationally. No injuries have been reported.
ModerateFDASep 23, 2026Philips2 products
- Medline Open Heart kit recalled for sterility risk
Medline's Open Heart Convenience Kit (DYNJ43008N) contains a coronary cannula with a potential sterile barrier breach. If a compromised pouch seal isn't caught before use, the non-sterile device could cause infection or serious complications including blood cell breakdown, blood clots, or organ damage.
SeriousFDASep 23, 2026Medline
- Nova Biomedical StatStrip glucose meters recalled for patient mix-up risk
Nova Biomedical is recalling five StatStrip glucose meter models, about 44,400 units in total. A software timing issue can cause a meter to keep the previous patient's or control solution's barcode on screen, so a glucose result may be recorded under the wrong patient. The glucose measurements themselves are not affected, and no injuries have been reported.
SeriousFDASep 23, 2026Nova Biomedical5 products
- Medline convenience kits recalled for fungal contamination
Medline is recalling seven groups of convenience kits because certain lots of sterile applicators may contain fungus (Aspergillus penicillioides). Skin preparation products contaminated with this fungus can cause serious infections, including sepsis, and death. The kits were distributed nationwide, and the FDA has classified this as a Class I recall, its most serious category.
SeriousFDASep 23, 2026Medline7 products
- Halyard IV start kit recalled for Aspergillus contamination
Halyard IV START KIT convenience kits (Lot 1658437) contain applicators contaminated with Aspergillus penicillioides, a fungus. The contamination was found in BD ChloraPrep Clear and FREPP Clear applicators included in the kits. This is a Class I recall affecting 288 units distributed in Nebraska.
SeriousFDASep 23, 2026AVID MedicalExpanded Sep 30
- BD Alaris infusion sets recalled for possible pump performance issues
BD Switzerland has recalled multiple models of Alaris Pump Infusion Sets due to discrepancies in performance data attributes. These discrepancies may cause the infusion pump to operate inappropriately. No injuries have been reported.
SeriousFDASep 23, 2026BD
- Bard dialysis catheters recalled as downstream Class I recall
Bard is recalling 2 product lines of short-term straight dialysis catheters, with about 172,500 units affected. The recall is downstream of a related Class I recall, RES 98777, and the filing does not describe the specific hazard. No injuries have been reported.
SeriousFDASep 23, 2026Bard2 products
- Windstone Medical AMS Convenience Kit recalled for detached port parts
Windstone Medical Packaging's AMS Convenience Kit DSAEK Pack contains saline solution bags with blue connector pieces that may detach from the infusion port during normal use. This could allow contamination or improper fluid delivery during ophthalmic surgery.
SeriousFDASep 16, 2026Windstone Medical Packaging
- Nipro MedicaLyte and Citric Complete dialysis concentrates recalled for endotoxin
Nipro is recalling four hemodialysis concentrate products because they may contain high levels of endotoxin, a substance from bacteria. The recall covers about 13,000 cases and 100 drums across the US, Guatemala, and Panama, with US distribution in FL, GA, IL, MD, MI, NC, NJ, NV, OH, PA. No injuries have been reported.
SeriousFDASep 16, 2026Nipro4 products
- HeartSine defibrillators recalled for electrical fault risk
HeartSine Technologies has recalled certain models of the SAM 350P, SAM 360P, and SAM 450P public access defibrillators (PADs) due to a damaged membrane component that can cause an intermittent electrical fault. The fault could prevent the device from operating properly when needed in a cardiac emergency. Approximately 11,912 affected units have been identified by serial number.
SeriousFDASep 16, 2026HeartSine
- Bard catheter trays recalled for contaminated pericare wipes
Bard is recalling six Foley and intermittent catheter tray products because the pericare wipes packed with them may have contamination on their surface. The recall followed customer complaints, and no injuries have been reported. The trays were distributed across the United States and in several other countries.
ModerateFDASep 16, 2026Bard6 products
- Getinge PVC thoracic catheters recalled for bacterial endotoxin
Atrium Medical is recalling three Getinge PVC thoracic catheter products, covering 11 model sizes, from a single lot. Testing found bacterial endotoxin levels above the allowed limit. No injuries have been reported.
ModerateFDASep 16, 2026Atrium Medical3 products
- Lutronic DermaV laser system recalled for burn risk
Lutronic's DermaV laser system may produce sparks and loud noise when used at higher energy levels (fluences) at the 1064 nm wavelength with current cooling settings. This spark hazard could result in patient burns during treatment.
ModerateFDASep 16, 2026Lutronic
- Halyard IV Start Kit recalled for seal integrity issue
Halyard IV Start Kit Non-Sterile (REF: KRIV26-02) convenience kits contain BD ChloraPrep Clear and FREPP Clear applicators with wrinkles in the paper lidding that may breach the seal. This compromises sterility assurance of the applicators. 98 kits in lot 1658438 (expiration 7/31/2026) were distributed nationally.
ModerateFDASep 16, 2026AVID MedicalExpanded Sep 30
- Medical Action epidural kits recalled for compromised sterility
Medical Action Industries is recalling two epidural kit products, 5,566 units in total. The kits contain saline ampules that are part of a separate recall, and the kits may not be sterile. No injuries have been reported.
SeriousFDASep 16, 2026Medical Action Industries2 products
- Given Imaging Bravo CF capsule devices recalled for possible detachment
The Bravo CF capsule delivery device is used during esophageal pH monitoring procedures to attach a small capsule to the esophagus. The capsule may fail to attach properly or detach from the delivery device during use. This can result in the capsule being inhaled, perforating the esophagus, blocking the airway, or causing bleeding.
SeriousFDASep 16, 2026Given Imaging
- Cardinal Health catheter trays recalled for possible contamination
Cardinal Health is recalling 12 procedure kits that include BD Urinary Catheter Trays with CliniClean Pericare Wipes. The wipes may be contaminated with a lubricant used during manufacturing. If a contaminated wipe is used on a patient, it may cause localized skin irritation, redness, or mild inflammation. No injuries have been reported.
ModerateFDASep 9, 2026Cardinal Health12 products
- GRANUFLO, NaturaLyte and Citrasate recalled for blue-green discoloration
Fresenius Medical Care is recalling three dialysis acid concentrate products, GRANUFLO, NaturaLyte and Citrasate, after complaints of blue-green discoloration in the product. No injuries have been reported. The products were distributed nationwide in the US and in Mexico.
ModerateFDASep 9, 2026Fresenius Medical Care3 products
- B Braun Discofix Stopcock Set recalled for mislabeled expiration date
Specific lots of B Braun's Discofix Four Way Stopcock extension sets have an incorrect expiration date printed on the carton shipper label. This labeling error could affect proper product tracking and usage verification.
ModerateFDASep 9, 2026B. Braun
- Medos Velys robotic surgery station recalled for arm movement risk
A software issue in the robotic controller of the Medos Velys Enabling Tech Robotic-Assist Station may cause the robotic arm to move unexpectedly during bone preparation or instrument positioning in orthopedic surgery. This affects 29 devices nationwide.
ModerateFDASep 9, 2026Medos